Description
1. Rapid absorbable precision suture for soft tissues
The multifilament absorbable suture PGLAR Poliglactina 910 RAPID Prime is the technical choice of the professional who requires a secure, manageable and accelerated absorption closure in soft tissue approximation procedures, ligatures and ophthalmic surgery. Its ½ circle needle with cylindrical body and taperpoint conical point (HR22) penetrates tissue with minimal resistance and less trauma, while the colourless fluted braided thread offers firm knotting and smooth sliding in each stitch. Presented in a box of 12 sterile units with double thread of 2×75 cm, it is a cost-effective and reliable alternative for operating theatres, purchasing centres and B2B tenders.
2. Technical Explanation
Polyglactin 910 is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, the same polymeric principle used by rapid absorption lines on the market. Its reabsorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the suture retains 50% of its tensile strength at 5 days and loses practically all its strength at 14 days, with absorption completed between days 40 and 45. This short support window makes it ideal for tissues that heal quickly and where it is desired to minimise residual foreign material. The Prime quality guarantees a homogeneous braided finish, constant needle passage and comfortable knot seating, with minimal tissue reaction. The double thread format (2×75 cm) increases performance per unit in the operating theatre.
3. Technical Specifications
| Thread material | Polyglactin 910 (PGLAR), copolymer 90 % glycolide / 10 % L-lactide |
| Structure | Braided multifilament, fluted (coated), colourless |
| Gauge | USP 3/0 |
| Thread length | 2 × 75 cm (double thread) |
| Needle shape | ½ circle, cylindrical body |
| Needle tip | Conical taperpoint (Taper Point) |
| Needle size | 22 mm (HR22) |
| Absorption | By hydrolysis, complete between 40 and 45 days |
| Residual tensile strength | 50% at 5 days; 0% at 14 days |
| Colour | Colourless |
| Quality | Prime |
| Sterilisation | Gamma radiation |
| Tissue reaction | Minimal |
| Indications | Approximation of soft tissues, ligatures and ophthalmic surgery |
| Manufacturer reference | P2L22HR2230.2 (SKU AMP2L22HR2230.2) |
| Presentation | Box of 12 sterile units |
The strength retention profile cited corresponds to the rapid absorption formulation (RAPID line); it should not be confused with standard absorption polyglactin 910 (PGLA line, 56–72 days), also available in our catalogue.
4. Equivalences and Cross-References
At category level, this rapid absorption polyglactin 910 suture corresponds to the accelerated absorption braided synthetic absorbable sutures on the market. The closest homologous family is VICRYL RAPIDE™ from Ethicon, with the same polymeric composition (90% glycolide / 10% L-lactide) and absorption profile (40–45 days, total loss of strength at 14 days). The following reference has been verified against the official Ethicon catalogue (EMEA code W9974) and regulatory databases:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (J&J MedTech) | VICRYL RAPIDE, USP 3/0, colourless, 75 cm, SH-1 ½ circle 22 mm taper point needle | W9974 | Comparable by category (not interchangeable 1:1) | Same composition (rapid absorption polyglactin 910), gauge 3/0, colourless thread of 75 cm and ½ circle conical point needle of 22 mm; differs in the configuration (1 thread × 75 cm versus the double thread 2×75 cm of our reference) and the coating/finish of the needle. Confirmed in official Ethicon EMEA catalogue and in GUDID. |
Other homologous lines of rapid absorption polyglactin 910 that can serve as a functional reference, although without a 1:1 equivalence published by the manufacturer, are the families of rapid braided absorbable sutures from B. Braun (Safil Quick / Novosyn Quick range). There is no equivalence table validated by the manufacturer for the exact configuration USP 3/0 + ½ circle conical point needle 22 mm + colourless double thread 2×75 cm; consult with the technical department before using in a tender.
5. Selection tips and variants of interest
To choose the correct suture, verify three parameters: the USP gauge (3/0 is ideal for soft tissues of fine-medium thickness and short-term support; gauges 4/0 to 6/0 for mucous membranes and delicate areas, and 2/0 to 1 for tissues requiring greater strength), the needle geometry (the ½ circle with taperpoint conical point is most suitable for soft tissues and confined spaces; reverse cutting triangular needles are reserved for skin or dense structures) and the thread length (the double 2×75 cm format maximises performance per unit). In our PGLAR Poliglactina 910 RAPID Prime range you will find variants with HR17 needle (USP 5/0), HR22 and HR26 (USP 2/0), HR30 (USP 0, double thread 2×75 cm), HR36 and HR48 (USP 0/1), as well as 3/8 circle formats with reverse cutting (DS16, DS19) for dermatological and cosmetic surgery. For hospital replenishment orders, consult volume conditions and stock availability with our commercial department.