Description
PGLAR Polyglactin 910 RAPID Prime suture, HR40 needle
The Absorbable Multifilament Suture PGLAR Polyglactin 910 RAPID Prime is the reference choice for professionals who need a synthetic absorbable thread with predictable behaviour, excellent handling and rapid strength loss, designed for fast-healing tissues. Its presentation with an attached needle reduces handling, optimises operating theatre times and guarantees batch traceability. Available in a box of 12 units, with manufacturer reference P2L23HR4001.
Technical product explanation
Polyglactin 910 is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, structured as a grooved multifilament braid that provides flexibility, smooth glide and great knot security. Its absorption occurs by hydrolysis, without depending on tissue enzymes, ensuring uniform kinetics among patients. The RAPID variant is optimised for short-term support: it maintains 50% of tensile strength at 5 days and loses 100% at 14 days, being completely eliminated in approximately 42–56 days. The ½ circle needle with cylindrical body and grooved conical point (taperpoint) of 40 mm penetrates soft tissues with minimal injury, ideal for tissue approximation, ligatures and ophthalmic surgery.
Technical specifications
| Material | Polyglactin 910 (PGLAR), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Braided multifilament, grooved, colourless |
| Gauge | USP 1 |
| Thread length | 90 cm |
| Needle | ½ circle, cylindrical body, taperpoint conical point, 40 mm (HR40) |
| Absorption | By hydrolysis, between 40 and 45 days |
| Tensile strength | 50% at 5 days; 0% at 14 days |
| Colour | Colourless |
| Quality | Prime |
| Sterilisation | Gamma radiation |
| Presentation | Box of 12 units |
| Reference | P2L23HR4001 (SKU AMP2L23HR4001) |
Equivalences and cross-references
At category level, polyglactin 910 RAPID corresponds to the braided synthetic absorbable sutures with rapid absorption on the market. The closest family is VICRYL RAPIDE™ (low molecular weight polyglactin 910, same composition 90% glycolide / 10% L-lactide and same absorption profile 40–45 days with strength loss at 14 days).
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence Level | Technical Differences / Remarks |
| Ethicon / J&J MedTech | VICRYL RAPIDE™ (polyglactin 910) | V9590H | Comparable by category (not interchangeable 1:1) | Same material, USP gauge 1, ½ circle needle conical point 40 mm, colourless and 90 cm. Different manufacturer and coating (VICRYL RAPIDE is coated, in this reference coating has not been verified); check IFU before substituting. |
| Ethicon / J&J MedTech | Coated VICRYL™ (polyglactin 910, standard absorption) | J959H | Comparable by category (not interchangeable 1:1) | Matches in USP gauge 1, ½ circle needle conical point 40 mm, colourless braided thread of 90 cm, but its absorption profile is standard VICRYL (longer strength retention), not RAPID. |
There is no equivalence table validated by the manufacturer; consult with the technical department before using in tenders. Any other brand of rapid absorption polyglactin 910 should be checked against its technical data sheet and IFU before proposing substitution.
Selection tips and variants of interest
To choose the correct suture, verify three parameters: USP gauge (this model, USP 1, is intended for medium thickness soft tissues and approximations that require short-term support), needle geometry (½ circle with taperpoint conical point for minimal trauma in soft tissue; the reverse cutting variant is reserved for denser tissues) and thread length (90 cm for wide suture manoeuvres). In the PGLAR Polyglactin 910 RAPID family, Prime and standard variants are available with HR40 needle in USP gauges 2, 1, 0, 2/0 and presentations of 75 cm, as well as double thread 2×90 cm for greater yield per unit. If the priority is cost per stitch without changing the configuration, consider the standard quality version of the same reference.





Reviews
There are no reviews yet.