Ensure the paediatric readiness of your Philips defibrillator with this pair of limited-energy defibrillation electrodes, reference Cahnos F7950WPL, designed as a compatible replacement for the HeartStream / HeartStar FR2 AED (M3840A, M3841A) and FR2+ AED (M3860A, M3861A) models. It is the recommended solution for schools, nurseries, sports centres, paediatric practices, ambulances and any public-access defibrillation point where children may be present: when this paediatric electrode is connected, the device delivers a safe shock adapted to the physiological characteristics of the smallest patients.
This consumable replicates the function of the original factory paediatric electrode Philips M3870A (Infant/Child Reduced-Energy Defibrillation Pads), maintaining the same connection type and the same limited-energy concept. Single-use product, available for shipping and with specialised technical support from Cahnos.
2. Technical explanation of the equipment and procedure
The automated external defibrillator (AED) analyses the cardiac rhythm through the multifunction electrodes connected to the patient and, if it detects a shockable rhythm (ventricular fibrillation or ventricular tachycardia), delivers a biphasic shock. In paediatric patients —under 8 years of age or weighing less than 25 kg— applying the energy intended for adults can be harmful to the child’s myocardium. Therefore, this electrode is limited-energy: the paediatric series replicates the behaviour of the original Philips M3870A electrode, which contains electronics that attenuate the device’s shock from 150 J to 50 J, a dose appropriate for paediatric resuscitation in accordance with international CPR guidelines (SMART Biphasic therapy).
The pads are placed following the anatomical diagrams on the packaging and allow, in addition to defibrillation, monitoring of the cardiac rhythm during care (max. 24 h). They incorporate low-impedance conductive adhesive gel on a medical foam backing with a conductive metal foil, ensuring optimal skin contact and efficient current transmission. They are single-use: they must be replaced after each intervention or upon reaching the expiry date printed on the packaging.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Reference Cahnos (SKU) | F7950WPL |
| Original factory reference that this reproduces | Philips M3870A (Infant/Child Reduced-Energy Defibrillation Pads) |
| Product type | Paediatric multifunction limited-energy defibrillation electrode, 1 pair (2 pieces), disposable |
| Compatible brand | Philips (HeartStream / HeartStar) |
| Compatible models | HEARTSTREAM / HEARTSTAR FR2 AED (M3840A, M3841A) and FR2+ AED (M3860A, M3861A) |
| Indicated population | Children under 8 years or weighing less than 25 kg (55 lb) |
| Energy | Limited for paediatric use; the paediatric series replicates the 150 J → 50 J attenuation of the original Philips electrode (SMART Biphasic therapy) |
| Functions | External defibrillation, synchronized cardioversion, transcutaneous pacing and ECG monitoring (max. 24 h) |
| Total surface (per electrode) | 75 cm² (paediatric series) |
| Active surface (per electrode) | 40 cm² (paediatric series) |
| Adhesive ring width | 0.7 cm |
| Supported shocks | 50–100 J (paediatric series) |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (1 mm thick) |
| Connection type | Safety anti-shock connector type W (Philips family) |
| Composition | Latex-free and phthalate-free |
| Use | Single patient, non-reusable |
| Sterility | Non-sterile |
| Presentation | 1 pair in heat-sealed polyethylene and aluminium pouch; 5 pairs per box (paediatric series) |
| Regulatory standard | Medical device class IIb (Directive 93/42/EEC); ANSI/AAMI DF-80; IEC/EN 60601-1, 60601-2-4; ISO 10993 |
| Shelf life | 24–30 months from production date depending on batch (see labelling) |
| Expiry | As indicated on the packaging; do not use after the expiry date |
| Operating temperature | 0 °C to +50 °C |
| Storage | Dry place, between +5 °C and +35 °C; do not use if the packaging is open or damaged |
Equipment compatibility:
| Brand / Family | Compatible models |
|---|---|
| Philips HeartStream / HeartStar | FR2 AED (M3840A, M3841A) |
| Philips HeartStream / HeartStar | FR2+ AED (M3860A, M3861A) |
Technical note: the compatibility data, target population and limited energy correspond to the manufacturer’s technical documentation for the paediatric series; before placing your order, confirm the exact model of your device.
