Description
1. Introduction
The adult SpO₂ pulse oximeter sensor (Ref. 18631) is the reference consumable for services that monitor blood oxygenation with equipment compatible with Nellcor technology. Its 9-pin (DB9) connector ensures direct and secure integration with most bedside, transport and multiparameter monitors that use this standard.
Supplied in a box of 25 units, it is designed for hospitals, clinics, emergency departments and intensive care units that need to maintain a constant rotation of sensors with an adjusted cost per unit and immediate availability, minimising the risk of cross-contamination as it is a single-use product per patient.
2. Technical explanation
Pulse oximetry non-invasively measures the functional oxygen saturation of arterial haemoglobin (SpO₂) and pulse rate by spectrophotometry: the sensor emits red light (660 nm) and infrared light (905 nm) through the vascular bed and detects differential absorption according to the degree of blood oxygenation. The signal is transmitted to the monitor via the 9-pin (DB9) connector, which follows the electrical scheme of the Nellcor/Oxismart standard.
This sensor is designed for adult patients, with a size and fixation profile adapted to that population, allowing comfortable and stable placement during continuous monitoring. As it is supplied in a presentation of 25 units, it facilitates stock rotation in high-demand units and simplifies service logistics.
3. Technical Specifications
| Parameter | Specification |
| Sensor type | SpO₂ pulse oximeter sensor for adult patients |
| Compatibility | Nellcor systems (Nellcor/Oxismart protocol) |
| Connection | 9-pin (DB9) |
| Patient population | Adults (profile comparable to original reference >40 kg) |
| Applications | Monitoring of blood oxygen saturation (SpO₂) and pulse rate |
| Contents | Box of 25 units |
| Use | Hospitals, clinics, ICU, emergency and transport |
| SpO₂ measurement range | 35–100 % (reference value of the Nellcor DB9 family; confirm in batch IFU) |
| SpO₂ accuracy | ≤ 3 % in range 70–100 % (ISO 9919 / ISO 80601-2-61, reference value) |
| Pulse rate range | 30–250 ppm (reference value; confirm in batch IFU) |
| Wavelengths | Red light 660 nm / infrared 905 nm (reference value) |
| Material | Latex-free |
| Sterility | Non-sterile (single-use product per patient) |
Note: the range and accuracy values marked as reference come from technical data sheets of the same family of Nellcor sensors with DB9 connector; always confirm them on the labelling and IFU of the supplied batch.
4. Equivalences and Cross-References
SKU 18631 is Cahnos’ own management code. The following references have been cross-checked against authorised distribution catalogues, public product documentation and the GUDID/FDA regulatory database; none is interchangeable 1:1 with this sensor without prior verification, since the electrical scheme and communication protocol must be confirmed with the specific monitor installed at the facility.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Nellcor (Covidien/Medtronic) | Oxisensor II – adult disposable adhesive sensor | D-25 / D-25L | Comparable by category (not interchangeable 1:1) | Original sensor from the protocol manufacturer for adult patients (>30/40 kg depending on version), DB9 9-pin connector and Oxismart technology; coding and electronics specific to the Nellcor standard. Confirm protocol (Oxisensor/OxiMax) with the monitor. |
| Nellcor (Covidien/Medtronic) | Nellcor OxiMax – adult adhesive sensor | MAX-A | Comparable by category (not interchangeable 1:1) | Adult sensor from the original manufacturer with OxiMax memory chip and 9-pin D-Sub connector (purple, double key); requires a monitor compatible with OxiMax technology. Verify protocol compatibility with each monitor before replacing. |
There is no equivalence table validated by the manufacturer that cross-references this sensor with codes from other brands in an interchangeable 1:1 manner; consult the technical department before using in tenders. Attention: the ear lobe sensors of the Nellcor family are another type of product, with a different geometry and application point, and should not be confused with this adult sensor.
5. Selection tips and variants of interest
Choice by population: within the same family of 9-pin Nellcor sensors, you have the paediatric sensor (SKU 18632), infant (SKU 18633) and adult/newborn (SKU 18634). If your service works with reusable sensors, consider the adult forceps compatible with Nellcor technology (Ref. AB78315040091, 1.9 m cable).
Compatibility verification: although the connector is 9-pin (DB9), the Nellcor protocol has variants (Oxisensor/OxiMax). Always indicate the brand and model of the monitor to confirm exact compatibility before placing the order.
Hygiene and storage: as it is a single-use sensor per patient, it eliminates the risk of cross-contamination. Store in a dry place, protected from direct sunlight and extreme temperatures.
Cost-effectiveness: the presentation in a box of 25 units reduces the cost per sensor and ensures continuous stock in high-turnover units. Consult our technical-commercial team for special B2B conditions by volume and supply contracts.




Reviews
There are no reviews yet.