Description
1. Introduction
Keep your CU Medical Systems defibrillator ready to treat even the smallest patients with this pair of standard paediatric defibrillator electrodes with external cable, reference Cahnos F7958PW. It is the clinical replacement indicated for emergency units, ambulances, paediatric practices, schools and any heart-safe location with children: it ensures rapid and safe care in children under 8 years of age or weighing less than 25 kg, with energy adapted to paediatric resuscitation.
Designed to connect to the Paramedic CU-ER2, CU-ER3 and Paramedic CU-ER5 (with ER-5 OA03 cable) equipment, as well as to the Lifegain CU-HD1, these electrodes are supplied in pairs, are single-use and are delivered in a heat-sealed package that preserves the gel and the expiry date.
2. Technical explanation of the equipment and procedure
The Paramedic CU-ER series defibrillators acquire and analyse the electrocardiogram through the multifunction electrodes connected to the patient: they detect shockable rhythms (ventricular fibrillation and ventricular tachycardia) and, in manual or semi-automatic mode, deliver the shock that reverses the arrhythmia. These standard paediatric electrodes are provided with external cable and safety anti-shock connector, ensuring a stable attachment to the device cable and efficient current transmission.
For paediatric care with CU-ER series equipment, the corresponding pair of paediatric electrodes must always be used: according to the Paramedic CU-ER2 operator’s manual, paediatric defibrillation requires selecting 150 joules (AED or Manual mode with energy at 150 J). Depending on the model, the paediatric electrodes may require the paediatric adapter (ER1-OA12 / ER2-OA12 depending on the device) in the ECG-DEFIB port; on the Paramedic CU-ER5 the connection is made with the ER-5 OA03 cable. The pads incorporate low-impedance conductive adhesive gel, which ensures optimal contact with the chest skin and also allows, in addition to defibrillation, heart rhythm monitoring for up to 24 hours. They are placed following the anatomical diagrams on the package (anterior-posterior position) and, after use or upon reaching the expiry date, must be replaced with a new pair.
3. Technical Specifications
| Parameter | Specification |
| Reference Cahnos (SKU) | F7958PW |
| Electrode type | Standard multifunction paediatric with external cable, 1 pair (2 pieces), disposable |
| Series / family | Standard paediatric series with external cable |
| Brand / compatible equipment | CU Medical Systems – Paramedic CU-ER2, CU-ER3, Paramedic CU-ER5 with cable ER-5 OA03 and Lifegain CU-HD1 |
| Indicated population | Children under 8 years or weighing less than 25 kg (55 lb) |
| Minimum width of the adhesive ring | 0.7 cm |
| Total area (per electrode) | 75 cm² |
| Active area (per electrode) | 40 cm² |
| Number of shocks supported | 50 shocks at 100 J |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Connection type | Safety anti-shock connector or clip (depending on model) – external cable |
| Functions | Rhythm analysis, defibrillation and ECG monitoring (max. 24 h) |
| Use | Single-use, disposable |
| Composition | Latex-free and phthalate-free |
| Primary packaging | 1 pair in heat-sealed polyethylene and aluminium bag, with laser pre-cut and easy-open notch |
| Sales format | 5 pairs in cardboard box (paediatric series) |
| Operating temperature | Between 0 °C and +50 °C |
| Storage | Dry place between +5 °C and +35 °C; occasional exposure from −30 °C to +65 °C for a maximum of 7 days |
| Shelf life | 30 months from the date of production |
| Expiry | As indicated on the packaging; do not use after the expiry date |
| Regulatory standard | Class IIb medical device compliant with Directive 93/42/EEC (MDD); ANSI/AAMI DF-80; IEC/EN 60601-1 and 60601-2-4; ISO 10993-10 |
Compatibility note: confirm the exact model of your device and the connector fit before use. The electrode corresponds to the Paramedic CU-ER and Lifegain CU-HD1 series; it is not interchangeable with consumables from other CU Medical AEDs (i-PAD SP1, SP2, CU-SPR, NFK200 or NF1200/NF1201 series), which use their own electrodes and connectors.
