Description
1. Introduction
Keep your defibrillator always operational with this pair of adult radiotransparent defibrillator electrodes with external cable (Ref. Cahnos F7755W). Designed as a multi-brand replacement for equipment from the Nihon Kohden Cardiolife TEC, Welch Allyn AED/PIC, Artema Innomed and Artema S&W CardioAid, Dräger Cardiolog 2000 and MRL families, they simplify the consumables stock for emergency departments, ICUs, ambulances, health centres and cardioprotected spaces. With their external cable configuration, you can connect the electrodes to the device in seconds, without opening or handling the packaging at the critical moment, and leave the patient’s chest visible for radiological imaging without interference. Available with immediate dispatch from Cahnos and specialised technical support for compatibility verification by model and connector.
2. Technical Explanation
These pads are disposable single-use multifunction electrodes: they perform external defibrillation, synchronised cardioversion, transcutaneous pacing and ECG monitoring. The low-impedance conductive adhesive gel ensures homogeneous contact with the skin, low interface resistance and optimal transmission of the discharge energy; the inner metal foil distributes the current uniformly and the 1 mm medical foam backing adapts to the contours of the chest.
Being radiotransparent, the electrode and its cable do not generate artefacts in fluoroscopy, tomography or intraoperative monitoring, allowing the healthcare team to assess the rib cage, pacemaker or vascular access without removing the pads. The safety anti-shock connector allows direct connection without adapters on compatible devices, and the external cable (connector and cable outside the bag) leaves the pair pre-connected and ready for emergencies. The condition of use is for a single patient: after resuscitation they must be disposed of and replaced, always respecting the expiry date on the packaging.
3. Technical specifications and compatibility
3.1 Technical specifications
| Parameter | Specification |
|---|---|
| Reference Cahnos (SKU) | F7755W |
| Patient type | Adult (older than 8 years / more than 25 kg) |
| Electrode type | Disposable multifunction: defibrillation, synchronised cardioversion, transcutaneous pacing and ECG monitoring |
| Radiotransparency | Yes (electrode and cable), suitable for imaging procedures |
| Cable configuration | External cable: connector and 120 cm cable outside the bag, immediate connection |
| Total surface per electrode | 136 cm² |
| Active surface per electrode | 94 cm² |
| Adhesive ring width | 0.7 cm |
| Supported shocks | Up to 50 shocks at 360 J |
| ECG monitoring | Max. 24 h |
| Composition | Medical foam (1 mm), low-impedance conductive adhesive gel, metal foil and safety anti-shock connector |
| Use | Single patient only, do not reuse |
| Latex / phthalates | Free from latex and phthalates |
| Regulatory classification | Medical device Class IIb (Directive 93/42/EEC) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
| Operating temperature | 0 °C to +50 °C |
| Storage temperature | +5 °C to +35 °C (dry place, not in direct sunlight) |
| Shelf life | 30 months from the date of manufacture |
| Presentation | 1 pair in heat-sealed polyethylene and aluminium pouch; 10 pairs per box |
3.2 Compatibility with defibrillators
| Equipment brand | Compatible models | Cable / required accessories |
|---|---|---|
| Nihon Kohden | Cardiolife TEC-5500 Series, TEC-5600 Series, TEC-7600 Series, TEC-7700 Series, TEC-2202, TEC-2203, TEC-7531, TEC-8251 and TEC-8300 Series | Cable JC-755V (TEC-5500, TEC-7600, TEC-7700, TEC-2202, TEC-2203, TEC-7531, TEC-8251) · Cable K342A / JC-855V (TEC-5600, TEC-8300) |
| Welch Allyn | AED-10, AED-20, PIC-30, PIC-40 and PIC-50 | Paediatric adapter Welch Allyn 002173-U (AED-10) · Paediatric adapter Welch Allyn 002174-U (AED-20) |
| Artema Innomed | CardioAid 360B, CardioAid 200B, CardioAid 200, CardioAid 100, CardioAid XL+ and CardioAid XL | Safety anti-shock connector |
| Artema S&W | CardioAid LS, CardioAid MC and CardioAid MC+ | Safety anti-shock connector |
| Dräger | Cardiolog 2000 | Safety anti-shock connector |
| MRL | LITE and Life Quest | Safety anti-shock connector |
Note: compatibility refers to the connector platform of each family. Always verify the correct fit of the connector and cable of your device before clinical use.
