Description
1. Introduction
The Absorbable Multifilament Suture PGLA (Polyglactin 910) from the Prime line, reference P1L13HRT3620, is the technical reference solution for the professional seeking a surgical thread with predictable behaviour, excellent handling and great knot security in the approximation of soft tissues, ligatures and ophthalmic surgery. Its ½ circle needle with cylindrical body and 36 mm trocar point (HRT36) allows firm and controlled penetration, while the 90 cm violet grooved thread guarantees optimal visibility over the surgical field.
It is a synthetic, absorbable, sterile and single-use medical device, with a very competitive price-quality ratio for B2B procurement of operating theatres, ophthalmic clinics and purchasing centres. Commercial presentation in a box of 12 units and a minimum order of 30 boxes, which ensures the continuous supply of your healthcare centre.
2. Technical explanation
Polyglactin 910 is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, presented in a braided multifilament structure with a grooved finish. This configuration provides a regular surface that facilitates the passage of the thread through the tissue, smooth handling and firm, secure knot tying, even in medium sizes such as USP 2/0.
Absorption occurs by hydrolysis, i.e., by degradation of the polymer in contact with organic fluids, over a predictable period of 56 to 72 days. This profile makes the suture ideal for tissues that require wound support during the first weeks of healing, maintaining the necessary tensile strength while the tissue recovers its own resistance. The ½ circle needle with cylindrical body and 36 mm trocar point (tapercut) penetrates tissues of moderate or slightly fibrous tension with controlled trauma. The product is supplied sterile by ethylene oxide, single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. General characteristics of reference P1L13HRT3620
| Characteristic | Specification |
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Braided multifilament with grooved finish |
| Quality | Prime (Premisut range) |
| Gauge | USP 2/0 (metric 3) |
| Thread colour | Violet |
| Needle shape | ½ circle, cylindrical body |
| Tip type | Trocar / Tapercut |
| Needle size | HRT 36 mm |
| Thread length | 90 cm |
| Absorption mechanism | Hydrolysis (56-72 days) |
| Sterilisation | Ethylene oxide |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
| Reference (SKU) | P1L13HRT3620 |
3.2. Tensile strength retention profile
| Time | Tensile strength retention |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 30 % |
Complete absorption by hydrolysis between 56 and 72 days. Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a fully predictable absorption rate. Price per box of 12 units.
4. Equivalences and Cross-References
Polyglactin 910 is one of the most well-established synthetic absorbable suture raw materials on the market and clinically corresponds to the same chemical family used by international reference lines:
| Manufacturer / Company | Equivalent line |
| Ethicon (Johnson & Johnson) | Vicryl (coated braided polyglactin 910, violet colour) |
| B. Braun | Safil (braided polyglactin 910, violet colour) |
| Medtronic | Polysorb (absorbable braided range of polyglactin 910) |
| Laboratory Aragó | Polyglactin 910 range with ½ circle needles (470xx series and HRT/HS series) |
Within the Cahnos catalogue, it is part of the HRT36 trocar needle family in violet with its corresponding sizes: P1L13HRT3601 (USP 1), P1L13HRT3610 (USP 0) and P1L13HRT3630 (USP 3/0), all in 90 cm violet thread. In the polyglycolic acid (PGA) family, the performance equivalent with the same HRT36 needle is reference AMP113HRT3620 (USP 2/0, violet, 90 cm), with an absorption profile of 60 to 90 days.
5. Selection tips and variants of interest
The USP 2/0 size (metric 3) is indicated for soft tissues of moderate tension: skin, subcutaneous tissue, mucosa, muscle and ligatures, and is widely used in ophthalmic surgery. The 36 mm HRT trocar point needle is especially recommended for firm or slightly fibrous tissues, where the passage must be clean and without tearing; if a fully atraumatic conical point is preferred, HR versions of 36 mm are available in the same range. For procedures on very friable tissues or where minimal thread visibility is required, the colourless version of the family (P1L23HRT3601) can be considered. For longer force retention needs (180-210 days), the polydioxanone (PDO) family is the recommended alternative. All references are supplied in boxes of 12 units; the minimum order is 30 boxes to ensure the continuous supply of your healthcare centre. For special B2B conditions and volume orders, contact our technical-commercial team.