1. Commercial Introduction
The Absorbable Multifilament Suture PGLAR Polyglactin 910 RAPID Prime (range PREMISUT RAPID) is the reference for accelerated-absorption closure for operating theatres, day hospitals, ambulatory minor surgery services, ophthalmology clinics and B2B purchasing centres that require a braided, flexible suture with fully predictable early strength loss. Its ⅜ circle needle with triangular reverse-cutting body and point of 16 mm ensures clean, controlled penetration in skin and firm, thin tissues, while the colourless fluted 75 cm suture offers maximum discretion on the surgical field. Practical and cost-effective presentation in a box of 12 sterile units, with batch traceability and a minimum order of 30 boxes for the B2B channel.
2. Technical Explanation
Polyglactin 910 (PGLAR) is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, spun into a coated braided multifilament structure with a fluted finish that favours suture glide, smooth handling and a firm, secure knot. In the RAPID variant (low molecular weight), the suture is formulated to lose tensile strength in an accelerated and programmed manner: it retains 50% of initial strength at 5 days and 0% at 14 days, with complete absorption by hydrolysis between 40 and 45 days, a predictable rate independent of the patient’s enzymatic activity. The reverse-cutting needle (reverse cutting) with triangular body and point directs the cutting edge away from the wound, creating a clean entry hole that preserves the skin edge and prevents the cut from widening in the puncture bed. The suture is coated with a mixture of polyglactin 370 copolymer (30% glycolide / 70% lactide) and calcium stearate (≤ 5%), which improves lubrication and smooth passage. It is non-antigenic and non-pyrogenic and meets the requirements of the European Pharmacopoeia (E.P.) and USP for sterile absorbable synthetic braided sutures. Single-use product, sterilised by gamma radiation, not re-sterilisable.
3. Technical Specifications
General characteristics (Ref. P2L22DS1650 · SKU AMP2L22DS1650)
| Parameter | Specification |
| Thread material | Polyglactin 910 (PGLAR), copolymer 90 % glycolide / 10 % L-lactide, RAPID variant low molecular weight |
| Thread structure | Braided coated multifilament, fluted finish |
| Quality | Prime (PREMISUT RAPID range) |
| Gauge | USP 5/0 (metric 1) |
| Thread diameter | 0.10 – 0.149 mm (approximate, according to Pharmacopoeia) |
| Thread colour | Colourless (undyed) |
| Thread length | 75 cm (mounted 1 needle / 1 thread) |
| Needle shape | ⅜ circle (3/8, 135°) |
| Needle body and point | Triangular, reverse cutting |
| Needle size | DS16 (16 mm), Prime quality |
| Absorption mechanism | Hydrolysis |
| Absorption period | 40 – 45 days |
| Sterilisation | Gamma radiation; single-use, not re-sterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Approximation of soft tissues, ligatures, closure of skin and mucous membranes, paediatric surgery, episiotomies, circumcision, oral mucosa and conjunctival suturing in ophthalmic surgery (short-term wound support, up to 7–10 days) |
| Presentation | Box of 12 units (minimum order 30 boxes) |
| Factory reference | P2L22DS1650 · SKU Cahnos AMP2L22DS1650 |
Tensile strength retention and absorption profile
| Post-implantation time | Retained tensile strength |
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% (total loss of tensile support) |
| Complete absorption | 40 – 45 days by hydrolysis (predictable and uniform rate) |
Gauge equivalence (USP)
| USP size | Metric size | Approximate diameter |
| 5/0 (this product) | 1 | 0.10 – 0.149 mm |
| 4/0 | 1.5 | 0.15 – 0.199 mm |
| 3/0 | 2 | 0.20 – 0.249 mm |
4. Equivalences and Cross-References
