Description
1. Commercial Introduction
The absorbable multifilament PGAR RAPID suture from the PREMISUT RAPID range is the rapid-absorption reference for B2B supply to operating theatres, plastic-cosmetic surgery clinics, dermatology services and purchasing centres that need a synthetic absorbable thread with predictable handling, high knot security and complete disappearance within a few weeks, without the need for removal.
This configuration combines a braided polyglactic acid thread in USP 3/0 gauge, colourless and 75 cm long, with a ⅜ circle needle with precision reverse cutting triangular point and square body of 19 mm (XDS19 profile): the ideal combination for tissue approximation, ligatures and precision work on skin and mucous membranes. Sterile single-use product, presented in a box of 12 units with individual sachets, ideal for hospital replenishment, tenders and continuous supply contracts.
2. Technical Explanation
PGAR suture (low molecular weight polyglycolic acid) is a synthetic polymer 100% absorbable by hydrolysis, presented in a braided multifilament structure and coated with polycaprolactone and calcium stearate. This coating reduces friction when passing through tissues, improves thread sliding and facilitates firm, secure knotting, with excellent first-knot retention.
Its rapid absorption profile is designed for short-duration healing: it retains 50% of its tensile strength at 5 days and loses practically all its resistance at 14 days, completing total absorption between 40 and 45 days through physiological hydrolysis, with minimal tissue reaction.
The XDS19 needle features a ⅜ circle curvature with precision reverse cutting triangular point (extra-long cutting edge) and square body: the cutting edge cuts in the opposite direction to the wound bed, generating a clean entry hole that preserves the skin edge and prevents the cut from widening. The square section of the body optimises grip in the needle holder, preventing rotation and improving the precision of the surgical gesture. Sterilisation by gamma radiation; single-use product, not re-sterilisable.
3. Technical Specifications
3.1. General characteristics
| Parameter | Specification |
| Factory reference | P222XDS1930 (Prime range) |
| Reference Cahnos (SKU) | AMP222XDS1930 |
| Brand / Range | PREMISUT RAPID |
| Thread material | Polyglactic acid (PGA) of low molecular weight, 100% absorbable |
| Thread structure | Braided multifilament, non-ribbed |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 3/0 (metric EP 2), conforming to USP/EP Pharmacopoeia |
| Approximate thread diameter | 0.20 – 0.249 mm |
| Thread length | 75 cm (free end, without loop) |
| Thread colour | Colourless |
| Needle shape | ⅜ circle (3/8 circular) |
| Tip type | Precision reverse cutting triangular (extra-long cutting edge), square body |
| Needle size | XDS 19 mm (Prime), 1 needle per thread |
| Swaged | 1 needle / 1 suture per unit |
| Sterilisation | Gamma radiation; sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures, general surgery and plastic-cosmetic surgery |
| Commercial presentation | Box of 12 units (sterile individual sachets), minimum order of 10 boxes |
| Regulatory standard | Conforming to USP/EP Pharmacopoeia for synthetic absorbable sutures; CE marking (MDR) |
3.2. Absorption profile and tensile strength retention
| Parameter | Value |
| Absorption type | Hydrolysis |
| Absorption time | 40 – 45 days |
| Tensile strength at 5 days | 50% of initial strength |
| Tensile strength at 14 days | 0% (total loss of tensile support) |
| Absorption rate | Predictable and uniform |
| Tissue reaction | Minimal |
3.3. Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 4/0 | 1.5 | 0.15 – 0.199 mm |
| 3/0 (this product) | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
The thread diameter complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without exceeding its limits.
3.4. Reference code decoding
The official reference P222XDS1930 breaks down as follows: P (Prime range) + 222 (75 cm thread) + XDS19 (⅜ circle needle with precision reverse cutting triangular point and square body of 19 mm, plastic/cosmetic profile) + 30 (USP 3/0 gauge). Cahnos markets the official reference without altering characters: no digit may be added, removed or reordered when ordering or tendering.
