Description
1. Commercial Introduction
The Absorbable Multifilament Suture PGAR RAPID from the PREMISUT RAPID range is the accelerated absorption solution for B2B supply to operating theatres, day hospitals, minor ambulatory surgery and purchasing centres that need a braided thread with early and fully predictable loss of strength, without the need for removal. This configuration combines a polyglycolic acid strand in USP 4/0 gauge, undyed and 75 cm long, with a ⅜ circle needle with triangular reverse cutting body and point of 16 mm (DS16 profile): firm cutting penetration in skin and mucosa without tearing the wound edge. It is presented in a box of 12 units in individual sterile sachets with batch traceability, at a very competitive price for hospital replenishment, tenders and continuous supply contracts.
2. Technical Explanation
PGAR (rapid-profile polyglycolic acid) is a low-molecular-weight synthetic polymer presented in a braided multifilament structure and coated with polycaprolactone and calcium stearate. This coating reduces friction when passing through tissues, improves thread sliding and facilitates firm, secure knotting, with excellent first-knot retention and minimal tissue reaction.
Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the polymer degrades into glycolic acid, a substance that the body metabolises without generating relevant tissue reaction. In the RAPID variant, complete absorption occurs between 40 and 45 days, retaining 50% of the initial tensile strength at 5 days and losing it completely at 14 days; this profile makes it suitable for tissues that only require transient support during the first phase of healing.
The ⅜ circle needle (approx. 135°) with 16 mm triangular reverse cutting body and point directs the cutting edge away from the wound: it creates a clean entry hole that preserves the skin edge and prevents the cut from widening in the puncture bed, being especially useful in skin, mucosa and firm or poorly distensible tissues. The undyed colour minimises visual interference on the field and is preferred in aesthetic areas or superficial tissue. Sterilisation by gamma radiation; single-use product, not re-sterilisable.
3. Technical Specifications
3.1. General characteristics (Ref. 222DS1640 · SKU AM222DS1640)
| Brand / Range | PREMISUT RAPID (standard quality) |
| Factory reference | 222DS1640 |
| SKU Cahnos | AM222DS1640 |
| Composition | 100% Polyglycolic Acid (PGAR) rapid absorption, absorbable by hydrolysis |
| Thread structure | Braided multifilament |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 4/0 (metric 1.5), conforming to USP/EP Pharmacopoeia |
| Approximate diameter | 0.15 – 0.199 mm |
| Thread colour | Colourless |
| Thread length | 75 cm (free end, without loop) |
| Needle shape | ⅜ circle (3/8 circular, approx. 135°), body and point |
| Tip type | Triangular reverse cutting (reverse cutting) |
| Needle size | DS 16 (16 mm), 1 needle per thread |
| Swaged | 1 needle / 1 suture per unit |
| Absorption mechanism | Hydrolysis; complete between 40 and 45 days |
| Tensile strength retention | 50% at 5 days; 0% at 14 days |
| Sterilisation | Gamma radiation; sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures and general surgery; skin and mucosa closure |
| Commercial presentation | Box of 12 units (individual sterile sachets) |
| Regulatory standard | Conforms to USP/EP Pharmacopoeia for synthetic absorbable sutures |
3.2. Absorption profile and tensile strength retention
| Post-implantation time | Retained tensile strength |
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% (total loss of tensile support) |
| Complete absorption | 40 – 45 days by hydrolysis (predictable and uniform rate) |
The cited retention profile corresponds to the rapid absorption formulation (RAPID line); it should not be confused with standard absorption polyglycolic acid (PGA line, 60–90 days) or with standard polyglactin 910 (56–72 days).
3.3. Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 5/0 | 1 | 0.10 – 0.149 mm |
| 4/0 (this product) | 1.5 | 0.15 – 0.199 mm |
| 3/0 | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
The strand diameter complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without extending their limits.
