Description
1. Introduction
The graduated 5-channel anorectal manometry catheter (ref. C66U) is the technical solution that your gastroenterology department, digestive physiology unit or motility laboratory needs to accurately record anal motility and assess pelvic floor disorders. Its graduated design, multichannel configuration and 200 ml distension balloon allow complete manometric studies to be performed with stable, reproducible and clinically reliable pressure curves. A professional reference choice for functional anorectal diagnostics in adult patients.
2. Technical explanation of the equipment and procedure
During anorectal manometry, the catheter is inserted rectally and, once positioned, the four perfusion ports are connected to the transducers of the recording system. Each port is perfused with sterile water at a constant flow rate, allowing pressure curves to be obtained at specific points of the anorectal canal: the positions closest to the distal end record the pressure of the proximal rectal region, while the ports located further away assess the pressure in the anal sphincter area. The distension balloon, with a maximum capacity of 200 ml, is inflated in a controlled manner to study the response of the rectal wall and the reflexes related to defecation. The independent lumens prevent interference between the inflation and measurement channels, ensuring clinically reliable signals throughout the examination. The graduated scale allows the position and depth of insertion to be controlled in each test.
3. Technical Specifications
The main features of the C66U catheter are summarised in the following table:
| Characteristic | Specification |
| Configuration | 5 channels (4 perfusion ports for pressure recording + distension balloon channel) |
| Perfusion ports | 4 separate ports for pressure recording at specific distances |
| Port distances | 1.5 cm, 5 cm, 6 cm and 7 cm from the distal end |
| Distension balloon | Maximum injected volume of 200 ml for controlled distension |
| Gauge | 6.6 mm (20 CV) |
| Total length | 120 cm |
| Measurement system | Constant-flow sterile water perfusion |
| Graduation | Graduated scale for controlling insertion position and depth |
| Indication | Anorectal manometry, anal motility recording and diagnostics of pelvic floor disorders |
Functional distribution of ports and channels:
| Channel / Port | Function |
| Ports at 1.5 cm and 5 cm from the distal end | Pressure recording in the proximal rectal region |
| Ports at 6 cm and 7 cm from the distal end | Pressure recording in the anal sphincter area |
| Balloon channel | Inflation of the distension balloon (max. 200 ml) for controlled distension studies |
4. Equivalences and Cross-References
The C66U meets the standard of multichannel perfusion catheter with balloon for anorectal manometry used by the main manufacturers of digestive motility equipment. It has functional correspondence with the water-perfused balloon-type catheters from companies such as Laborie and Albyn Medical in 5-channel configurations for simultaneous recording of rectal pressure, sphincter pressure and distension, as well as with the perfused anorectal measurement catheters of the urodynamics series from the Rüsch / Teleflex group (colour-coded accessory connection, sterile and latex-free). Within the Cahnos catalogue, its closest equivalent is the graduated manometry and compliance catheter C56U, which adds radiopaque markers and rectal distensibility study, and the non-graduated version C74U of the same 5-channel configuration. For exclusive compliance studies, it is complemented with the bidirectional compliance rectal tube C55U.
5. Selection tips and variants of interest
Select the C66U when your protocol requires a graduated 5-channel catheter with a 200 ml distension balloon for standard recording of anal motility and sphincter pressure. If you also need to measure rectal distensibility with radiographic localisation of the measurement points, consider the graduated C56U catheter; if you do not need graduation or radiographic reference, the C74U offers the same functional configuration. For higher resolution in adult patients, the 7-channel C91AU catheter and the 4-channel Arhan C54U model are available; for the paediatric population, the graduated C69U catheter with a 100 ml balloon. For bubble-free filling of the balloons, complement the equipment with the C57U filling needle. Always verify the compatibility of connectors and transducers with your manometry system; at Cahnos we supply on request and with a formalised quotation under special conditions for B2B professionals.





Reviews