Description
1. Original backup power for your Cardiac Science defibrillator
Keep your defibrillator always ready for action with the original Cardiac Science battery, 12 V and 7500 mAh, supplied under the reference Cahnos 88888768, corresponding to factory code 9146-202/302. This is the OEM replacement for the Powerheart G3 / G3 Plus and G3 Elite equipment family, with primary lithium chemistry (Li-SO2), a shelf life of up to 4 years and the Intellisense chip that verifies battery status in the defibrillator’s daily self-test.
A direct, traceable replacement solution for hospitals, ambulances, workplaces, schools and public spaces with AEDs, backed by Cahnos technical-commercial support for medical electronics services, equipment fleets and tenders.
2. Technical explanation: how this battery works
The Cardiac Science Powerheart G3 / G3 Plus defibrillators (model 9146; series 9300A/9300E and 9390A/9390E) are automated external defibrillation devices that analyse cardiac rhythm and deliver the shock semiautomatically or automatically. The original battery 9146-202/302 is a non-rechargeable primary lithium pack (Li-SO2, sulphur dioxide) of 12 V and 7.5 Ah (90 Wh) that supplies the energy for analysis, monitoring and discharge throughout its service life.
Unlike a rechargeable battery, this battery cannot be recharged: it must be replaced with a new unit when the device indicates so. Its built-in Intellisense chip communicates the battery status, installation date and end of service life to the defibrillator, so that the daily self-test confirms the device is operational, with stable performance over a wide temperature range (from -20 °C to 55 °C according to manufacturer documentation). Installation takes seconds, by fitting the battery into the compartment without tools.
3. Technical Specifications
| Parameter | Detail |
| Reference Cahnos (SKU) | 88888768 |
| Original manufacturer reference | 9146 (designations of the same pack: 9146-001, 9146-002, 9146-102, 9146-202, 9146-302 and 9146-ABE) |
| Product type | Original replacement battery (OEM) for defibrillator |
| Technology / chemistry | Primary lithium (sulphur dioxide, Li-SO2), non-rechargeable |
| Nominal voltage | 12 V |
| Capacity | 7500 mAh (7.5 Ah) |
| Energy | 90 Wh (12 V × 7.5 Ah) |
| Service life (standby) | Up to 4 years from installation |
| Manufacture expiry | Approx. 5 years from the date of manufacture |
| Compatibility | Cardiac Science Powerheart G3, G3 Plus and G3 Elite: models 9146 and series 9300A/9300E, 9390A/9390E (units shipped after April 2004, according to manufacturer documentation) |
| Management system | Intellisense chip: daily self-test of the equipment and battery status reading |
| Temperature range | From -20 °C to 55 °C (according to the manufacturer’s documentation) |
| Installation | Direct replacement in the compartment, no tools required |
| Rechargeable | No (single-use primary battery; do not connect to a charger) |
| Presentation | 1 unit |
| Recycling | Do not dispose of as municipal waste; recycle in accordance with RAEE/WEEE regulations |
The main specifications (12 V, 7.5 Ah / 7500 mAh, 90 Wh, Li-SO2 chemistry, 4-year service life) come from the manufacturer’s technical documentation and authorised distribution of the original 9146 replacement; always confirm the reference stamped on the depleted battery before replacing.
4. Equivalences and Cross-References
The original reference 9146-202/302 corresponds to the primary lithium battery that Cardiac Science fits in the Powerheart G3 / G3 Plus family. Cardiac Science does not publish an official equivalence table authorising third-party batteries; the cited OEM designations are alternative numbers for the same factory pack, while the replacement references indicated below share chemistry (Li-SO2), voltage (12 V) and capacity (7.5 Ah) and are classified as comparable by category, not 1:1 interchangeable, and must be validated on a case-by-case basis (reading chip, connector and dimensions) before replacement.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
| Cardiac Science (factory designations) | Powerheart G3 / G3 Plus / G3 Elite (9146; 9300A/9300E, 9390A/9390E) | 9146 = 9146-001 = 9146-002 = 9146-102 = 9146-202 = 9146-302 (9146-ABE) | Exact reference equivalence (same manufacturer) | OEM numbers for the same primary lithium Li-SO2 pack, 12 V, 7.5 Ah (90 Wh), yellow casing with Intellisense chip; same performance and service life (4 years in service / 5 years shelf life); consistently confirmed in authorised distribution. Verify the exact variant printed on the label of the installed battery. |
| BBM Battery | Replacement battery for Powerheart AED G3 / G3 Plus | 7064 | Comparable by category (not interchangeable 1:1) | Lithium (Li-SO2), 12 V, 7500 mAh, non-rechargeable; replaces OEM 9146-102, 9146-202 and 9146-302; certified by Health Canada. Different replacement manufacturer (not original Cardiac Science); verify Intellisense chip reading, connector and dimensions against the device. Confirmed in distributor datasheet. |
| AMCO / Medical Battery Pros | Powerheart G3 / G3 Plus replacement battery | AM9146-2 (9146-2) | Comparable by category (not interchangeable 1:1) | Lithium (Li-SO2), 12 V, 7.5 Ah, non-rechargeable, yellow casing; crosses with 9146-001, 9146-002, 9146-102, 9146-202, 9146-302, 9142, 9143 and the 9300A/9300E series; approximate dimensions 233 × 52 × 70 mm. Different replacement manufacturer (not original Cardiac Science); validate reading chip, connector and dimensions before replacing. Confirmed in distributor datasheet. |
There is no equivalence table validated by the manufacturer (Cardiac Science) authorising third-party batteries for reference 9146; the cross-references indicated come from replacement manufacturers and specialised distributors, so it is recommended to consult with the technical department before use in tenders.
5. Selection tips and variants of interest
- Confirm the device model: this battery corresponds to the Cardiac Science Powerheart G3 / G3 Plus / G3 Elite defibrillator (model 9146; series 9300A/9300E and 9390A/9390E). Check the reference stamped on the depleted battery (9146-001, 9146-002, 9146-102, 9146-202, 9146-302 or 9146-ABE) before ordering.
- Non-rechargeable: this is a single-use primary lithium battery. Do not connect it to any charger or attempt to recondition it; replace it when the device indicates low battery or when it approaches the end of its service life (4 years from installation).
- Symptoms of depletion: daily self-test with failure or low battery warning, reduced analysis time or service indicator illuminated indicate that the pack must be replaced to keep the device operational in emergencies.
- Storage: store the battery in a dry place, at room temperature and inside its original packaging until installation; respect the factory expiry date (approx. 5 years).
- Environmental management: at the end of its service life, dispose of the battery at an authorised WEEE collection point; never discard it as household waste or incinerate it.
- Variants of the same family at Cahnos: for the Cardiac Science Powerheart 9145 defibrillator, there is a lithium battery 12 V / 7.5 Ah (Ref. Cahnos 88888683) and for the Powerheart G5, Ref. 88881203. Do not interchange these references between models without technical validation: each device requires its specific pack.
- Institutional orders: for consortia, tenders or maintenance contracts, contact the technical-commercial department of Cahnos to confirm exact compatibility with your batch of defibrillators and obtain a formalised quotation with special conditions for volume.





Reviews