Description
1. Commercial Introduction
The multifilament absorbable suture PREMISUT RAPID in Prime quality is the accelerated absorption closure solution for general and ambulatory surgery theatres, emergency departments, minor surgery and B2B purchasing centres that require a braided, flexible thread with early and fully predictable loss of tensile strength, without the need for removal. This configuration combines a 75 cm undyed strand in USP 3/0 gauge with a ⅜ circle, reverse cutting triangular point needle of 16 mm (DS16 Prime profile): clean penetration in skin and firm tissues with excellent aesthetic outcome and minimal tissue reaction. It is supplied in a box of 12 sterile units with batch traceability, designed for hospital replenishment, clinics, distribution and tenders.
2. Technical Explanation
Rapid-profile polyglycolic acid (PGAR) is a low molecular weight synthetic polymer, spun into a braided multifilament structure and coated with polycaprolactone and calcium stearate. This coating reduces friction when passing through tissue, improves thread glide and facilitates firm, secure knotting, with minimal tissue reaction. In its RAPID variant, absorption occurs by hydrolysis —a predictable process independent of the patient’s enzymatic activity— within 40 to 45 days: it retains 50 % of the initial tensile strength at 5 days and loses it completely at 14 days, an ideal profile for tissues that only require temporary support during the first phase of healing.
The DS16 needle has a ⅜ circle curvature (approx. 135°) with a reverse cutting triangular body and point: the cutting edge is oriented outward from the wound, creating a clean entry hole that preserves the skin edge and prevents the cut from widening in the puncture bed. It is especially useful in skin, mucous membranes and firm or non-distensible tissues, with a reduced radius that is highly manageable in small surgical fields. The undyed (uncoloured) thread provides maximum aesthetic discretion on the field. Sterilisation by gamma radiation; single-use product, not to be re-sterilised.
3. Technical Specifications
3.1. Technical data sheet of the reference
| Parameter | Specification |
| Factory reference | P222DS1630 (Prime range) |
| Reference Cahnos (SKU) | AMP222DS1630 |
| Brand / Range | PREMISUT RAPID – Prime quality |
| Thread material | 100% rapid absorption polyglycolic acid (PGAR), absorbable by hydrolysis |
| Thread structure | Coated braided multifilament |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 3/0 (metric EP 2), conforming to USP/EP Pharmacopoeia |
| Approximate thread diameter | 0.20 – 0.249 mm |
| Thread colour | Colourless (violet on demand) |
| Thread length | 75 cm (free end, without loop) |
| Needle shape | ⅜ circle (3/8 circular, approx. 135°), body and point |
| Tip type | Reverse cutting triangular (reverse cutting) |
| Needle size | DS 16 Prime (16 mm), 1 needle per strand |
| Swaged | 1 needle / 1 suture per unit |
| Absorption mechanism | Hydrolysis; complete between 40 and 45 days |
| Tensile strength | 50% at 5 days; 0% at 14 days |
| Sterilisation | Gamma radiation; sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Approximation of soft tissues, ligatures and general surgery |
| Commercial presentation | Box of 12 units (individual sterile sachets); minimum order 30 boxes |
| Regulatory standard | Conforming to USP/EP Pharmacopoeia for synthetic absorbable sutures; CE marking (MDR) |
3.2. Tensile strength retention profile
| Post-implantation time | Retained tensile strength |
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% (total loss of tensile support) |
| Complete absorption | 40 – 45 days by hydrolysis (predictable and uniform rate) |
The profile cited corresponds to the rapid absorption formulation (RAPID line); it must not be confused with standard absorption polyglycolic acid (PGA line, 60–90 days) nor with standard polyglactin 910 (56–72 days).
3.3. Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 5/0 | 1 | 0.10 – 0.149 mm |
| 4/0 | 1.5 | 0.15 – 0.199 mm |
| 3/0 (this product) | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
The strand diameter complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without extending its limits.