4. Equivalences and Cross-References
CU Medical Systems documents in Appendix C of the Paramedic series operator manuals its original paediatric electrodes: reference ER2-PP01 (Reduced-energy Pediatric Defibrillator Electrode Pads) for the Paramedic CU-ER2, with its paediatric connection adapter ER2-OA12; for the CU-ER1 it uses reference ER1-PP01 with the ER1-OA12 adapter. The Cahnos distribution code F7958PW reproduces the function of these standard paediatric electrodes for CU-ER2, CU-ER3, CU-ER5 and Lifegain CU-HD1. CU Medical does not publish an official equivalence table authorising third-party electrodes for these devices: the connector, analysis algorithm and energy management are proprietary. Any comparison with other brands must be understood as comparable at category level, not 1:1 interchangeable, verifying connector, energy range and algorithm before any replacement.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| ZOLL Medical | Pedi-Padz II (paediatric electrode) | 8900-0810-01 | Comparable by category (not interchangeable 1:1) | Same paediatric range (<8 years / <25 kg) and automatic switching to paediatric energy, but proprietary ZOLL connector and algorithm; exclusive to AED Plus and AED Pro. Code confirmed in the official ZOLL store. |
| Philips | HeartStart FRx SMART Pads II (adult/child electrode) | 989803139261 | Comparable by category (not interchangeable 1:1) | Universal adult/child electrodes: require the Infant/Child Key to activate paediatric mode; proprietary Philips connector and algorithm. Code confirmed in the official Philips store. |
| Defibtech | Lifeline AED Child / Pediatric Electrode Pads | DDP-200P | Comparable by category (not interchangeable 1:1) | Paediatric electrodes with integrated energy attenuator for children under 8 years or under 25 kg; connector and management electronics proprietary to Lifeline equipment. Code confirmed at authorised distributors. |
| Cardiac Science | Powerheart G3 Pediatric Electrode Pads | 9730-002 | Comparable by category (not interchangeable 1:1) | Reduced energy (approx. 50 J) for patients under 8 years or under 25 kg; exclusive to Powerheart G3 series (9200/9300), with own connector and attenuation. Code confirmed at authorised distributors. |
Other homologous families in the sector with paediatric AED electrodes (e.g. Philips M3870A Infant/Child Reduced Energy Pads or the Child Pads from Stryker/Physio-Control LIFEPAK) are comparable only by category: check one by one the equipment connector, the paediatric energy range and the algorithm before considering any alternative. There is no equivalence table validated by the manufacturer (CU Medical) for this electrode; consult the technical department at Cahnos before using it in tenders.
5. Selection tips and variants of interest
Confirm the equipment model: this electrode is indicated for the Paramedic CU-ER2, CU-ER3 and Paramedic CU-ER5 defibrillators with ER-5 OA03 cable, as well as for the Lifegain CU-HD1. It is not valid for the CU Medical i-PAD AEDs (SP1, SP2, CU-SPR, NFK200 or NF1200/NF1201), which use their own consumables.
Check the expiry date: before installing, check the expiry date on the packaging and do not force the use of an expired electrode: the gel may lose conductivity and compromise the effectiveness of the discharge.
Paediatric energy setting: on the Paramedic CU-ER2, paediatric defibrillation requires selecting 150 joules (AED or Manual mode) according to the operator’s manual; on other models in the CU-ER series, follow the equipment manufacturer’s instructions for the paediatric configuration and connection scheme (paediatric adapter or ER-5 OA03 cable).
Paediatric provision: in centres with a paediatric population, always keep a pair of paediatric electrodes with the defibrillator and periodically check that the equipment remains operational with the pads in date.
Variants within the Cahnos catalogue: F7958P (standard CU Medical paediatric electrode with internal cable for the same equipment), F7958PL (reduced-energy paediatric electrode for Paramedic CU-ER1, original ref. ER1-PP01) and F7959PL / F7959WPL (limited-energy paediatric electrodes for the i-PAD NF1200/NF1201 AEDs from the Saver family, factory ref. CUA0512P).
Stock and tender management: establish FIFO rotation by expiry date and always replace the electrode after use or expiry. For volume orders or tenders, contact the technical department at Cahnos stating the exact model and serial number of the equipment.