Description
1. Commercial Introduction
Ensure the operability of your defibrillation point with this pair of preconnected adult/paediatric multifunction electrodes, designed for the automated external defibrillators (AED) of the brand Nihon Kohden from the Cardiolife AED 2100, 2150K and 2151K family (the accessory manufacturer also lists them for the 2152K). Their lead-out or preconnected configuration keeps the cable and connector accessible from the outside of the sealed pouch, so the device is ready for analysis and shock delivery without opening the packaging until the moment of use: a decisive advantage in the management of cardiorespiratory arrest in emergency services, workplaces, industrial estates, schools and public spaces with AEDs.
2. Technical explanation of the equipment and procedure
The disposable multifunction electrodes allow, with a single pair of adhesive pads: external transthoracic defibrillation, synchronised cardioversion, transcutaneous cardiac pacing and ECG monitoring. Their low-impedance adhesive conductive gel ensures stable adhesion and a reliable signal, while the 1 mm medical foam backing and the internal metal foil guarantee a uniform distribution of current with minimal risk of burns.
As it is an adult/paediatric model, the same pair of electrodes serves patients over 10 kg: in Nihon Kohden Cardiolife devices, energy switching is performed using the physical ‘adult/paediatric’ switch on the defibrillator itself, which automatically reduces the shock in paediatric mode. The connector, with anti-shock safety, is inserted into the device socket following the anatomical diagrams on the packaging. It is a single-use product: it must be replaced after each intervention or upon reaching the expiry date printed on the packaging.
3. Technical Specifications
| Product type | Disposable adult/paediatric multifunction electrodes, pre-connected (lead-out) |
| Compatibility | Nihon Kohden Cardiolife AED 2100, AED 2150K and AED 2151K (also listed for AED 2152K according to the accessory manufacturer’s catalogue) |
| Population | Adults and paediatric \u003E 10 kg (\u003E 12 months); contraindicated in children under 12 months / \u003C 10 kg |
| Functions | Defibrillation, synchronized cardioversion, transcutaneous pacing and ECG monitoring (max. 24 h) |
| Conductive gel | Low-impedance conductive adhesive gel |
| Backing material | Medical foam (FOAM) 1 mm thick; conductive metal foil |
| Connector | Anti-shock safety, accessible from outside the pouch (preconnected / lead-out) |
| Presentation | 1 pair in heat-sealed polyethylene and aluminium pouch (10 pairs per box) |
| Classification | Medical device class IIb (EU Regulation 2017/745 – MDR) |
| Materials | Latex and phthalate free; non-sterile |
| Shelf life | Approx. 30 months from the date of manufacture; expiry date indicated on the packaging |
| Storage | Dry place, +5 °C to +35 °C; operation 0 °C to +50 °C |
Note: the active surface area and number of shocks fields for the adult/paediatric series (≈ 95 cm² active in adult and 50 shocks) correspond to typical values for the EuroDefiPads family; confirm with the technical department for this reference before tendering.
4. Equivalences and Cross-References
This product is a consumable compatible with Nihon Kohden defibrillators, not an original spare part from the equipment manufacturer: the connector and the monitoring system are proprietary to Nihon Kohden, so replacing the original pad with a compatible accessory must always be validated against the equipment’s technical data sheet and IFU. The original equipment manufacturer references that this electrode functionally covers are:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Nihon Kohden | Disposable adult/paediatric pad for Cardiolife AED (original equipment) | P-740K | Comparable by category (not interchangeable 1:1) | Covers AED-2100K, 2150K, 2151K, 2152K, AED-3100 and TEC-5500/5600/7600/7700/8300 series (official IFU P-740K); 1.5 m cable. Factory original: connector, gel and labelling proprietary to Nihon Kohden |
| Nihon Kohden | Preconnected disposable adult/paediatric electrode (original equipment) | P-530K | Comparable by category (not interchangeable 1:1) | Original preconnected variant listed as an option for the AED-2100K and for the AED-3100; the electrodes remain connected to the AED in standby |
There is no equivalence table validated by Nihon Kohden that authorises replacing the P-740K/P-530K references with third-party consumables: any comparison must be understood as comparable at category level, not interchangeable 1:1, and must be confirmed with the technical department before its use in tendering. The correspondence of this SKU (F7989UW) with the official Nihon Kohden construction is not published by the equipment manufacturer.
5. Selection tips and variants of interest
If your AED is a Nihon Kohden Cardiolife AED 2100, 2150K or 2151K and you serve the general population (adults and children), this pair of preconnected adult/paediatric electrodes is the most versatile option: a single pair of pads covers both populations and the paediatric energy is selected with the device switch, without the need for an additional set. For AED fleets or centres with high usage volume, manage stock rotation by expiry (≈ 30 months) and always replace the electrode after use or expiry.
As interesting variants in the catalogue, consider the electrode compatible with the same devices that also covers the 2152K model (SKU F7948UW), the version for the Mediana and Tecno-Gaz brands (F7986UW) if you have other AEDs, or the radiotransparent variant for imaging procedures if your centre performs interventional procedures. Consult the Cahnos technical department to confirm exact compatibility with your defibrillator model before placing large-volume orders or tenders.