4. Equivalences and Cross-References
Rapid absorption polyglycolic acid (PGAR) belongs to the segment of braided multifilament absorbable sutures with early tensile strength loss, whose international clinical benchmark is the rapid absorption line of major manufacturers. Only references whose codes and attributes have been verified in the official catalogue and authorised distribution are listed; as they are from other manufacturers, they are declared comparable by category (not 1:1 interchangeable) and must be validated against the official datasheet before any substitution.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (J&J MedTech) | VICRYL RAPIDE™ (polyglactin 910, braided, undyed) | W9932 – USP 3/0, 75 cm thread, 26 mm Prime reverse cutting needle, ⅜ circle, undyed, box of 12 | Comparable by category (not interchangeable 1:1) | They share rapid absorption profile (≈42 days), USP 3/0 gauge, 75 cm length, ⅜ circle curvature, reverse cutting point family and undyed colour; they differ in base polymer (polyglactin 910 versus polyglycolic acid) and needle size (26 mm versus DS16 16 mm). Code confirmed in Ethicon EMEA catalogue and authorised distribution. |
| B. Braun / Aesculap | Safil Quick™ (low molecular weight polyglycolic acid, braided and coated) | C1046421 – USP 3/0, 70 cm thread, 19 mm ⅜ circle cutting needle, undyed | Comparable by category (not interchangeable 1:1) | Same base material (PGAR), rapid absorption profile and USP 3/0 gauge; they differ in needle size (19 mm versus 16 mm) and thread length (70 cm versus 75 cm). Code confirmed in authorised distribution; verify exact needle geometry before substituting. |
Homologous ranges in the rapid absorption segment also include Novosyn Quick (B. Braun), DemeCRYL Fast (DemeTECH) and Surgicryl Fast Absorb (SMI), comparable by category by material and absorption profile; for the exact combination USP 3/0 / DS16 mm needle / 75 cm thread, their specific codes must be confirmed with each manufacturer. The standard VICRYL range (absorption 56–72 days) is not comparable to this RAPID version. The brands mentioned belong to their respective manufacturers and are listed only as comparative reference. There is no equivalence table validated by the manufacturer; consult with the technical department of Cahnos before using in a tender, verifying in each official datasheet the USP 3/0 gauge (metric 2), the ⅜ circle curvature, the type and size of needle (reverse cutting triangular 16 mm), the thread length (75 cm) and the undyed colour.
5. Selection tips and variants of interest
The USP 3/0 gauge (metric 2) offers an optimal balance between strength and fineness, indicated for tissues with moderate tension: skin, subcutaneous tissue, mucosa and soft tissue in general, as well as for ligatures of small-calibre vessels. The DS16 ⅜ circle reverse cutting point needle is the recommended option when firm, precise penetration is required in firm tissues without enlarging the exit hole, typical in general surgery, minor surgery and surgical fields with restricted access. Prime quality guarantees consistency of sharpness and finish batch to batch. Within the PREMISUT system itself, the PGAR RAPID family offers the same material and absorption profile in other gauges, geometries and lengths, fully interchangeable with each other while respecting the gauge and needle geometry:
| SKU | Variant of interest |
| AMP222DS1630 (P222DS1630) | PGAR RAPID Prime DS16 – USP 3/0 undyed – 75 cm (this product) |
| AM222DS1630 (222DS1630) | PGAR RAPID standard quality, same configuration DS16 – USP 3/0 undyed – 75 cm (direct equivalent within the same manufacturer) |
| AMP222DS1640 (P222DS1640) | PGAR RAPID Prime DS16 – USP 4/0 undyed – 75 cm (immediately finer gauge) |
| AMP222DS1650 (P222DS1650) | PGAR RAPID Prime DS16 – USP 5/0 undyed – 75 cm (fine gauge for delicate tissue) |
| AMP222DS2630 (P222DS2630) | PGAR RAPID Prime DS26 – USP 3/0 undyed – 75 cm (same strand, 26 mm needle for larger fields) |
| AMP222DRT1630 (P222DRT1630) | PGAR RAPID Prime DRT16 trocar point – USP 3/0 undyed – 75 cm (Tapercut point for medium-consistency tissues) |
If the protocol prefers atraumatic penetration in lax tissues, consider the conical point series (HR / taperpoint); if greater thread visibility over the surgical field is required, request the same configuration in violet on demand. Sterile single-use product: check the integrity of the pouch and the expiry date before each procedure, keep the reference alongside the batch number for traceability in the operating theatre, and store below 30 °C in a cool, dry place, protected from direct sunlight. For special B2B conditions, volume orders or hospital supply agreements, contact the technical-commercial team at Cahnos.





Reviews