Description
1. Commercial Introduction
Keep your Philips CodeMaster and HeartStart family defibrillator or monitor always ready to care for the smallest patients too. This pair of standard paediatric defibrillator electrodes with internal cable (Cahnos reference F7950P) is the replacement consumable indicated for paediatric defibrillation with CodeMaster 100, XE, XL, XL+ equipment and the HeartStart/HeartStream family. It is sold as a pair (2 units) in an individually heat-sealed pouch and is ready to be integrated into cardiopulmonary resuscitation protocols in emergency departments, paediatric intensive care units (PICU), ambulances, health centres and cardioprotected areas with Philips equipment. Available with immediate dispatch from Cahnos and specialised technical support for compatibility verification by model and connector.
2. Technical explanation of the equipment and procedure
Multifunction defibrillation electrodes are the interface element between the defibrillator and the patient: through them the device analyses the cardiac rhythm, delivers the transthoracic shock and, in many models, enables emergency ECG monitoring. In paediatric patients (under 8 years of age or weighing less than 25 kg) it is essential to attenuate the delivered energy to avoid myocardial injury; therefore this electrode has a reduced active surface area of 40 cm² per pad and must be used with the device’s paediatric mode switching or with the reduced energy setting (50–100 J) indicated in the defibrillator manual.
The integrated internal cable configuration provides a stable, direct connection to the device’s electrode socket, with no loose cables during resuscitation. The recommended placement is anterolateral (anteroapical position), so that both pads do not come into contact with each other on the infant’s chest. The low-impedance conductive adhesive gel ensures stable adhesion and a reliable signal, while the metal foil and 1 mm thick medical foam guarantee a uniform distribution of current. They are single-use and disposable: do not reuse under any circumstances and replace after each intervention or upon reaching the expiry date printed on the packaging. The electrode allows external defibrillation, synchronised cardioversion and ECG monitoring for up to 24 h.
3. Technical Specifications
| Parameter | Specification |
| Reference Cahnos (SKU) | F7950P (Cahnos distribution code; the equivalent original factory reference is indicated in block 4) |
| Product type | Paediatric multifunction defibrillator electrode pair, standard, with internal cable |
| Brand / compatible equipment | Philips – CodeMaster 100 (M2475B), CodeMaster XE (M1724A), CodeMaster XL (M1723A/B), CodeMaster XL+ (M1722A/B), HeartStart/HeartStream family |
| Target population | Paediatric (< 8 years / < 25 kg) |
| Total surface (per electrode) | 75 cm² (paediatric series) |
| Active surface (per electrode) | 40 cm² (paediatric series) |
| Adhesive ring width | 1 cm |
| Supported shocks | 50 shocks (energy 50–100 J in paediatric configuration; verify according to the equipment’s indications) |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Connection type | Safety anti-shock connector / clip, with integrated internal cable |
| Functions | External defibrillation, synchronised cardioversion and ECG monitoring |
| ECG monitoring | Up to 24 h |
| Operating temperature | Between 0 °C and +50 °C |
| Use | Single-use, disposable, non-sterile; latex-free and phthalate-free |
| Presentation | Price per pair (2 units); 5 pairs per box (paediatric series) |
| Primary packaging | 1 pair in heat-sealed polyethylene and aluminium bag, with laser pre-cut and opening notch |
| Shelf life | Approx. 30 months from the date of production; expiry indicated on the packaging |
| Storage | Dry place, between +5 °C and +35 °C |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC; EU Regulation 2017/745) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
Note: the values for active surface area (40 cm²), total area (75 cm²), shocks (50–100 J) and adhesive ring (1 cm) correspond to the standard paediatric electrode series with internal cable in this product family; confirm them against the technical data sheet of the supplied batch and the equipment manual before use in tendering.
Compatibility with defibrillators
| Equipment | Models | Cable / required accessories |
| Philips CodeMaster 100 | M2475B | Electrode cable M3507A (direct connection via internal cable) |
| Philips CodeMaster XE | M1724A | Direct connection |
| Philips CodeMaster XL / XL+ | M1723A/B · M1722A/B | Direct connection |
| Philips HeartStart / HeartStream | MRx (M3535A/M3536A) · XL (M4735A) · XLT (M3500B) · XL+ (861290) · HeartStart 4000 (M5500B) | Hands-free electrode cable M3508A depending on model; verify paediatric function and energy in manual mode |
Note: always check the connector fit and the cable of your equipment before clinical use.
