Description
The Sengstaken-Blakemore C120PVCS catheter is the reference device in emergency departments, gastroenterology and intensive care for the immediate control of upper gastrointestinal bleeding due to oesophageal varices. Designed by the French manufacturer Marquat Génie Biomédical, it combines a semi-rigid, graduated PVC body with two highly elastic natural latex balloons, achieving effective, rapid and reliable mechanical tamponade when first-line endoscopic treatment is unavailable or has failed. An essential resource on the emergency trolley of any digestive endoscopy unit.
Technical explanation and mechanism of action
The catheter incorporates three independent channels: a main gastric suction channel, fitted with three orifices, for evacuating gastric contents and lavage; and two secondary channels dedicated to the controlled inflation of the oesophageal and gastric balloons. Each inflation channel has a pilot valve with a female Luer connector, allowing the inflated volume to be checked and regulated. The gastric (or cardia) balloon is inflated first to anchor the catheter against the gastric fundus; the oesophageal balloon is then inflated, exerting direct pressure on the bleeding varices. The sliding fixation collar maintains the necessary traction and the radiopaque line on the body allows position to be verified by radiological control. The recommended working pressure is between 30 and 40 mmHg, with a maximum use time of 2 days and continuous patient monitoring.
Technical specifications by model
| Model | Diameter | Gauge | Length | Oesophageal balloon | Gastric balloon |
|---|---|---|---|---|---|
| C12020PVCS | 6.60 mm | 20 CH | 150 cm | 50 ml | 150 ml |
| C12018PVCS | 6.00 mm | 18 CH | 100 cm | 35 ml | 100 ml |
| C12016PVCS | 5.30 mm | 16 CH | 125 cm | 25 ml | 50 ml |
| C12014PVCS | 4.60 mm | 14 CH | 100 cm | 15 ml | 40 ml |
Balloon dimensions by reference
| Model | Oesophageal balloon (approx. length) | Gastric balloon (approx. length) |
|---|---|---|
| C12020PVCS | ≈ 140 mm | ≈ 68 mm |
| C12018PVCS | ≈ 110 mm | ≈ 54 mm |
| C12016PVCS | ≈ 98 mm | ≈ 45 mm |
| C12014PVCS | ≈ 74 mm | ≈ 43 mm |
- Graduated PVC body with radiopaque line along its entire length
- Two natural latex balloons (oesophageal and gastric) with independent inflation
- Gastric suction channel with three orifices
- Two pilot valves with female Luer connector for inflation control
- Sliding fixation collar
- Single-use, sterile by beta irradiation at 25 kGy; maximum use time: 2 days
- Presentation: box of 5 units
- Medical device class IIa with CE marking
Equivalences and cross-references
The C120PVCS reference corresponds to the three-channel Sengstaken-Blakemore catheter from the French manufacturer Marquat Génie Biomédical, available in four sizes: C12014PVCS, C12016PVCS, C12018PVCS and C12020PVCS. As a fully 100% pure silicone alternative, our catalogue includes the Type 32 range (refs. 800-320-0160, 800-320-0180 and 800-320-0210) and the four-lumen Type 42 models (refs. 002.182.0400, 002.182.0420 and 002.182.0440), recommended when maximum biocompatibility or absence of PVC is required. Clinically, the device is functionally equivalent to the Sengstaken-Blakemore and Minnesota type oesophagogastric tamponade catheters from other manufacturers on the market (BD, Bard), making it ideal as an approvable reference in hospital tenders and bids. In the same field of functional gastroenterology, the catalogue offers graduated oesophageal manometry catheters (refs. C58U, C59U and C76U) for the study of digestive motility.
Selection tips and variants of interest
Select the size according to the patient: C12014PVCS (14 CH) for infant patients, C12016PVCS (16 CH) for paediatric patients and C12018PVCS or C12020PVCS (18-20 CH) for adults. Before insertion, verify the integrity of both balloons and their filling system. Always lubricate with silicone gel or aqueous emulsion; never with petroleum jelly or greasy lubricants, as they deteriorate latex. Do not use the catheter in patients with known latex allergy: in such cases consider the silicone alternatives in the catalogue. During tamponade, control the oesophageal balloon pressure between 30 and 40 mmHg, secure the tube with the sliding collar and avoid excessive inflation to prevent mucosal injury. Remember that the maximum use time is 2 days and that the catheter is single-use, not re-sterilisable.





Reviews