Description
1. Product Presentation
The absorbable multifilament polyglactin 910 (PGLA) suture from the Premisut line is the trusted choice for professionals seeking predictable performance, smooth handling and absolute knot security. Reference 1L18JR3020 combines a J-hook needle with cylindrical body and conical (taperpoint) point of 30 mm with a violet braided thread of USP gauge 2/0 and 70 cm length, ideal for tissue approximation, ligatures and ophthalmic surgery.
It is presented in a box of 12 units with a minimum order of 30 boxes, designed for hospitals, clinics and purchasing centres that need a stable supply, with full traceability and guaranteed sterility in each unit.
2. Technical explanation
Polyglactin 910 is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, presented as a braided multifilament coated with a grooved finish that facilitates thread sliding and proper knot seating. It is absorbed by hydrolysis over a period of 56 to 72 days, with a predictable absorption rate and minimal tissue reaction.
The tensile strength profile ensures the necessary support during the critical healing phase: 75% at 2 weeks, 50% at 3 weeks and 30% at 4 weeks. The J-hook needle, with cylindrical body and conical point, penetrates soft tissues without cutting effect, minimising trauma and being especially suitable for delicate tissues and ophthalmic surgery. The suture is sterilised by ethylene oxide and maintains excellent tensile and knot tying strength.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Material | Polyglactin 910 (PGLA): copolymer 90% glycolide / 10% L-lactide |
| Structure | Braided coated multifilament, fluted finish |
| Thread gauge | USP 2/0 |
| Colour | Violet (also available colourless) |
| Thread length | 70 cm |
| Needle shape | Hook / J-Hook (J shape) |
| Needle body and tip | Cylindrical body with conical point (taperpoint) |
| Needle size | JR 30 mm |
| Absorption | By hydrolysis, between 56 and 72 days |
| Tensile strength retention | 75 % (2 weeks), 50 % (3 weeks), 30 % (4 weeks) |
| Sterilisation | Ethylene oxide (sterile) |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
| Reference | 1L18JR3020 |
4. Equivalences and Cross-References
Polyglactin 910 is the classic synthetic absorbable suture material of the Vicryl type. This Cahnos reference is functionally equivalent to the coated polyglactin 910 sutures of the main international manufacturers, such as Ethicon’s Vicryl (J needle with conical point), as well as the polyglactin 910 lines from Assut Sutures, SMI, Kono and G21, all with a hydrolysis absorption profile of 56 to 72 days and similar tensile retention.
In the Cahnos catalogue itself, the same configuration with Prime finish corresponds to reference P1L18JR3020. For alternative gauges and needles from the same PGLA family, see references 1L18JR3001 (USP 1 with J-hook needle), 1L12HR3020 (1/2 circle needle) and 1L18FR3620 (5/8 circle needle).
5. Selection tips and variants of interest
Select USP gauge 2/0 when you need a fine thread with good strength in soft or friable tissues, especially in ophthalmic surgery and approximation of delicate tissues. If the procedure requires greater tensile strength, consider the USP 1 version (1L18JR3001). The 30 mm J-hook needle with conical point is especially recommended where the cutting effect is to be avoided.
For surgical fields with deeper access or greater curves, Cahnos offers the same polyglactin 910 with 1/2 circle (HR) and 5/8 circle (FR) needles. All references are supplied in boxes of 12 units with a minimum order of 30 boxes, guaranteeing traceability and sterility by ethylene oxide.





Reviews