Description
1. Commercial Introduction
Keep your automated external defibrillator Mediana HeartOn AED A10 always ready to also treat the smallest patients. This pair of standard paediatric electrodes with external cable (reference Cahnos F7969PW) is the replacement consumable indicated for paediatric defibrillation: it is sold as a pair (2 units) in an individual heat-sealed package and is ready to be integrated into the cardiopulmonary resuscitation protocols of emergency services, ambulances, health centres, schools and cardio-protected spaces equipped with the HeartOn AED series. Its original factory replacement is identified by the code Mediana A0647-0 (old reference A0257-0).
2. Technical explanation of the equipment and procedure
Multifunction defibrillation electrodes are the interface element between the defibrillator and the patient: they transmit the energy of the transthoracic shock and allow, in many models, emergency ECG monitoring. In paediatric patients (under 8 years of age or weighing less than 25 kg) it is essential to attenuate the delivered energy to avoid myocardial injury; therefore this electrode features a reduced active surface of 40 cm² per electrode and must be used with the device’s paediatric mode switch or with the reduced energy configuration indicated in the HeartOn AED A10 manual.
The recommended placement is anterolateral (anteroapical position), so that both pads do not come into contact with each other on the infant’s chest. The low-impedance conductive adhesive gel ensures stable adhesion and a reliable signal, while the metal foil and 1 mm medical foam guarantee a uniform distribution of the current. They are single-use and disposable: do not reuse under any circumstances and replace after each intervention or upon reaching the expiry date printed on the packaging.
3. Technical Specifications
| Specification | Detail |
|---|---|
| Reference Cahnos (SKU) | F7969PW |
| Factory code of the original Mediana spare part | A0647-0 (old reference A0257-0) |
| Brand / compatible equipment | Mediana – AED HeartOn AED A10 |
| Electrode type | Paediatric standard with external cable, multifunction |
| Target population | Paediatric (< 8 years / < 25 kg) |
| Total surface (per electrode) | 75 cm² |
| Active surface (per electrode) | 40 cm² |
| Adhesive ring width | 1 cm |
| Supported shocks | 50 shocks (energy 50–100 J in paediatric configuration; verify according to the equipment’s indications) |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Conductive material | Metal foil |
| Conductive gel | Low-impedance conductive adhesive gel |
| Connection type | Safety anti-shock connector or clip (depending on model), external cable |
| Maximum ECG monitoring time | 24 h |
| Use | Single-use, disposable, non-sterile |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, 60601-2-4; ISO 10993-10 |
| Latex / phthalates | Latex-free |
| Shelf life | 30 months from the date of production; expiry indicated on the packaging |
| Storage | Dry place, between +5 °C and +35 °C |
| Presentation | Price per pair of electrodes (2 units); 5 pairs per box in the paediatric series |
Note: the values for active surface (40 cm²), total area (75 cm²), shocks (50–100 J) and adhesive ring (1 cm) correspond to the series of standard paediatric electrodes with external cable in this product family; confirm them with the technical data sheet of the batch and the HeartOn AED A10 manual before use in tendering.
4. Equivalences and Cross-References
This electrode (SKU F7969PW) belongs to the family of standard paediatric consumables with external cable for defibrillation. The original replacement from the equipment manufacturer, Mediana HeartOn AED A10, is identified by the codes A0647-0 (old reference A0257-0), verified at authorised distributors of original Mediana spare parts (Defibshop and LivCor, Australia); the manufacturer confirms that they are the same pad with an updated order code. Equivalences with other brands are comparable only at category level: the connector, the analysis algorithm and energy management are proprietary to each device, so they are not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Mediana (original equipment spare part) | Paediatric AED electrodes HeartOn AED A10 (child pads) | A0647-0 / A0257-0 | Comparable by category (not interchangeable 1:1) | Original factory spare part for the same AED HeartOn AED A10; the manufacturer indicates that A0647-0 replaces the old A0257-0 (same pad, new order code). Verified at authorised distributors of original spare parts; there is no manufacturer validation authorising the use of the compatible Cahnos spare part as a substitute for the original: the substitution must be validated against the technical data sheet and IFU of the device. |
| ZOLL Medical | Pedi·padz® II (original paediatric electrodes) | 8900-0810 / 8900-0810-01 | Comparable by category (not interchangeable 1:1) | Factory original for AED Plus and AED Pro (pre-connected, without external cable); hydrogel with tin conductive element and paediatric levels of 50/70/85 J. Code confirmed on the manufacturer’s website and GUDID (UDI 00847946001438). Not suitable for Mediana equipment due to proprietary connector. |
Within the Cahnos catalogue there are other references in the same family (standard paediatric with external cable) intended for equipment from other manufacturers, so they are not interchangeable with each other due to differing connector and algorithm: F7950PW (Philips), F7961PW (Nihon Kohden), F7964PW (Weinmann MEDUCORE), F7967PW (Schiller), F7984PW (Instramed), F7986PW (Mediana D500) and F7987PW (ZOLL). As contrast references of the same clinical category in other equipment, the original spare parts Weinmann WM45419 (MEDUCORE), Instramed 79048 (i.ON / CardioMax / ISIS) and Nihon Kohden P-740K stand out: all comparable by category and not interchangeable 1:1 due to proprietary connector and energy management. The adult version of the same Mediana HeartOn AED A10 equipment is the F7969W.
There is no equivalence table validated by the manufacturer; consult the technical department before use in tendering.
5. Selection tips and variants of interest
Confirm the device model: verify that your defibrillator is a Mediana HeartOn AED A10 and that the external cable connector fits correctly into the device socket before clinical use. If you have a Mediana D500, use reference F7986PW; they are different connection systems.
Do not confuse the paediatric and adult versions: this pair is for patients under 8 years of age or under 25 kg; for adult patients of the same HeartOn AED A10 use reference F7969W.
Paediatric mode: ensure the AED switches to paediatric mode/energy when connecting these electrodes or select it manually according to the device manual; the energy in paediatric configuration must be limited to 50–100 J.
Check the expiry date: always check the date printed on the packaging and store in a dry place between +5 °C and +35 °C; expired electrodes or those with damaged packaging must not be used (the gel may lose conductivity).
Plan your replacement: as this is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock according to the expiry date. For volume orders or tenders, contact the technical department of Cahnos stating the exact model and serial number of your device.






Reviews