Description
1. Introduction
The ZOLL AED PLUS is one of the lightest and most proven automatic external defibrillators (AED) on the market: just 3.1 kg, with a design intended to allow any rescuer, even without prior medical training, to intervene safely in the event of cardiac arrest. This configuration is supplied as a unit with protective cover and disposable type 123 lithium battery, without electrodes, designed for defibrillation programmes that already have original ZOLL electrodes in stock or wish to complete their provision according to their target population (Stat-Padz II, CPR-D-Padz or Pedi-Padz II).
It is the recommended solution for workplaces, schools, gyms, public administrations, homeowners’ associations and emergency services that need an accessible, reliable AED with minimal maintenance, with the freedom to choose the most appropriate electrodes for their environment.
2. Technical explanation of the equipment
The AED PLUS automatically analyses the patient’s cardiac rhythm through the multifunction electrodes and determines whether defibrillation is necessary. Its analysis algorithm evaluates the electrode connection and the electrocardiogram, and only recommends a shock for shockable rhythms: ventricular fibrillation with an average amplitude greater than 100 microvolts and wide complex ventricular tachycardia with rates greater than 150 lpm in adults or 200 lpm in children.
The energy is automatically selected in a preprogrammed manner (120 J, 150 J and 200 J) with rectilinear biphasic waveform, ZOLL’s proprietary technology that adjusts the current to the patient’s impedance. The unit has complete automatic checks: battery capacity, electrode connection, electrocardiogram, charge and discharge circuits, microprocessor hardware and software, CPR circuit and CPR-D sensor, and CPR audio circuit. All patient connections are electrically isolated, ensuring the safety of the rescuer and the patient. With the CPR-D-Padz electrodes (not included), the unit also offers the Real CPR Help system, which guides in real time the depth and rate of chest compressions through voice prompts and a metronome.
3. Technical Specifications
3.1. General specifications of the unit
| Field | Specification |
| Manufacturer | ZOLL Medical Corporation |
| Set type | Semi-automatic automated external defibrillator (AED) |
| Dimensions (height x width x depth) | 13.3 cm x 24.1 cm x 29.2 cm |
| Weight | 3.1 kg (one of the lightest on the market) |
| Waveform | Biphasic rectilinear |
| Energy selection | Automatic preprogrammed: 120 J, 150 J, 200 J |
| Weight-bearing time | Less than 10 seconds with new batteries |
| Charge retention time | 30 seconds |
| Patient safety | All patient connections electrically isolated |
| Automatic check | Configurable from 1 to 7 days (default value: every 7 days); monthly check at full energy (200 J) |
| Battery included | Disposable lithium battery type 123, pack of 10 cells, non-rechargeable (ref. 8000-0807-01; alternative 1008-0107-01) |
| Electrodes | Not included. Compatible: CPR-D-Padz, Pedi-Padz II or Stat-Padz II from ZOLL |
| Kit contents | AED PLUS unit + protective cover + disposable battery (without electrodes) |
| Presentation | Price per unit of equipment |
3.2. Automatic unit checks
| System checked | Detail |
| Battery capacity | Automatic verification of status and available charge |
| Electrode connection | Detection of correct connection of the multifunction pad |
| Electrocardiogram | Check of the ECG signal acquisition circuit |
| Charge and discharge circuits | Verification of the defibrillator energy circuits |
| Microprocessor hardware and software | Comprehensive system self-test |
| CPR circuit and CPR-D sensor | Check of the CPR circuit and compression sensor |
| CPR audio circuit | Verification of the speaker and voice prompts |
3.3. CPR guidance (Real CPR Help, with CPR-D-Padz electrodes)
| Parameter | Value |
| Frequency metronome | Variable between 60 and 100 compressions per minute (CPM) |
| Compression depth | From 3/4” to 3.5”; from 1.9 to 8.9 cm |
| Sensor | CPR-D sensor integrated in the electrode |
| Feedback | Voice prompts to adjust the depth and rate of compressions |
3.4. Shockable rhythms
| Rhythm | Criterion |
| Ventricular fibrillation | Average amplitude greater than 100 microvolts |
| Wide-complex ventricular tachycardia (adults) | Rates greater than 150 bpm |
| Wide-complex ventricular tachycardia (children) | Rates greater than 200 bpm |
For information on the sensitivity and specificity of the ECG analysis algorithm, refer to the AED Plus Administrator Guide.
