Description
1. Commercial Introduction
The absorbable multifilament suture PREMISUT RAPID PGAR is the safe choice when the procedure requires short-term wound support, predictable resorption and excellent tissue handling. Made from 100% polyglycolic acid and coated with polycaprolactone and calcium stearate, it offers smooth penetration, secure knotting and minimal tissue response. Supplied in a box of 12 sterile units, it is a reliable and cost-effective solution for operating theatres, hospital purchasing departments and clinics, with special terms for B2B supply contracts.
2. Technical Explanation
Polyglycolic acid (PGA) is a synthetic absorbable polymer spun into a coated braided multifilament structure. Its RAPID variant (low molecular weight) degrades by hydrolysis —a predictable process independent of the patient’s enzymatic activity— within 40 to 45 days, retaining approximately 50% of the initial tensile strength at 5 days and losing it completely at 14 days. This makes it suitable for superficial tissues and temporary approximations where material removal is not required. The ½ circle needle with cylindrical body and taperpoint conical tip (HR 22 mm) penetrates soft tissues without cutting, minimising tissue damage, ideal for surgical fields with limited access in general surgery. The colourless thread provides maximum discretion on the field. Sterilisation by gamma radiation; single-use product, not re-sterilisable.
3. Technical Specifications
3.1. Technical data sheet for reference 222HR2220
| Material | 100% polyglycolic acid (PGA / PGAR) braided multifilament |
| Coating | Polycaprolactone and calcium stearate |
| Thread structure | Braided multifilament |
| Gauge | USP 2/0 (metric 3) |
| Colour | Colourless |
| Thread length | 75 cm (single strand) |
| Needle shape | ½ circle |
| Needle body / tip | Cylindrical body, taperpoint conical tip |
| Needle size | HR 22 (22 mm) |
| Absorption mechanism | Hydrolysis, 40-45 days |
| Tensile strength | 50% at 5 days; 0% at 14 days |
| Sterilisation | Gamma radiation; single-use, not re-sterilisable |
| Tissue reaction | Minimal |
| Indications | Approximation of soft tissues, ligatures and general surgery |
| Factory reference | 222HR2220 |
| SKU Cahnos | AM222HR2220 |
| Commercial format | Box of 12 units |
3.2. Tensile strength retention profile
| Post-implantation time | Retained tensile strength |
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% (total loss of tensile support) |
| Total absorption | 40-45 days, by hydrolysis (predictable and uniform rate) |
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
| 2/0 | 3 | 0.30 – 0.339 mm (indicative, according to USP Pharmacopoeia) |
4. Equivalences and Cross-References
Fast-absorbing polyglycolic acid (low molecular weight) is one of the most widespread categories of absorbable multifilament in the surgical market. At category level, this reference corresponds to synthetic absorbable braided sutures with rapid strength loss: the closest homologous family is Safil Quick (B. Braun), composed of low molecular weight polyglycolic acid with the same absorption profile of 40-45 days and total strength loss at 14 days, as well as Vicryl Rapide (Ethicon / Johnson \u0026 Johnson MedTech), fast-absorbing polyglactin 910 with an equivalent degradation profile. None of these brands are cited as interchangeable 1:1 without prior verification: needle nomenclatures (shape, tip and attachment) vary by manufacturer.
There is no manufacturer-validated equivalence table cross-referencing this specific configuration (½ circle cylindrical taperpoint needle HR 22 mm, USP gauge 2/0, colourless thread 75 cm) with references from other commercial houses; consult the technical department before using in a tender. Any code from Safil Quick, Vicryl Rapide or another fast-absorbable brand must be cross-checked against its technical data sheet and IFU before proposing substitution. The brands cited belong to their respective manufacturers and are mentioned only as a category equivalence reference, not interchangeable 1:1.
5. Selection tips and variants of interest
To choose the correct suture, verify three parameters: USP gauge (this model, USP 2/0, is intended for thin to medium thickness soft tissues with short-term support), needle geometry (½ circle with 22 mm taperpoint conical tip for reduced fields and delicate tissues; cutting tips are reserved for firm tissues) and attachment (single strand of 75 cm; double strand offers greater yield per unit).
Variants of interest within the Cahnos catalogue in the same PGAR RAPID ½ circle cylindrical family: AM222HR2230 (HR 22, USP 3/0, colourless, 75 cm) for a finer gauge with the same needle; AM222HR2240 (HR 22, USP 4/0, colourless, 75 cm); AM222HR1730 (HR 17, USP 3/0, colourless, 75 cm) for a shorter needle; and AMP222HR1740 / AMP222HR1750 (Prime quality, HR 17, USP 4/0 and 5/0) for precision surgery. If the protocol requires longer support, contrast with medium-duration polyglactin 910 (PGLA, absorption 56-72 days).
As this is a sterile single-use product, check the integrity of the packaging and the expiry date before each procedure and store the suture in a cool, dry place, protected from sunlight. For special B2B terms by volume, hospital supply agreements or enquiries about equivalences from other commercial houses, contact the technical-commercial team at Cahnos.





Reviews