Description
1. Commercial Introduction
The absorbable multifilament suture PREMISUT RAPID of polyglactin 910 (PGLAR) in Prime quality is the reference for accelerated absorption surgical closure for general and ambulatory surgery operating theatres, emergency services, ophthalmology clinics and B2B purchasing centres that need a braided, flexible thread with early and totally predictable loss of strength. Its ⅜ circle 30 mm needle with triangular body and reverse cutting point penetrates firm and non-compliant tissues with a clean and controlled passage, without tearing the skin edge. Presentation in a box of 12 sterile units with batch traceability and factory reference P2L23DS3010.
2. Technical Explanation
Polyglactin 910 (PGLAR) is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, spun into a braided multifilament structure with a grooved finish that favours thread sliding, very smooth handling and a firm, secure knot. In the RAPID variant (low molecular weight), absorption occurs by hydrolysis —a predictable process independent of the patient’s enzymatic activity— within 40 to 45 days: it retains 50% of the initial tensile strength at 5 days and loses it completely at 14 days. This profile makes it especially suitable for superficial tissues with rapid healing (skin closure, mucous membranes, subcutaneous tissue), ligation of small-calibre vessels and conjunctival suture, in many cases without the need to remove the stitches. The ⅜ circle needle (135°) with triangular body and reverse cutting point directs the cutting edge away from the wound, creating a clean entry orifice that preserves the skin edge. Product sterile by gamma radiation, single-use and non-reesterilizable: check the integrity of the packaging and the expiry date before each procedure.
3. Technical Specifications
3.1. General characteristics (Ref. P2L23DS3010 · SKU AMP2L23DS3010)
| Parameter | Specification |
|---|---|
| Material | Polyglactin 910 (PGLAR): copolymer 90% glycolide / 10% L-lactide, RAPID low molecular weight variant |
| Thread structure | Braided multifilament with grooved finish |
| Quality | Prime (PREMISUT RAPID range) |
| Gauge | USP 0 (metric 3.5; approximate diameter 0.35 – 0.399 mm according to Pharmacopoeia) |
| Thread colour | Colourless (undyed) |
| Thread length | 90 cm |
| Needle shape | ⅜ circle (3/8, 135°) |
| Needle body and point | Triangular reverse cutting (reverse cutting), size DS30 (30 mm) |
| Absorption mechanism | Hydrolysis |
| Absorption period | 40 – 45 days |
| Tensile strength retention | 50 % at 5 days / 0 % at 14 days |
| Sterilisation | Gamma radiation; single-use, not re-sterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
| Factory reference | P2L23DS3010 |
| SKU Cahnos | AMP2L23DS3010 |
3.2. Tensile strength retention profile
| Post-implantation time | Retained tensile strength |
|---|---|
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% (total loss of tensile support) |
| Complete absorption | 40 – 45 days by hydrolysis (predictable and uniform rate) |
Profile corresponding to the RAPID formulation; it must not be confused with standard polyglactin 910 (PGLA line, absorption 56 – 72 days).
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
|---|---|---|
| 0 | 3.5 | 0.35 – 0.399 mm (indicative, according to Pharmacopoeia) |
4. Equivalences and Cross-References
Rapid-absorption polyglactin 910 (low molecular weight) is one of the most widespread categories of fast-acting braided multifilament absorbable sutures in the surgical market. The international clinical reference is the Coated VICRYL RAPIDE™ (Ethicon / Johnson & Johnson MedTech) line, which shares composition (90% glycolide / 10% L-lactide), braided structure, colourless thread and the same accelerated absorption profile (loss of resistance at 14 days and absorption around 42 days). Only codes whose attributes have been verified in catalogues of authorised distributors and in the official Ethicon EMEA datasheet are listed below; no reference should be considered interchangeable 1:1 without validation by the technical department.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| Ethicon (J&J MedTech) | VICRYL RAPIDE (braided coated polyglactin 910, colourless) | W9963 – USP 0, 90 cm, colourless, V-34 needle 36 mm in ½ circle, tapercut point | Comparable by category (not interchangeable 1:1) | Same material, rapid absorption profile, USP 0 gauge, 90 cm colourless thread and box of 12 units; the needle geometry differs (½ circle tapercut 36 mm versus ⅜ circle reverse cutting 30 mm). Code confirmed in Ethicon EMEA and authorised distributors. |
| Ethicon (J&J MedTech) | VICRYL RAPIDE (braided coated polyglactin 910, colourless) | W9947 – USP 0, 90 cm, undyed, needle V-39 (non-cutting) | Comparable by category (not interchangeable 1:1) | Same material, USP 0 gauge, 90 cm colourless thread and accelerated absorption; the needle type differs (non-cutting body V-39 versus triangular reverse cutting point DS30 of 30 mm). Code confirmed in authorised distributors. |
There is no equivalence table validated by the manufacturer that allows declaring 1:1 interchangeability for the exact configuration USP 0 / ⅜ circle reverse cutting needle of 30 mm / 90 cm thread; consult the technical department before using in tenders. As homologous ranges in the rapid absorption segment, Safil Quick™ and Novosyn Quick™ (B. Braun), DemeCRYL™ Fast (DemeTECH), Surgicryl Fast Absorb (SMI) and AssuCryl (Assut Medical) are listed, comparable by category by material and absorption profile; their specific codes for this exact combination are pending confirmation with each manufacturer. Standard VICRYL (absorption 56 – 72 days) is not comparable to this RAPID version due to its longer resistance retention profile.
5. Selection tips and variants of interest
USP 0 (metric 3.5) gauge is one of the most versatile: indicated for approximation of subcutaneous tissues, superficial fascia, skin and mucosal closure, ligation of small and medium-calibre vessels and ophthalmic surgery, with good initial strength and rapid reabsorption that avoids removal in fast-healing tissues. The ⅜ circle reverse cutting needle of 30 mm is especially recommended in reduced working fields and firm tissues, where a clean passage without tearing the edge is required. Always verify three parameters in the surgical report: USP gauge, needle geometry and size, and thread length. Within the same PGLAR Poliglactina 910 RAPID range with reverse cutting needle, Cahnos offers complementary configurations: the same DS30 needle in Normal quality (SKU AM2L23DS3010) and in Prime double-thread format 2 × 90 cm (SKU AMP2L23DS3010.2), the DS30 needle in USP 2/0 gauge (AM2L22DS3020 and Prime AMP2L22DS3020), the DS36 needle in USP 2/0 and USP 0 (AM2L22DS3620 / AM2L22DS3610) and the ½ circle reverse cutting HS40 needle in USP 0 (AMP2L23HS4010) for deeper tissues. If your protocol prefers a conical point (taperpoint), consult the HR family (e.g. AMP2L22HR3010); and if you require the same absorption profile in rapid polyglycolic acid, consider the PGAR RAPID family. The colourless version offers maximum discretion on the surgical field; if you require greater visibility, request the same configuration in dyed thread on demand. Sterile single-use product: store it in a cool, dry place, protected from sunlight, and check the integrity of the packaging and the expiry date before each procedure. For special B2B conditions, volume orders or equivalents from other commercial houses, contact the technical-commercial team of Cahnos.





Reviews
There are no reviews yet.