Description
1. Commercial Introduction
The multifilament absorbable suture PGAR RAPID Prime (PREMISUT RAPID) is the accelerated absorption surgical closure solution designed for general surgery operating theatres, ambulatory surgery, day hospitals and B2B purchasing centres that need a braided, flexible thread with early and fully predictable loss of tensile strength. Its ½ circle needle with cylindrical body and 26 mm Tapercut trocar point penetrates medium-consistency tissue firmly and without a broad cutting effect, while the 75 cm colourless thread offers maximum discretion on the surgical field. Cost-effective presentation in a box of 12 units, with a minimum order of 30 boxes for the professional channel.
2. Technical Explanation
Polyglycolic acid (PGAR) is a 100% absorbable synthetic polymer spun into a braided multifilament structure and coated with polycaprolactone and calcium stearate, providing flexibility, low shape memory, smooth passage through tissue and firm, secure knot tying. In its RAPID variant (low molecular weight), absorption occurs by hydrolysis —a predictable process independent of the patient’s enzymatic activity— within 40 to 45 days: it retains approximately 50% of the initial tensile strength at 5 days and loses it completely at 14 days, ideal for tissues that heal quickly and where it is desired to minimise residual foreign material.
The ½ circle (160°) needle with cylindrical body and trocar point (Tapercut, HRT26 profile) combines the atraumatic penetration of the round body with a triangular sharpening at the tip, capable of traversing somewhat firmer tissues than a conical point without tearing or enlarging the exit hole. The USP 3/0 (metric 2) gauge provides an optimal balance between strength and fineness for fine and medium thickness sutures. Gamma radiation sterilisation guarantees final sterility with minimal alteration of mechanical properties; single-use product, not re-sterilisable.
3. Technical Specifications
3.1. Technical data sheet of the reference
| Parameter | Specification |
|---|---|
| Reference (SKU) | AMP222HRT2630 · Factory ref. P222HRT2630 (box of 12 units) |
| Brand / Quality | PREMISUT RAPID – PGAR Prime |
| Material | 100% polyglactic acid (PGAR), synthetic and absorbable, rapid-absorption variant |
| Thread structure | Braided multifilament, undyed |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 3/0 (metric 2; approximate diameter 0.20–0.249 mm according to Pharmacopoeia) |
| Thread colour | Colourless |
| Thread length | 75 cm |
| Needle shape | ½ circle (1/2, 160°), cylindrical body |
| Needle tip | Trocar / Tapercut (triangular) |
| Needle size | 26 mm (HRT26 Prime), 1 needle per thread |
| Swaged | 1 needle / 1 suture per unit |
| Absorption mechanism | Hydrolysis; complete between 40 and 45 days |
| Sterilisation | Gamma radiation; sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures and general surgery |
| Commercial format | Box of 12 units (individual sterile sachets); minimum order 30 boxes |
| Regulatory standard | Conforms to USP/EP Pharmacopoeia for synthetic absorbable sutures |
3.2. Tensile strength retention profile
| Post-implantation time | Retained tensile strength |
|---|---|
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% (total loss of tensile support) |
| Total absorption | 40–45 days, by hydrolysis (predictable and uniform rate) |
The cited retention profile corresponds to the rapid absorption formulation (RAPID line); it must not be confused with standard absorption polyglactin 910 (PGLA line, 56–72 days), which is not comparable in terms of tensile support window.
3.3. Gauge equivalence according to USP/EP Pharmacopoeia (synthetic absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
|---|---|---|
| 0 | 3.5 | 0.35 – 0.399 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
| 3/0 (this product) | 2 | 0.20 – 0.249 mm |
The thread diameter complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without exceeding its limits.
4. Equivalences and Cross-References
Rapid absorption polyglycolic acid (PGAR) is one of the most widespread braided multifilament absorbable suture families in the surgical market, comparable at clinical category level with the rapid absorption lines of major manufacturers. Only the reference whose code and attributes have been verified in the manufacturer’s own sources (Ethicon EMEA catalogue) and authorised distributors is included:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Ethicon (Johnson & Johnson MedTech) | VICRYL RAPIDE (polyglactin 910, braided, colourless) | W9932 – USP 3/0, 75 cm, colourless, 26 mm PS Prime needle | Comparable by category (not interchangeable 1:1) | They match in rapid absorption material (≈42 days), USP 3/0 gauge, colourless colour, 75 cm length, 26 mm needle, 12-unit format and tensile retention profile. They differ in needle geometry: 3/8 circle reverse cutting (PS) in Ethicon versus ½ circle with Tapercut trocar point (HRT26) of this reference, and the braid coating. Code confirmed in Ethicon EMEA (W9932) and authorised distributors; verify needle geometry in the official datasheet before substituting. |
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. Homologous ranges in the rapid absorption segment include Safil Quick and NOVOSYN QUICK (B. Braun), DemeCRYL Fast (DemeTECH) and Surgicryl Fast Absorb (SMI), comparable by category by material and absorption profile; for the exact combination USP 3/0 · 26 mm trocar needle · 75 cm thread, their specific codes must be confirmed with each manufacturer before any substitution. Standard VICRYL (absorption 56–72 days) is not comparable to this RAPID version.
5. Selection tips and variants of interest
The USP 3/0 (metric 2) gauge offers an optimal balance between strength and fineness, indicated for tissues with moderate tension: skin, subcutaneous tissue, mucosa and soft tissue in general, as well as for ligation of small-calibre vessels. The 26 mm needle with Tapercut trocar point is the recommended option when firm penetration without a broad cutting effect is required in medium-consistency tissues.
| Variant (SKU) | Description |
|---|---|
| AM222HRT2630 (222HRT2630) | PGAR RAPID HRT26 – USP 3/0 colourless – 75 cm (standard quality, same configuration) |
| AMP222HRT2620 (P222HRT2620) | PGAR RAPID Prime HRT26 – USP 2/0 colourless – 75 cm (higher gauge for more tension) |
| AMP228HRT2610 (P228HRT2610) | PGAR RAPID Prime HRT26 – USP 0 colourless – 75 cm (larger gauge) |
| AMP2L22HRT2630 (P2L22HRT2630) | PGLAR Polyglactin 910 RAPID Prime – HRT26 – USP 3/0 colourless – 75 cm (same geometry, glycolide/L-lactide copolymer) |
| AMP222HRT3620 (P222HRT3620) | PGAR RAPID Prime HRT36 – USP 2/0 colourless – 75 cm (36 mm needle) |
| AMP223HRT4020 (P223HRT4020) | PGAR RAPID Prime HRT40 – USP 2/0 colourless – 90 cm (long trocar needle) |
If the protocol prefers even more atraumatic penetration in lax tissues, consider the conical point series (HR / HRP taperpoint) from the same range; if a larger working field is required, consult the HRT36/HRT40 needles. The colourless version offers maximum discretion in the field; if greater visibility is required, consult the same configuration in dyed thread on request. As this is a sterile single-use product, check the integrity of the packaging and the expiry date before each procedure and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or equivalences from other commercial houses, contact the technical-commercial team at Cahnos.