Rapid-absorption polyglactin 910 (low molecular weight) is the most widespread category of fast-acting braided multifilament absorbable in the surgical market, whose international clinical benchmark is the VICRYL RAPIDE line (Ethicon / Johnson & Johnson MedTech), with which it shares composition (90% glycolide / 10% L-lactide), hydrolysis absorption mechanism and accelerated strength-loss profile (≈50% at 5 days, 0% at 14 days). Only codes whose complete attributes have been verified in authorised distribution catalogues and verifiable data sheets are included; none of these references should be considered interchangeable 1:1 without validation by the technical department, since the exact needle nomenclature (FS-3 vs DS16) and the braid coating differ.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (J&J MedTech) | VICRYL RAPIDE (polyglactin 910, braided, colourless) | W9915 – USP 5/0, 16 mm needle in ⅜ circle, reverse cutting Prime, colourless thread 75 cm, box of 12 | Comparable by category (not interchangeable 1:1) | Material, rapid absorption (≈42 days), USP gauge 5/0, needle size (16 mm), ⅜ circle curvature, reverse-cutting point, colourless 75 cm suture and presentation in a box of 12 units all coincide. The needle profile nomenclature (FS-3 Prime vs DS16) and the coating differ. Code confirmed in authorised distribution. |
| Ethicon (J&J MedTech) | VICRYL RAPIDE (polyglactin 910, braided, colourless) | VR2296 – USP 5/0, FS-3 needle 16 mm in ⅜ circle, reverse cutting, colourless thread 75 cm | Comparable by category (not interchangeable 1:1) | Same material, USP gauge 5/0, FS-3 needle of 16 mm in ⅜ circle with reverse cutting and colourless 75 cm suture. The needle profile nomenclature (FS-3 vs DS16) and the commercial house differ. Code confirmed in multiple authorised distribution sources. |
Also included as homologous ranges in the rapid-absorption segment are Safil Quick / Novosyn Quick (B. Braun), DemeCRYL Fast (DemeTECH) and Surgicryl Fast Absorb (SMI), comparable by category by material and absorption profile; for the exact combination USP 5/0 / DS16 reverse-cutting needle / 75 cm suture, their specific codes must be confirmed with each manufacturer. Standard VICRYL (absorption 56–72 days) is not comparable to this RAPID version. There is no manufacturer-validated equivalence table cross-referencing this specific reference with codes from other brands; consult the technical department before use in tenders, verifying on each official data sheet the USP gauge 5/0 (metric 1), the ⅜ circle curvature, the needle type and size (triangular reverse-cutting 16 mm), the suture length (75 cm) and the colourless colour.
5. Selection tips and variants of interest
The USP 5/0 (metric 1) gauge is ideal for fine and delicate tissues: thin skin, mucous membranes, conjunctiva, ophthalmic surgery, paediatric surgery and superficial approximations where maximum fineness is required with short-term support and early absorption, in many cases without suture removal. The ⅜ circle reverse cutting needle of 16 mm is especially recommended when a firm and clean passage through skin or mucous membranes is required without tearing the edge, typical in microsurgery, dermatology, minor surgery and ophthalmology. Within the same PGLAR Poliglactina 910 RAPID Prime family with DS16 needle, Cahnos offers complementary gauges: USP 4/0 (ref. P2L22DS1640) and USP 3/0 (ref. P2L22DS1630). For a longer needle with the same 5/0 gauge, the DS19 variant (ref. P2L22DS1950) is available, and for smaller access points the DS12 (ref. P2L22DS1250). If the protocol prefers a precision reverse cutting point with a square body, consider the XDS range; if a conical point is preferred for looser tissues, the HR family. The colourless version offers maximum aesthetic discretion; if greater visibility is required, request the same configuration in dyed thread on demand. If your protocol requires longer tensile support (56–72 days), consider the standard PGLAR Poliglactina 910 range (refs. P1L22DS1650 / 1L22DS1650). As this is a sterile single-use product, check the integrity of the packaging and the expiry date before each procedure and store it in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or equivalents from other commercial brands, contact the technical-commercial team at Cahnos.