Description
The absorbable multifilament poliglactin 910 (PGLA) suture, reference 1L12HR4020 from the Premisut range, is the reference surgical closure solution for hospitals, ophthalmic clinics, private operating theatres and purchasing centres seeking a flexible, resistant synthetic thread with fully predictable absorption. Its ½ circle needle with cylindrical body and 40 mm Taperpoint conical point penetrates soft tissues by separating the fibres without cutting them, which minimises trauma and makes it especially suitable for tissue approximation, ligatures and ophthalmic surgery. It is a synthetic absorbable medical device, sterile and single-use, presented in a box of 12 units with a minimum order of 30 boxes, which guarantees continuous and traceable supply for your healthcare centre.
2. Technical explanation of the product and procedure
Poliglactin 910 is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, presented in a braided multifilament structure with a grooved finish. This configuration provides a regular surface that facilitates the passage of the thread through tissue, very smooth handling, comfortable sliding and firm, secure knot tying, with high initial tensile strength and a low coefficient of friction.
Absorption occurs by hydrolysis, i.e. by degradation of the polymer in contact with organic fluids, over a predictable period of 56 to 72 days, independent of the patient’s enzymatic activity. The thread maintains wound support during the critical 2 to 4 weeks of healing, with minimal tissue reaction. The ½ circle needle with cylindrical body and conical point (Taperpoint) penetrates by separating the fibres without cutting them, minimising trauma to delicate structures, while the violet colour of the thread ensures optimal visibility in the surgical field. Sterilisation by ethylene oxide; sterile single-use product, not suitable for re-sterilisation.
3. Technical Specifications
3.1. Technical data sheet
| Characteristic | Specification |
| Reference (SKU) | 1L12HR4020 |
| Range / Quality | Premisut, Normal quality |
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Braided multifilament with grooved finish |
| Colour | Violet (also available in undyed) |
| Gauge | USP 2/0 (metric 3) |
| Thread length | 75 cm |
| Needle shape | ½ circle |
| Needle body | Cylindrical |
| Needle tip | Conical (Taperpoint) |
| Needle size | 40 mm (HR 40) |
| Absorption mechanism | Hydrolysis |
| Absorption period | 56-72 days |
| Sterilisation | Ethylene oxide (EO); sterile, single use |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
3.2. Tensile strength retention profile
| Time after implantation | Retained tensile strength |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 30 % |
| Complete absorption | 56-72 days, by hydrolysis |
Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a fully predictable absorption rate.
3.3. Presentation and ordering conditions
| Value | Value |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
| Format | Sterile, single-use |
4. Equivalences and Cross-References
Poliglactin 910 is one of the most widely used synthetic absorbable materials in the European market. This reference is clinically comparable, in terms of composition, absorption profile and tension retention, to the braided multifilament sutures of the main manufacturers:
| Manufacturer / Company | Equivalent line |
| Ethicon (Johnson & Johnson) | Vicryl / Vicryl Plus (braided polyglactin 910, violet) |
| B. Braun | Safil (braided polyglactin 910, violet) |
| Medtronic / Covidien | Polysorb (absorbable braided range of polyglactin 910) |
| SMI | Surgicryl (braided polyglactin 910, violet) |
| Assut | Assucryl 910 (polyglactin 910) |
| Laboratory Aragó | Polyglactin 910 range with 1/2 circle needles (HR series) |
Within the Cahnos catalogue, this reference is directly related to: P1L12HR4020 (Prime quality, same HR 40 mm needle, USP 2/0 and 75 cm), AM112HR4020 (alternative in polyglycolic acid with the same needle configuration), 1L18HR4020 (same HR 40 mm needle with 70 cm thread), 1L12HR4520 and 1L12HR4820 (same gauge with HR 45 and HR 48 mm needles), as well as the Heavy versions 1L12HR4001H and 1L12HR4002H (USP sizes 1 and 2). The indicated equivalences are for guidance; for a formal substitution of references, please consult our technical-commercial department.
5. Selection tips and variants of interest
USP gauge 2/0 (metric 3) is very versatile: suitable for subcutaneous tissue, superficial muscle planes, soft tissue approximation and ophthalmic surgery procedures, with an excellent balance between initial strength and predictable absorption during a 2 to 4 week healing period. For finer tissues, gauges 3/0 or smaller are recommended, while gauges 1 and 2 are reserved for areas subject to greater tension (Heavy versions 1L12HR4001H and 1L12HR4002H).
In ophthalmic surgery, prioritise ½ circle needles: the conical point (Taperpoint) separates the fibres without cutting them, ideal for soft and delicate tissues; the blunt point offers maximum atraumaticity and the trocar or reverse cutting triangular point (HS) is reserved for fibrous and dense structures. Check the needle radius (HR 20, 30, 40, 45 or 48 mm) according to the depth and surgical area to be treated. If you require a longer thread, consult the 90 cm versions (P1L13HR4810) or the 150 cm loop format (P1L15HR48L10). For longer force retention needs (180-210 days), the polydioxanone (PDO) family is the recommended alternative.
As it is a sterile single-use product, always check the integrity of the packaging before each procedure and comply with the traceability required by the centre. Consult our technical-commercial team for special volume conditions: minimum order of 30 boxes, with scheduled deliveries for hospitals, clinics and purchasing centres.





Reviews