4. Equivalences and Cross-References
The original factory reference that this compatible replacement substitutes is the paediatric electrode Philips M3870A (Infant/Child Reduced-Energy Defibrillation Pads, for HeartStart FR2/FR2+), verified in the GUDID/FDA regulatory database (UDI-DI 00884838023734; FDA class III; type: single-use paediatric external defibrillation electrode). Philips does not publish an official equivalence table authorising the interchange of pads between brands: the connector, energy attenuator and analysis algorithm are proprietary. The references listed below share the same type (limited-energy paediatric electrode for AED) and are classified as comparable by category, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Philips | Infant/Child Reduced-Energy Defibrillation Pads (HeartStart FR2 / FR2+) | M3870A (also 989803108851) | Comparable by category (not interchangeable 1:1) | Original factory reference that this compatible replacement substitutes for the same FR2/FR2+ devices; attenuates the energy from 150 J to 50 J (SMART Biphasic therapy). Verified in GUDID/FDA (UDI-DI 00884838023734) and on the official Philips website. |
| Philips | HeartStart OnSite / Home / HS1 Infant/Child SMART Pads Cartridge | M5072A | Comparable by category (not interchangeable 1:1) | Paediatric cartridge with attenuator 150 → 50 J; proprietary cartridge format and connector for the OnSite/HS1 series. Does not fit FR2/FR2+. |
| ZOLL Medical | Pedi-Padz II (AED Plus / AED Pro) | 8900-0810-01 | Comparable by category (not interchangeable 1:1) | Same paediatric range (0–8 years / <25 kg) with 50/70/85 J levels; connector, algorithm and management electronics exclusive to ZOLL. |
| Cardiac Science (ZOLL) | Powerheart Pediatric Pads (series 9200/9300) | 9730-002 | Comparable by category (not interchangeable 1:1) | Integrated attenuator at ~50 J for Powerheart G3 and GE Responder AED; proprietary connector; verified in GUDID/FDA. |
| Defibtech | Lifeline DDU-1XX Pediatric Rescue Pads | DDP-200P | Comparable by category (not interchangeable 1:1) | Paediatric electrodes with integrated attenuator for children under 8 years or <25 kg; proprietary Defibtech connector; verified in GUDID/FDA (UDI-DI 00815098020430). |
| CU Medical Systems | Paediatric electrode i-PAD NF1200 / NF1201 (Saver family) | CUA0512P | Comparable by category (not interchangeable 1:1) | Pre-connected paediatric electrode with automatic energy reduction (~50 J); exclusive to CU Medical i-PAD equipment. |
There is no equivalence table validated by Philips that cross-references the M3870A reference with third-party electrodes for HeartStart FR2/FR2+: the connector, energy attenuator and analysis algorithm are proprietary. Any comparison with other brands must be understood as comparable at category level, not interchangeable 1:1, checking one by one the equipment connector, energy range, target population and packaging expiry before any replacement. Consult the Cahnos technical department before use in tenders.
5. Selection tips and variants of interest
- Confirm the equipment model: this electrode corresponds to the Philips AEDs HeartStream / HeartStart FR2 (M3840A, M3841A) and FR2+ AED (M3860A, M3861A). Check the model on the equipment plate before ordering; it is not valid for equipment with a different connector (OnSite/HS1, FR3 in standby, etc.).
- Check the reference and expiry: verify on the packaging the code F7950WPL and the expiry date; do not force the installation of an expired electrode, as the gel may lose conductivity and compromise the discharge.
- Paediatric provision: in centres with a child population always keep a pair of paediatric electrodes with the AED, periodically check that the equipment is operational with the pads in date and immediately replace the pair after any actual use.
- Variants within the Cahnos catalogue: F7950PL (same paediatric family compatible with Philips with standard connector) and, for other equipment in the same category, F7952WPL (Physio-Control LIFEPAK pink connector), F7957WPL (Cardiac Science Powerheart / GE Responder AED), F7959WPL (CU Medical i-PAD NF1200/NF1201) and F7966WPL (Defibtech Lifeline), all with limited energy and not interchangeable 1:1 with each other.
- Maintenance: manage stock rotation by expiry (24–30 months depending on batch) and always replace the electrode after use or expiry. For volume orders or tenders, contact the Cahnos technical department stating the exact model and serial number of the equipment.