4. Equivalences and Cross-References
Rapid-absorption polyglycolic acid (PGAR) is one of the most widespread braided absorbable suture families on the market, comparable at clinical category level with the rapid-absorption lines of major manufacturers. The reference P222XDS1930 shares low molecular weight absorbable material, USP 3/0 gauge, ⅜ circle curvature, reverse cutting needle and colourless thread with the homologous lines in the segment. However, the comparison is declared at category level, not interchangeable 1:1, because the needle profile and body differ (XDS with square body and precision cutting edge versus standard reverse cutting profiles), as do the coating, thread length and commercial presentations.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon / Johnson & Johnson MedTech | VICRYL RAPIDE™ (braided polyglactin 910, colourless) | NW2732 – USP 3/0, 3/8 circle, reverse cutting 26 mm, 70 cm | Comparable by category (not interchangeable 1:1) | Same material and rapid absorption profile (50% at 5 days, 0% at 10-12 days according to the data sheet); the needle differs (26 mm, round body, versus XDS19 with extra-long cutting edge and square body) and the thread length (70 cm versus 75 cm). Code confirmed in the official Ethicon catalogue and authorised distribution. |
| Ethicon / Johnson & Johnson MedTech | Coated VICRYL® (braided polyglactin 910, colourless) | J427H – USP 3/0, PS-2 needle 19 mm, 3/8 circle, reverse cutting, 70 cm | Comparable by category (not interchangeable 1:1) | Same material, gauge and needle size (19 mm); the absorption profile differs (standard VICRYL longer versus the RAPID variant), the thread length (70 cm) and the needle body (PS-2 without square body). Code verified in authorised distribution catalogues. |
| B. Braun / Aesculap | Safil Quick™ (low molecular weight polyglycolic acid, braided and coated) | C1046657 – USP 3/0, 3/8 circle needle, reverse cutting with micro-point 24 mm, 70 cm undyed, box of 36 | Comparable by category (not interchangeable 1:1) | Same base material and rapid absorption profile; the needle size (24 mm), length (70 cm) and presentation (36 units versus 12 units) differ. Code confirmed in authorised distribution data sheet (WMS Medical) and in the NHS product database (dmd-browser). |
| B. Braun / Aesculap | Safil Quick™ (low molecular weight polyglycolic acid) | C1046421 – USP 3/0, 3/8 circle needle, cutting 19 mm, 70 cm undyed | Comparable by category (not interchangeable 1:1) | Same material and needle size (19 mm); the point family differs (standard cutting versus XDS precision reverse cutting with extra-long cutting edge), the needle body and the thread length (70 cm versus 75 cm). Code verified in authorised distribution. |
Other homologous families in the segment identified on the market, whose technical data sheets must be consulted before any substitution and for which no verified code is available in this specific configuration: DemeCRYL™ Fast from DemeTECH and Surgicryl Fast Absorb from SMI. The cited brands belong to their respective manufacturers and are mentioned solely as comparative reference.
There is no equivalence table validated by the manufacturer for this reference; consult the technical department before using in tenders.
5. Selection tips and variants of interest
USP size 3/0 (metric 2) is indicated for tissues with moderate tension: skin, subcutaneous tissue, mucous membranes and soft tissue closures. The XDS19 Prime needle (precision reverse cutting with square body) is especially recommended in plastic-cosmetic surgery, dermatology and ophthalmology, where the needle passage must be clean and without tearing.
| Variant / Reference | Description |
| AMP222XDS1940 (P222XDS1940) | PGAR RAPID Prime XDS19 – USP 4/0 undyed – 75 cm (same needle, finer gauge) |
| AMP221XDS1930 (P221XDS1930) | PGAR RAPID Prime XDS19 – USP 3/0 undyed – 45 cm (same needle and gauge, short strand) |
| AMP222DS1930 | PGAR RAPID, classic reverse cutting DS19 needle – USP 3/0 undyed – 75 cm (profile without square body, economical alternative) |
| AMP2L22XDS1930 (P2L22XDS1930) | PGLAR Polyglactin 910 RAPID Prime XDS19 – USP 3/0 undyed – 75 cm (same needle geometry, PGLAR polymer) |
| AMP122DS1930 (P122DS1930) | PGA Prime standard DS19 – USP 3/0 undyed – 75 cm (medium absorption 60-90 days for greater tensile support) |
The undyed version offers maximum aesthetic discretion on the surgical field; if greater visibility is required, request the same configuration in violet on demand. Sterile single-use product: check the integrity of the packaging and the expiry date before each procedure, keep the batch number for traceability and store below 30 °C, away from moisture and direct sunlight. For special B2B conditions, volume orders or hospital supply agreements, contact the technical-commercial team at Cahnos.