Description
1. Product Presentation
The Absorbable Multifilament Suture PGLA (Polyglactin 910) from the Premisut range, reference 1L13HRT3630, is the reference technical choice for the professional seeking a surgical thread with predictable behaviour, excellent handling and maximum knot security in the approximation of soft tissues, ligation and ophthalmic surgery.
Its 1/2 circle needle, cylindrical body and 36 mm trocar point (HRT 36) allows firm and controlled penetration even in tissues of a certain density, while the 90 cm violet fluted thread ensures optimal visibility on the surgical field and smooth sliding throughout the suture. It is a synthetic absorbable medical device, sterile and single-use, with a very competitive price-quality ratio for B2B supply to operating theatres, ophthalmic clinics and purchasing centres.
2. Technical Explanation
Polyglactin 910 is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, presented in a braided multifilament structure with a fluted finish. This configuration provides a regular surface that favours the passage of the thread through tissue, smooth handling and firm, secure knot tying, even in fine gauges such as USP 3/0.
Absorption occurs by hydrolysis, i.e., by degradation of the polymer in contact with organic fluids, over a predictable period of 56 to 72 days. This profile makes the suture suitable for tissues requiring wound support during the first weeks of healing, maintaining the necessary tensile strength while the tissue recovers its own resistance. The 1/2 circle needle with cylindrical body and trocar point (tapercut) penetrates tissues of moderate tension with controlled trauma, being especially indicated for tissue approximation, ligation and ophthalmic surgery procedures. The product is supplied sterile by ethylene oxide, single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. General characteristics of reference 1L13HRT3630
| Characteristic | Specification |
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Braided multifilament with grooved finish |
| Quality | Normal (range PREMISUT) |
| Gauge | USP 3/0 |
| Colour | Violet |
| Needle | 1/2 circle, cylindrical body, trocar point (tapercut) |
| Needle size | HRT 36 mm |
| Thread length | 90 cm |
| Absorption mechanism | Hydrolysis (56-72 days) |
| Sterilisation | Ethylene oxide |
| Tissue reaction | Minimal |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
| Reference (SKU) | 1L13HRT3630 |
3.2. Tensile strength retention profile
| Time | Tensile strength retention |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 30 % |
Complete absorption by hydrolysis between 56 and 72 days. Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a fully predictable absorption rate. Indicated for tissue approximation, ligation and ophthalmic surgery. Price per box of 12 units.
4. Equivalences and Cross-References
This multifilament polyglactin 910 suture is clinically positioned in the same family as the market-leading synthetic absorbable braided sutures, commonly used in general surgery, gynaecology, urology and specialised surgery:
| Manufacturer / Company | Equivalent line |
| Ethicon (Johnson & Johnson) | Vicryl (coated braided polyglactin 910, violet colour) |
| B. Braun | Safil (braided polyglactin 910, violet colour) |
| Medtronic | Polysorb (absorbable braided range of polyglactin 910) |
| Laboratory Aragó | Range of polyglactin 910 with 1/2 circle needles (470xx series and HRT/HS series) |
Within the Cahnos catalogue, its direct equivalent in structure and needle within the same brand is reference P1L13HRT3630 (Prime quality PGLA, HRT36 mm trocar needle, USP 3/0, violet, 90 cm). In the polyglycolic acid family, the performance equivalent is AM113HRT3630 (multifilament PGA, HRT36 mm trocar needle, USP 3/0, violet, 90 cm).
5. Selection tips and variants of interest
To choose the most suitable suture, consider the wound support time required by the tissue to be sutured and the necessary gauge according to the anatomical area. USP 3/0 gauge is indicated for tissues with moderate tension: skin, subcutaneous tissue, mucosa, fascia and soft tissue in general. The 36 mm trocar point needle is especially recommended for firm tissues and in ophthalmic surgery procedures, where the passage of the needle must be clean and without tearing.
| Reference | Description |
| P1L13HRT3630 | PGLA Prime 1/2 circle trocar point HRT36 mm, USP 3/0, violet, 90 cm (premium range) |
| 1L13HRT3620 | PGLA 1/2 circle trocar point HRT36 mm, USP 2/0, violet, 90 cm (smaller gauge) |
| 1L13HRT3610 | PGLA 1/2 circle trocar point HRT36 mm, USP 0, violet, 90 cm (larger gauge) |
| 1L13HRT3601 | PGLA 1/2 circle trocar point HRT36 mm, USP 1, violet, 90 cm (higher initial strength) |
| 1L13HR3630 | PGLA 1/2 circle conical point HR36 mm, USP 3/0, violet, 90 cm (conical point for soft tissues) |
| 1L15HRT48L01 | PGLA 1/2 circle trocar point HRT48 mm with loop, USP 1, violet, 150 cm (loop format) |
For longer force retention needs (180-210 days), the polydioxanone (PDO) family is the recommended alternative. All references are supplied in boxes of 12 units; the minimum order is 30 boxes to ensure continuous supply to your healthcare centre. For special B2B conditions and volume orders, contact our technical-commercial team.