Description
1. Commercial Introduction
The absorbable monofilament suture PREMISUT Prime made of poliglecaprone 25 (PGCL / caprolactone) is the technical solution for the approximation of soft tissues and surgical ligatures that require controlled absorption, minimal inflammatory response and particularly smooth passage of the thread. This double-strand presentation includes 2 identical 75 cm threads, each with its ½ circle needle with cylindrical body and conical Taperpoint point of 30 mm (HR30 Prime), in USP 3/0 gauge and fluted violet colour for high visibility in the field.
It is supplied in a box of 12 individual sterile sachets, ready to use, and is the ideal choice for operating theatre services, purchasing centres and continuous hospital supply programmes with a very competitive price-quality ratio in B2B procurement.
2. Technical explanation
Poliglecaprone 25 (PGCL) is a synthetic copolymer of glycolide and ε-caprolactone that degrades by hydrolysis, without enzymatic activity, which guarantees predictable absorption between 90 and 120 days and minimal tissue reaction. Being a fluted monofilament, the strand offers a smooth surface that slides easily through tissue, reduces the sawing effect and hinders bacterial adherence compared with multifilament threads, while maintaining secure tying with stable knots.
Tensile strength retention is approximately 60 % at the first week and 30 % at the second, covering the critical phase of tissue healing with contained tension. The ½ circle needle with cylindrical body and conical Taperpoint point is designed for soft tissues: it separates and displaces fibres without cutting effect, minimising damage to delicate structures such as vessels, peritoneum and subcutaneous tissue. The USP 3/0 gauge (metric 2) provides an optimal balance between handling and initial strength. The double-thread format (2×75 cm) reduces material waste per unit and is particularly cost-effective in procedures requiring multiple ligatures.
3. Technical Specifications
3.1. Technical data sheet of the reference
| Parameter | Specification |
| Material | Polyglecaprone 25 (PGCL) · copolymer of glycolide and ε-caprolactone |
| Thread structure | Synthetic absorbable monofilament, fluted finish |
| Quality | Prime (premium range) |
| Gauge / size | USP 3/0 (metric EP 2); conforms to USP and European Pharmacopeia |
| Swaged | 2 threads of 75 cm per unit, each with its needle (2×75 cm) |
| Thread colour | Ribbed violet (high visibility in the field) |
| Needle shape | ½ circle (1/2 circular) |
| Body and point | Cylindrical body, Taperpoint conical tip, atraumatic |
| Needle size | HR30 Prime (30 mm) |
| Absorption mechanism | Hydrolysis, without enzymatic activity |
| Absorption time | 90 – 120 days |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Indications | Soft tissue approximation and general ligatures |
| Factory reference | P312HR3030.2 (PREMISUT PGCL PRIME) |
| Reference Cahnos (SKU) | AMP312HR3030.2 |
| Presentation | Box of 12 units (individual sterile sachets) |
3.2. Tensile strength retention profile
| Period after implantation | Tensile strength retention |
| 1 week | 60 % of initial strength |
| 2 weeks | 30 % of initial strength |
| Complete absorption | 90–120 days, by hydrolysis |
3.3. Gauge equivalence according to USP/EP Pharmacopoeia (absorbable suture)
| USP size | Metric measurement (EP) | Approximate diameter |
| 5/0 | 1 | 0.10 – 0.149 mm |
| 4/0 | 1.5 | 0.15 – 0.199 mm |
| 3/0 (this product) | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
4. Equivalences and Cross-References
Poliglecaprone 25 is the polymer used in the MONOCRYL® line from Ethicon (Johnson & Johnson MedTech), a clinical reference in the category of synthetic absorbable monofilament of medium absorption, along with homologous families from DemeTECH (DemeCAPRONE®), SMI, Assut Medical and B. Braun (Monosyn, based on gluconate). For this specific configuration (PGCL monofilament, USP 3/0 gauge, ½ circle conical point needle of 30 mm and double violet thread of 75 cm, ref. P312HR3030.2) the following correspondence has been verified against official catalogue and authorised distribution:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (J&J MedTech) | MONOCRYL® – poliglecaprone 25 monofilament, violet, USP 3/0, 70 cm, SH needle ½ circle conical point of 26 mm | Y316H | Comparable by category (not interchangeable 1:1) | They match in material (poliglecaprone 25), USP 3/0 gauge, violet colour, monofilament structure and conical point family. They differ in needle size (26 mm SH vs HR30 of 30 mm), length and configuration (70 cm with single thread vs double thread 2×75 cm) and commercial presentation (box of 36 units vs 12 units). Code verified in official Ethicon catalogue; validate in technical data sheet/IFU before substituting in tender. |