4. Equivalences and Cross-References
Rapid absorption polyglycolic acid (PGAR) belongs to the segment of braided multifilament absorbables with early strength loss, whose international clinical reference is the rapid absorption line of the major suture manufacturers. The references indicated below have been verified in authorised distribution catalogues and coincide with this data sheet in USP 4/0 gauge, ⅜ circle curvature, reverse cutting point family, undyed colour and strand length within the same support window; as they are from other manufacturers and may present some difference in measurement or composition, they are declared comparable by category (not interchangeable 1:1) and must be validated against the official data sheet before any substitution.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (J&J MedTech) | VICRYL RAPIDE™ (polyglactin 910, braided, undyed) | W9922 – USP 4/0, 75 cm, undyed, ⅜ circle reverse cutting needle PS-2 of 19 mm | Comparable by category (not interchangeable 1:1) | They match rapid absorption profile (≈42 days), USP 4/0 gauge, ⅜ circle curvature, reverse cutting point family, undyed thread and 75 cm length; they differ in chemical composition (polyglactin 910 vs. PGA homopolymer) and needle size (19 mm vs. DS16 of 16 mm). Code confirmed in authorised distribution (box of 12 units). |
| B. Braun / Aesculap | Safil Quick™ (low molecular weight polyglycolic acid, braided and coated) | C1046213 – USP 4/0, 16 mm, 70 cm, undyed, ⅜ circle reverse cutting needle | Comparable by category (not interchangeable 1:1) | They match base material (low molecular weight PGA), rapid absorption profile, USP 4/0 gauge, 16 mm needle size and reverse cutting point family; they differ in thread length (70 cm vs. 75 cm) and manufacturer. Code confirmed in authorised distribution; verify exact needle geometry before substituting. |
There is no equivalence table validated by the manufacturer that cross-references this reference with other brands; consult the technical department of Cahnos before use in tenders, checking on each official datasheet the USP 4/0 gauge (metric 1.5), the ⅜ circle curvature, the needle type and size (16 mm reverse cutting triangular), the suture length (75 cm) and the undyed colour. Comparable ranges in the rapid-absorption segment also include Novosyn Quick (B. Braun), DemeCRYL Fast (DemeTECH) and Surgicryl Fast Absorb (SMI), comparable by category, material and absorption profile; for the exact USP 4/0 / DS16 mm needle / 75 cm suture combination, their specific codes must be confirmed with each manufacturer. The standard VICRYL range (56–72 day absorption) is not equivalent to this RAPID version. The brands cited belong to their respective manufacturers and are mentioned solely as comparative reference; no unverified cross-reference codes are published.
5. Selection tips and variants of interest
The USP 4/0 gauge (metric 1.5) is a common choice for fine skin, mucous membranes, subcutaneous tissue and soft tissues under moderate tension, offering an optimal balance between initial strength and fineness. The 16 mm ⅜ circle reverse cutting needle is especially recommended when a firm, precise passage through skin or mucous membranes is required without tearing the edge, typical in general surgery, minor surgery and small surgical fields. The undyed suture offers maximum aesthetic discretion; if maximum visibility in the field is preferred, request the same configuration in dyed colour on demand.
| SKU | Variant of interest |
| AM222DS1640 (222DS1640) | PGAR RAPID USP 4/0, 3/8 triangular DS16 needle, undyed, 75 cm (current product) |
| AMP222DS1640 (P222DS1640) | PGAR RAPID Prime USP 4/0, same DS16 needle, undyed, 75 cm (superior sharpening quality) |
| AM222DS1630 | PGAR RAPID USP 3/0, DS16 needle, undyed, 75 cm (next larger gauge) |
| AM222DS1650 | PGAR RAPID USP 5/0, DS16 needle, undyed, 75 cm (fine gauge for delicate tissue) |
| AM222DS1940 | PGAR RAPID USP 4/0, DS19 needle, undyed, 75 cm (same needle in a larger size) |
| AM122DS1640 | Standard PGA (60–90 day absorption) USP 4/0, DS16 needle, undyed, 75 cm (prolonged tensile support) |
Sterility and storage: product sterile by gamma radiation and single-use; check the integrity of the pouch and the expiry date before each procedure, keep the reference together with the batch number for traceability in the operating theatre and store below 30 °C, away from moisture and direct sunlight. For B2B volume conditions, hospital supply agreements or equivalents from other manufacturers, contact the technical-commercial team at Cahnos.





Reviews
There are no reviews yet.