Description
1. Commercial Introduction
Ensure safe and effective cardiopulmonary resuscitation in the smallest patients with this pair of standard paediatric defibrillator electrodes with external cable (reference Cahnos F7951PW), a high-reliability consumable compatible with the ZOLL Medical defibrillators and monitors of the PD, E and M series. Designed for emergency departments, paediatric intensive care, medical transport and settings with professional-use AEDs, it combines the robustness of the standard series with the operability of the pre-connected external cable: the connector is always accessible and the device ready for immediate connection, a decisive advantage in paediatric cardiorespiratory arrest. A cost-optimised solution, compliant with European medical device standards (class IIb) and latex-free and phthalate-free.
2. Technical explanation of the equipment and procedure
Disposable multifunction electrodes are the interface between the defibrillator and the patient: through them the device analyses the rhythm, delivers therapy and monitors the response. This paediatric model is indicated for external transthoracic defibrillation, synchronised cardioversion and ECG monitoring (up to 24 h), adapted to paediatric physiology (children under 8 years or weighing less than 25 kg).
In paediatric patients it is essential to attenuate the delivered energy to avoid myocardial injury; therefore this electrode incorporates a reduced active surface of 40 cm² per electrode and must be used with the device’s paediatric mode switching or with the reduced energy setting (50–100 J) indicated in the defibrillator manual. The external cable variant allows the connector to be left accessible and the pad ready, reducing preparation time in emergencies. The low-impedance conductive adhesive gel ensures stable adhesion and a reliable signal, while the metal foil and 1 mm medical foam guarantee a uniform distribution of current. They are single-use: they must be replaced after each intervention or upon reaching the expiry date printed on the packaging, placed according to the anatomical diagrams (sternal and apical position) on the pouch.
3. Technical Specifications
Standard paediatric multifunction electrode from the paediatric defibrillator pad series, in its 100 cm external cable (pre-connected) variant.
| Field | Specification |
|---|---|
| Reference Cahnos (SKU) | F7951PW |
| Product type | Standard paediatric multifunction defibrillator electrode pair with external cable (pre-connected) |
| Indicated population | Children under 8 years or weighing less than 25 kg |
| Cable | External (pre-connected), 100 cm |
| Functions | Defibrillation, synchronized cardioversion and ECG monitoring (max. 24 h) |
| Supported shocks | 50 to 100 J (paediatric series) |
| Total surface (per electrode) | 75 cm² |
| Active area (per electrode) | 40 cm² |
| Adhesive ring width | 0.7 cm |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (1 mm thick) |
| Connection type | Safety anti-shock connector |
| Presentation | 1 pair in heat-sealed polyethylene and aluminium pouch; 5 pairs per box (paediatric series) |
| Use | Single-use, disposable; latex-free and phthalate-free |
| Shelf life | Approx. 30 months from date of manufacture (check printed expiry date) |
| Conditions of use and storage | Operating temperature from 0 °C to +50 °C; store in a dry place between +5 °C and +35 °C |
| Classification | Class IIb medical device (Directive 93/42/EEC and amendments; EU Regulation 2017/745) |
| Regulatory standard | ANSI/AAMI DF-80; IEC/EN 60601-1, 60601-2-4; ISO 10993-10 |
Compatibility (equipment models)
| Equipment manufacturer | Compatible models |
|---|---|
| ZOLL Medical | PD 1200, PD 1400, PD 1600, PD 1700, PD 2000 |
| ZOLL Medical | E Series, M Series |
Technical note: always verify the correct insertion of the connector into the device socket before clinical use.
4. Equivalences and Cross-References
This electrode (SKU F7951PW) is a consumable compatible with the cited ZOLL equipment, not an original factory spare part. The correspondence with the equipment manufacturers’ references is as follows; any comparison between brands must be understood as comparable at category level, not 1:1 interchangeable.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| ZOLL Medical | Pedi·Padz II — multifunction paediatric electrodes (original equipment) | 8900-0810-01 | Comparable by category (not interchangeable 1:1) | Factory original with intelligent paediatric capability for AED Plus and AED Pro; in M/E/R Series it is used in manual mode. Hydrogel with tin conductive element, pre-connected, paediatric range 0–8 years / less than 25 kg with levels of 50/70/85 J. Code confirmed in the official ZOLL store and in GUDID (UDI 00847946001438). |
| Stryker / Physio-Control | Infant/Child Reduced Energy Defibrillation Electrodes (for LIFEPAK with pink connector) | 11101-000016 | Comparable by category (not interchangeable 1:1) | Paediatric electrode with energy reduction (200 J → ≈50 J) for the LIFEPAK 500, 1000, EXPRESS and CR Plus AEDs; proprietary QUIK-COMBO connector and algorithm, not suitable for ZOLL equipment. Code confirmed in the official Stryker store and in GUDID (UDI 00721902629013). |
| Defibtech | Lifeline AED Child / Pediatric Electrode Pads | DDP-200P | Comparable by category (not interchangeable 1:1) | Paediatric electrodes with integrated energy attenuation for children under 8 years or less than 25 kg; proprietary connector, algorithm and electronics of Lifeline equipment. GUDID/FDA UDI-DI registration 00815098020430. |
There is no equivalence table validated by ZOLL Medical that authorises the replacement of the original Pedi·Padz electrodes with third-party pads: the connector, the analysis algorithm and the energy management are proprietary. Any replacement must be validated against the technical data sheet and IFU of the defibrillator. Consult the technical department of Cahnos before use in tenders.
5. Selection tips and variants of interest
Confirm the model and connector of the device: this pair corresponds to the ZOLL defibrillators and monitors PD 1200, PD 1400, PD 1600, PD 1700, PD 2000, E Series and M Series. Verify the reference and expiry date printed on the packaging before replacement; do not use an expired electrode or reuse a pair after an intervention.
Choose the variant according to your workflow: the external cable of this model allows the device to be left pre-connected and ready in seconds; if you prefer an internal cable, the F7951P reference is available (same standard paediatric series). For artefact-free imaging requirements, consider the radiotranslucent series F7751P (internal cable) or F7751PW (external cable), with the same ZOLL compatibility. The adult version of this same family is the F7951W.
Maintenance and provisioning: manage stock rotation by expiry date (approx. 30 months from manufacture) and always keep a spare pair next to the device in centres with a paediatric population. For volume orders or tenders, contact the Cahnos technical department stating the exact model and serial number of the defibrillator.