Description
1. Commercial Introduction
At Cahnos we offer the 100 % synthetic polyglycolic acid (PGA) multifilament absorbable suture, reference 118FR2710 (SKU AM118FR2710), designed for the professional who demands reliable, manageable and predictable performance in the operating theatre. This presentation combines a ⅝ circle needle with cylindrical body and Taperpoint conical point of 27 mm (FR 27), USP 0 gauge (metric 3.5) and a 70 cm violet fluted braided thread: maximum visibility on the surgical field and precision in delicate tissues.
It is the reference choice for hospitals, private clinics, general surgery operating theatres and purchasing groups seeking a high-performance synthetic absorbable thread with an excellent quality-price ratio. Each unit is supplied sterile, ready to use and single-use, in a box of 12 units with a minimum order of 30 boxes, ensuring continuity and traceability of supply for your healthcare facility.
2. Technical Explanation
Polyglycolic acid (PGA) is a 100 % absorbable synthetic polymer, structured as a braided multifilament with a fluted finish and a coating of polycaprolactone and calcium stearate. This configuration provides high flexibility, low memory and a low coefficient of friction, which favours smooth passage of the thread through tissue, easy handling and firm, secure knot tying.
Absorption occurs by hydrolysis, i.e. by degradation of the polymer in contact with organic fluids, over a predictable period of 60 to 90 days, maintaining apposition of the wound edges during the critical 2 to 4 weeks of healing while the tissue recovers its own strength. The ⅝ circle needle with cylindrical body and Taperpoint conical point penetrates by separating the fibres without a cutting effect, minimising trauma to delicate structures; it is therefore especially recommended for approximation of friable tissues, ligatures and general surgery. The fluted violet colour ensures optimal visibility of the thread on the surgical field. Product sterile by ethylene oxide (EO), single-use and not re-sterilisable.
3. Technical Specifications
3.1. Technical data sheet for reference 118FR2710
| Characteristic | Specification |
| Reference (SKU) | 118FR2710 (AM118FR2710) |
| Range / Quality | Premisut · Normal quality |
| Material | 100% polyglycolic acid (PGA), absorbable by hydrolysis |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 0 (metric 3.5) |
| Thread colour | Fluted violet (colourless version available in the family) |
| Thread length | 70 cm |
| Needle shape | ⅝ circle (5/8) |
| Needle body | Cylindrical |
| Needle tip | Conical (Taperpoint), atraumatic |
| Needle size | 27 mm (FR 27) |
| Absorption mechanism | Hydrolysis |
| Absorption period | 60 – 90 days |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Tissue reaction | Minimal |
| Indications | Approximation of soft tissues, ligatures and general surgery |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Complete absorption | 60 – 90 days by hydrolysis |
Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a fully predictable and controlled absorption rate.
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
| 0 | 3.5 | 0.35 – 0.399 mm (indicative according to Pharmacopoeia) |
3.4. Presentation and ordering conditions
| Value | Value |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
| Format | Sterile, single-use |
4. Equivalences and Cross-References
This 100 % multifilament polyglycolic acid (PGA) suture belongs to the family of synthetic braided absorbable sutures with medium absorption (60-90 days by hydrolysis). The technical cross-references verified in the manufacturer’s official catalogue for the configuration needle 5/8 taper point FR 27 / USP 0 gauge / 70 cm / violet are as follows:
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
| Resorba (Germany) | PGA-RESORBA | PA1015 (FR 27, taper point, USP 0, 70 cm, violet) | Comparable by category (not interchangeable 1:1) | Same polymer (coated braided PGA), same 5/8 taper point FR 27 needle, same USP 0 gauge, 70 cm and violet colour. They differ in presentation (24 units), coating (resolactona) and declared retention profile (approx. 50 % strength at 18 days). Code confirmed in the official Resorba supply programme (2025); verify against the official data sheet before use in tenders. |
| Assut Europe (Italy) | ASSUFIL (PGA) | FS43036 (FRG 27, reinforced taper point, USP 0, 75 cm, violet) | Comparable by category (not interchangeable 1:1) | Same coated braided multifilament PGA, violet, absorption 60-90 days. It matches the 5/8 taper point geometry of 27 mm (heavy reinforced version) and USP 0 gauge; the thread length (75 cm vs 70 cm) and presentation (box of 36) differ. Code verified in the official Assut Europe 2024 catalogue. |
By composition and mechanical behaviour, this suture is comparable at category level with the braided PGA lines on the market (Safil from B. Braun, Surgicryl PGA from SMI, DemeSORB PGA from DemeTECH, among others) and with polyglactin 910 (Vicryl / Vicryl Plus from Ethicon, Novosyn from B. Braun, Polysorb from Medtronic, Surgicryl 910 from SMI, Glicofil LAC from Assut). The reference codes of these manufacturers for the exact FR 27 / USP 0 / 70 cm configuration could not be conclusively verified in public sources with the same needle configuration, and are therefore not declared here. There is no equivalence table validated by the manufacturer for this cross-reference; consult the technical department before use in tenders.
5. Selection tips and variants of interest
USP 0 gauge (metric 3.5) offers an excellent balance between strength and delicacy: it is suitable for subcutaneous tissue, soft tissue, superficial muscle planes, tissue approximation and ligatures, with predictable absorption during a healing period of 2 to 4 weeks. The ⅝ circle needle with conical point (Taperpoint) of 27 mm provides a pronounced curve ideal for deep and restricted access spaces, recommended in delicate tissues where a cutting effect is to be avoided; for fibrous structures (sclera, dense fasciae) cutting or trocar points are recommended.
| Reference | Variant |
| AMP118FR2710 | PGA Prime USP 0, FR 27 mm needle, violet, 70 cm (superior quality, same configuration) |
| AM118FR2701 | PGA USP 1, FR 27 mm needle, violet, 70 cm (larger gauge) |
| AM118FR2720-2 | PGA USP 2/0, double needle FR 27 mm, violet, 70 cm (reduced access) |
| AM118FR3620 | PGA USP 2/0, FR 36 mm needle, violet, 70 cm (larger needle radius) |
| 1L18FR2710 | PGLA (polyglactin 910) USP 0, FR 27 mm needle, violet, 70 cm (absorption 56-72 days) |
Always check the needle radius (FR 27, FR 36 or FR 40 mm) according to the depth and surgical area to be treated; for greater tensile strength consider USP 0 or higher gauges, and if a longer strength retention is required (180-210 days), the polydioxanone (PDO) family is the recommended alternative. As this is a sterile single-use product, always check the integrity of the packaging before each procedure, respect the expiry date and store the suture in a cool, dry place, protected from sunlight. All references in the range are supplied in boxes of 12 units with a minimum order of 30 boxes, with scheduled deliveries for hospitals, clinics and purchasing groups. For special B2B conditions by volume or equivalents from other commercial manufacturers, contact our technical-commercial team.





Reviews