4. Equivalences and Cross-References
Reference F7755W is confirmed in the manufacturer’s official catalogue of the electrodes as a pair of adult radiotransparent multifunction electrodes with external cable (cable outside the bag) for Nihon Kohden, Welch Allyn, Artema Innomed, Artema S&W, Dräger and MRL equipment, with a presentation of 10 pairs per box. This is a consumable compatible with the aforementioned defibrillators, not an original spare part certified by each equipment manufacturer: the connector and energy management of each defibrillator are proprietary. After verifying official manufacturer catalogues and regulatory databases, the following cross-references with confirmed codes in public sources are identified.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Nihon Kohden | Disposable adult/child/infant pad (original equipment) | P-740K | Comparable by category (not interchangeable 1:1) | Code confirmed in official Nihon Kohden documentation and authorised distribution: factory original with 1.5 m cable; covers AED-2100 series, AED-3100 and TEC-5500/5600/7600/7700/8300 series. Connector, gel and labelling proprietary to Nihon Kohden; it is not radiotransparent and the cable is inside the packaging. Verify connector fit before replacement. |
| Nihon Kohden | Disposable AED pad adult/child/infant (original for the device) | P-711 (order code H329) | Comparable by category (not interchangeable 1:1) | Code confirmed on the official Nihon Kohden Europe website and authorised distribution: original with 1.5 m cable, aimed at the Cardiolife AED and TEC-8300 families (including TEC-8342). Proprietary connector; its radiotransparent nature has not been verified. Verify compatibility before replacement. |
| Welch Allyn (AED line) | Original adult pad for AED-10 and AED-20 | 001855-U | Comparable by category (not interchangeable 1:1) | Original reference for AED-10 and AED-20, confirmed in authorised technical distribution: cable inside the packaging and without verified radiotransparent nature. It shares the AED-10/20 platform but is not interchangeable 1:1 with this electrode due to cable and connector configuration. |
| HeartSync (Nissha Medical Technologies) | Defibrillation pads compatible with Nihon Kohden TEC-5500/5600/7600/7700/8300 | T100AC-NIHONKOHDEN (ref. 6600265H) | Comparable by category (not interchangeable 1:1) | Code confirmed in the manufacturer’s official catalogue: adult radiolucent (radiotransparent) pair, total surface area 371 cm², with connector comparable to the original Nihon Kohden and cable integrated inside the packaging (2 m). Presentation and areas different from this reference; verify connector and cable (JC-755V / JC-855V depending on series) before replacement. |
For Dräger Cardiolog 2000, MRL LITE / Life Quest and Artema CardioAid equipment, no verifiable original electrode code has been found in public sources, so equivalences for these families remain pending confirmation with the manufacturer. The original paediatric adapters Welch Allyn 002173-U (AED-10) and 002174-U (AED-20) appear consistently in the manufacturer’s documentation and are the accessories required for paediatric use on these devices. There is no equivalence table validated by the manufacturer that authorises the replacement of original electrodes with third-party pads: the connector and energy management of each defibrillator are proprietary. Any replacement must be validated against the defibrillator’s technical data sheet and instructions for use; consult the technical department at Cahnos before use in tenders.
5. Selection tips and variants of interest
Identify the model and cable of your device: on Nihon Kohden, the TEC-5600 and TEC-8300 series use the K342A / JC-855V cable, while the remaining series in the table use the JC-755V. Always check the connector fit before clinical use.
Paediatric use on Welch Allyn AED-10 / AED-20: requires the original energy-reducing adapter 002173-U or 002174-U (2 units); without this accessory, therapy must not be applied to paediatric patients. For patients under 8 years of age or weighing less than 25 kg, the paediatric version of the same family is available.
Choose radiotransparent when imaging tests are anticipated: if the patient may require X-ray, fluoroscopy or CT during care, this version avoids artefacts without removing the electrode. If imaging is not required, the standard version with external cable from the same platform (F7955W) may be considered, at lower cost.
Variants of the same Cahnos family: radiotransparent version with internal cable (F7755) if you prefer a direct connection without a loose cable; paediatric radiotransparent version with external/internal cable (F7755P); and standard non-radiotransparent version with external cable (F7955W). For other platforms in the radiotransparent with external cable category, F7750W (Philips), F7751W (Zoll Medical / M&D), F7752W (Physio-Control / Bexen / Mindray), F7753W (Agilent HP / Laerdal) and F7760W (Metrax Primedic) are available, not interchangeable with each other due to proprietary connector.
Storage and expiry: store in a dry place between +5 °C and +35 °C and respect the expiry date indicated on the packaging; the conductive gel may dry out and reduce its conductivity after the expiry date. As this is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock (FIFO). For volume orders or tenders, contact the technical department at Cahnos stating the exact model and serial number of your device.