4. Equivalences and Cross-References
The reference F7950P is the Cahnos distribution code for this compatible electrode; it is not an official Philips reference. Philips Healthcare does not publish an F7950P code: the equivalence with the manufacturer’s official nomenclature corresponds to the original paediatric electrodes of the family with barrel connector (M3504A and, for the infant range of 10–25 kg, M3501A) and to the alternative ordering code 989803106951 (M3504A), verified in the Philips technical specification sheet, in GUDID/FDA (DI 00884838022034) and in authorised distributors of original spare parts. Any comparison between brands must be understood as comparable at category level, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Philips | Infant Pads with barrel connector – original paediatric electrodes for HeartStart MRx/XL/XL+ and CodeMaster family | M3504A (also 989803106951) | Comparable by category (not interchangeable 1:1) | Original factory replacement with barrel connector, conductive surface of 75 cm² and 60 cm cable; requires M3507A/M3508A cables depending on the equipment. Code confirmed in Philips technical documentation, GUDID (DI 00884838022034) and authorised distributors; there is no Philips validation authorising the replacement of the original with the compatible spare F7950P. |
| Philips | Adult/Child Pads with barrel connector – original adult/child electrodes (≥ 10 kg) | M3501A (also 989803106921) | Comparable by category (not interchangeable 1:1) | Original adult/child with barrel connector; also covers the infant range of 10–25 kg. Code confirmed in the Philips technical specification sheet and authorised distributors. |
| Cardinal Health (Covidien/Kendall) | Medi-Trace™ Cadence Pediatric RTS – radiolucent paediatric electrodes with barrel connector compatible with Philips | 22661P | Comparable by category (not interchangeable 1:1) | Radiolucent paediatric pair for the CodeMaster XL+ family (M1722A/B, M1723A/B), HeartStream XLT, HeartStart XL, MRx and HeartStart 4000; conductive hydrogel with Ag/AgCl technology and 5 pairs per box. Code confirmed in GUDID (DI 20884527022356) and the official Cardinal Health catalogue; verify connector, cable and surfaces before replacement. |
| Nissha Medical Technologies (HeartSync) | 97-DEFIB F/G PEDIATRIC-PHILIPS – paediatric patch for the HeartStart XL/T, XL+, MRx and Intrepid family | 6600066H (D321) | Comparable by category (not interchangeable 1:1) | Radiolucent paediatric pair with tin and tin-copper conductive element; total surface of 78.4 cm²; packaged per pair with external cable in heat-sealed bag. Code confirmed in the manufacturer’s official catalogue; not suitable for use with AED: verify compatibility with the exact model of the equipment. |
| ZOLL Medical | Pedi-Padz II – original paediatric electrodes | 8900-0810 / 8900-0810-01 | Comparable by category (not interchangeable 1:1) | Original for AED Plus and AED Pro (preconnected, without external cable); hydrogel with tin conductive element and paediatric energy levels of 50/70/85 J. Code confirmed on the manufacturer’s website and GUDID (UDI 00847946001438). Not suitable for Philips equipment due to proprietary connector. |
| Stryker / Physio-Control | Infant/Child Reduced Energy Defibrillation Electrodes (LIFEPAK, pink connector) | 11101-000016 | Comparable by category (not interchangeable 1:1) | Paediatric electrode with energy reduction (200 J → ≈50 J) for the LIFEPAK 500, 1000, EXPRESS and CR Plus AEDs; proprietary QUIK-COMBO connector and algorithm, not suitable for Philips equipment. Code confirmed in the official Stryker store and GUDID (UDI 00721902629013). |
There is no equivalence table validated by the manufacturer authorising the replacement of original Philips electrodes with third-party patches: the connector, analysis algorithm and energy management are proprietary. Any replacement must be validated against the defibrillator’s technical data sheet and IFU; consult the Cahnos technical department before use in tenders. Within the Cahnos catalogue there are other references of the same family (standard paediatric with internal cable) intended for equipment from other manufacturers, not interchangeable with each other due to differing connector and algorithm: F7959P (CMOS Drake), F7967P (Schiller), F7970P (Laerdal Snap connection) and F7951P (ZOLL).
5. Selection tips and variants of interest
Confirm the equipment model: verify that your defibrillator is a Philips CodeMaster 100 (M2475B), XE (M1724A), XL (M1723A/B), XL+ (M1722A/B) or a compatible HeartStart/HeartStream and that the internal cable connector fits correctly into the equipment’s electrode socket before clinical use; on HeartStart 4000, XL/XLT and MRx models also verify the hands-free electrode cable (M3508A).
Do not confuse paediatric and adult versions: this pair is for patients under 8 years of age or under 25 kg; the adult version of the same Philips connector family is supplied under reference F7950, and the paediatric variant with external cable under F7950PW (original function M3504A).
Paediatric mode: ensure the equipment switches to paediatric mode or energy (reduced energy setting 50–100 J) according to the defibrillator manual; on manual adjustment models, always select it before the shock.
Check the expiry date: always check the date printed on the packaging and store in a dry place between +5 °C and +35 °C; expired electrodes or those with damaged packaging must not be used (the gel may lose conductivity). For the paediatric supply of Philips HeartStart FR2/FR2+ AEDs, limited energy references F7950PL (internal cable) and F7950WPL (external cable) are available, which reproduce the 150 J → 50 J attenuation of the original M3870A. If X-rays or CT scans are anticipated during care, consider the paediatric radiolucent series F7750P from the same connector family.
Plan your replacement: as this is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock according to the expiry date. For volume orders or tenders, contact the Cahnos technical department indicating the exact model and serial number of your equipment.