Description
1. Rapid absorbable suture for precision surgery
The absorbable multifilament suture PGLAR of polyglactin 910 rapid absorption (line PREMISUT RAPID Prime) is the technical choice for professionals who need a short-term suture support, excellent handling and totally predictable reabsorption. Its ½ circle needle with cylindrical body and atraumatic Ultra Plus taper point of 36 mm penetrates soft tissues with minimal resistance and less trauma, a quality especially valued in ophthalmic surgery, ligatures, approximation of superficial tissues and general surgery. Presented in a box of 12 sterile units, it is a cost-effective and reliable alternative for operating theatres and B2B purchasing centres.
2. Technical Explanation
Polyglactin 910 is a synthetic absorbable copolymer composed of 90 % glycolide and 10 % L-lactide, which is reabsorbed by hydrolysis, i.e., by chemical degradation in contact with tissue fluids, without dependence on enzymes. The rapid absorption variant presents an accelerated and programmed loss of tensile strength kinetics: it retains 50 % of its tensile strength at 5 days and loses practically all its strength at 14 days, with complete absorption between days 40 and 45. This short support window makes it ideal for rapidly healing tissues and where it is desired to minimise residual foreign material. The multifilament braided grooved format, soft and colourless, facilitates sliding, secure knotting and comfortable knot seating, with minimal tissue reaction.
3. Technical Specifications
| Thread material | Polyglactin 910 (copolymer 90% glycolide / 10% L-lactide), synthetic, absorbable |
| Structure | Braided grooved multifilament, undyed |
| Gauge | USP 0 |
| Thread length | 90 cm |
| Colour | Colourless |
| Needle shape | ½ circle, cylindrical body |
| Tip type | Atraumatic Ultra Plus taper point |
| Needle size | 36 mm (HRP36) |
| Absorption | By hydrolysis, complete between 40 and 45 days |
| Residual tensile strength | 50% at 5 days; 0% at 14 days |
| Sterilisation | Gamma radiation |
| Tissue reaction | Minimal |
| Indications | Approximation of soft tissues, ligatures and ophthalmic surgery |
| Manufacturer reference | P2L23HRP3610 |
| Presentation | Box of 12 sterile units |
The retention strength profile cited corresponds to the rapid absorption formulation (RAPID line); it should not be confused with standard absorption polyglactin 910 (PGLA line, 56-72 days), also available in our catalogue.
4. Equivalences and Cross-References
The technical profile of this suture (polyglactin 910 rapid absorption, USP 0 gauge, colourless thread 90 cm and ½ circle needle 36 mm with taper point) coincides with a commercial presentation of reference in the European market, verified against official manufacturer catalogue and authorised distributor sources:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (J&J MedTech) | VICRYL RAPIDE, needle V-34 ½ circle 36 mm, USP 0, 90 cm, colourless | W9963 | Comparable by category (not interchangeable 1:1) | Same composition (polyglactin 910 rapid absorption), USP 0 gauge, colourless thread 90 cm and ½ circle needle 36 mm; the Ethicon tapercut needle versus the atraumatic Ultra Plus taper point of our line and the coating may differ slightly. Confirmed in official Ethicon catalogue and in GUDID. |
Other homologous lines of polyglactin 910 rapid absorption that can serve as functional reference for the same indication, although without 1:1 equivalence published by the manufacturer, are the families of rapid absorbable braided sutures from B. Braun (Safil Quick/Novosyn Quick range). There is no equivalence table validated by the manufacturer; consult the technical department before using in a tender.
5. Selection tips and variants of interest
Before ordering, verify three parameters: the USP gauge (USP 0 is reserved for tissues requiring higher initial strength; gauges 2/0 to 6/0 for finer areas or mucous membranes), the needle geometry (the ½ circle with atraumatic taper point is the most suitable for soft tissues and deep spaces; triangular reverse cutting needles are reserved for skin or dense structures) and the thread length (90 cm for general surgery and material replacement). In our PGLAR range of rapid polyglactin 910 you will find variants of ½ circle cylindrical needle of 26 mm (HR26), 36 mm (HR36) and 48 mm (HR48), as well as 3/8 circle formats with reverse cutting in sizes DS16, DS19 and DS24 and gauges from 5/0 to 2/0. For hospital volumes, consult B2B conditions per box and aggregated orders with our commercial department.





Reviews
There are no reviews yet.