View AED Plus catalogue | View AED PLUS technical specifications
4. Equivalences and Cross-References
Exact reference equivalence (kit consumables): the included battery corresponds to the official factory construction of ZOLL 8000-0807-01 (pack of 10 type 123 lithium batteries, non-rechargeable), confirmed in the official ZOLL store and at authorised distributors, with listed alternative 1008-0107-01.
Comparison with other brands: there is no equivalence table validated by ZOLL Medical Corporation that would allow the AED PLUS unit to be replaced by an AED from another manufacturer: the electrode connector, analysis algorithm and Real CPR Help system are proprietary to ZOLL. Any comparison with other leading AEDs (e.g. Philips HeartStart, Stryker/Physio-Control LIFEPAK or Defibtech) should be understood as comparable at category level, not interchangeable 1:1: they differ in waveform, algorithm, electrodes and accessories, although they share the same general clinical indication for cardiopulmonary resuscitation with biphasic defibrillation. Any specific cross-reference is pending confirmation with the Cahnos technical department before use in tenders.
| Reference | Product | Related items |
| 8000-0807-01 / 1008-0107-01 | Type 123 lithium battery pack (10 cells) | Official AED PLUS battery (included in this kit) |
| 8900-0801-01 | Stat-Padz II adult electrode (1 pair) | Compatible adult replacement (not included) |
| 8900-0800-01 | CPR-D-Padz adult single piece with Real CPR Help | Alternative with integrated CPR sensor (not included) |
| 8900-0810-01 | Pedi-Padz II paediatric electrode (1 pair) | Complementary for children < 8 years or < 25 kg (not included) |
| 8900-0802-01 | Stat-Padz II adult electrode (box of 12 pairs) | Same family, multipack format (not included) |
There is no equivalence table validated by the manufacturer between brands; consult the Cahnos technical department before using any cross-reference in a tender.
5. Selection tips and variants of interest
Complete the kit: this unit is supplied without electrodes; to make it operational, a pair of original electrodes from the ZOLL family must be added. For the general adult population, the Stat-Padz II (8900-0801-01) is the standard option; for real-time compression guidance, the CPR-D-Padz (8900-0800-01) with Real CPR Help technology, highly recommended for non-healthcare rescuers.
Paediatric population: if the setting cares for children (schools, sports centres), add a pair of Pedi-Padz II (8900-0810-01): the AED Plus detects their connection and automatically switches to paediatric energy levels.
Maintenance: replace the electrodes after each use or upon expiry (check the date on the packaging; service life of 18 to 24 months) and replace the battery pack (8000-0807-01) when the unit indicates; the configurable automatic check allows the status to be monitored without manual intervention.
Variants of the same unit at Cahnos: SER0A0PL000 (with adult Stat-Padz II electrode), SER0A0PL001 (with CPR-D-Padz electrode), SER0A0PL002 (with Stat-Padz II + Pedi-Padz II) and SER0A0PL003 (with CPR-D-Padz + Pedi-Padz II). This variant SER0A0PL000SE (only case and battery, without electrodes) is the lowest-cost option for those who already manage electrodes in stock.
Training: for CPR training programmes, the ZOLL AED PLUS TRAINER 2 (8008-0050-10) replicates the operation of the unit without delivering real shocks; always use training electrodes, never clinical ones. For public tenders that require equivalences, request documentary confirmation from our technical department before formalising the order.