| Ethicon (J&J MedTech) | MONOCRYL® Plus Antibacterial – poliglecaprone 25 with triclosan, violet, USP 3/0, 70 cm, SH needle of 26 mm, ½ circle, conical point | MCP316H | Comparable by category (not interchangeable 1:1) | Same base configuration as Y316H (3/0 · 70 cm · SH needle of 26 mm, ½ circle, conical) with added antibacterial triclosan load, a relevant functional difference compared with the untreated thread; needle size 26 mm vs HR30 of 30 mm and single-thread configuration vs double strand. Code verified in authorised distribution sources; validate in technical data sheet/IFU before substituting. |
There is no manufacturer-validated equivalence table that cross-references this specific reference (PGCL monofilament, HR30 Prime needle of 30 mm, USP 3/0 gauge, double violet thread of 2×75 cm, ref. P312HR3030.2) with references from other houses in the exact configuration; consult the technical department before using in tender or direct substitution. The factory reference PREMISUT of this item does not appear in public regulatory databases (GUDID/EUDAMED) and must be confirmed with the supplier before use in technical specifications. Before validating any candidate from another brand, check one by one: exact material (poliglecaprone 25 vs gluconate, PGA or PDO), USP gauge, configuration (number of threads and needles), ½ circle curvature, conical point family, needle size of 30 mm, thread length, colour, strength retention and current CE/MDR classification. The cited brands belong to their respective manufacturers and are mentioned only as comparative reference.
5. Selection tips and variants of interest
USP 3/0 (metric 2) with a 30 mm needle is very versatile: it is indicated for the approximation of soft tissues, subcutaneous tissue, superficial fascia and ligation of fine-to-medium calibre vessels, with reliable wound support during the first 2–3 weeks. The 30 mm half-circle needle with Taperpoint conical point is the most atraumatic option for delicate tissues; if greater penetration in firm or slightly fibrous tissues is required, consider the trocar-point versions (Tapercut/Ultra Plus) from the same family (HRP series), and if you prefer a larger needle radius, the HR36 and HR40 versions in the catalogue. The double-thread format (2×75 cm) provides twice as much useful material per sachet, reducing the cost per ligation.
| Reference | Variant of interest |
| AMP312HR3030.2 | PGCL Prime, HR30 30 mm conical point needle, USP 3/0, violet, 2 strands of 75 cm (product analysed) |
| AMP312HR3030 | PGCL Prime, same HR30 30 mm needle, USP 3/0, violet, 1 strand of 75 cm (single strand) |
| AM312HR3030 | PGCL standard quality, HR30 30 mm needle, USP 3/0, violet, 75 cm |
| AMP312HR3020.2 | PGCL Prime, HR30 30 mm needle, USP 2/0, violet, 2 strands of 75 cm (larger calibre) |
| AMP312HR3010.2 | PGCL Prime, HR30 30 mm needle, USP 0, violet, 2 strands of 75 cm (larger calibre) |
| AMP312HR3001.2 | PGCL Prime, HR30 30 mm needle, USP 1, violet, 2 strands of 75 cm (high initial tensile strength) |
| AMP312HR3620.2 / AMP312HR4020.2 | PGCL Prime, HR36 36 mm / HR40 40 mm conical point needle, USP 2/0, violet, 2 strands of 75 cm (larger needle radius) |
If the procedure requires longer wound support, consider polydioxanone PDO (180–210 days); if a shorter absorption profile is needed, consider PGA (60–90 days) or polyglactin 910 (56–72 days). As this is sterile single-use material, check the integrity of the sachet and the expiry date before each procedure, and store the suture in a cool, dry place below 30 °C, protected from direct sunlight. For special B2B volume conditions, hospital supply contracts or equivalence validation with other commercial brands, contact the technical-commercial team at Cahnos.