Description
1. Commercial Introduction
The absorbable multifilament polyglycolic acid (PGA) suture from the Premisult range, reference 112DS1950, is the trusted choice for the professional seeking a synthetic absorbable thread with predictable behaviour, excellent handling and maximum knot security in tissue approximation, ligatures and general surgery. Its ⅜ circle needle with 19 mm reverse cutting triangular point (DS 19) offers firm, controlled penetration in medium-consistency tissues, while the 75 cm fluted violet braided thread ensures optimal visibility on the surgical field and smooth sliding during suturing.
It is a synthetic absorbable medical device, sterile and single-use, with an excellent quality-price ratio for B2B procurement for operating theatres, purchasing centres and distributors of surgical supplies. Commercial presentation in a box of 12 units.
2. Technical Explanation
Polyglycolic acid (PGA) is a synthetic absorbable polymer presented in a braided multifilament structure with a fluted finish and a coating of polycaprolactone and calcium stearate. This configuration provides a regular surface that facilitates the passage of the thread through tissue, reduces the sawing effect, offers smooth handling and guarantees firm, secure knot tying, even in fine working calibres such as USP 5/0.
Absorption occurs by hydrolysis, i.e. by degradation of the polymer in contact with organic fluids, over a predictable period of 60 to 90 days and independent of the patient’s enzymatic activity. The tensile strength retention profile maintains the necessary support during the critical phase of wound healing, while the tissue recovers its own resistance. The ⅜ circle needle with 19 mm reverse cutting triangular point (reverse cutting) penetrates firm, non-compliant tissues with controlled trauma and avoiding tearing. The violet colour improves the visibility of the thread on the surgical field. The product is supplied sterile by ethylene oxide (EO), is single-use and is not resterilisable.
3. Technical Specifications
3.1. Technical data sheet for reference 112DS1950
| Parameter | Specification |
| Brand / Range | Premisult (standard quality) |
| Catalogue reference | 112DS1950 |
| SKU Cahnos | AM112DS1950 |
| Composition | 100 % polyglycolic acid (PGA) absorbable by hydrolysis |
| Coating | Polycaprolactone and calcium stearate |
| Thread structure | Braided coated multifilament, fluted finish |
| Gauge | USP 5/0 (metric 1) |
| Thread colour | Violet (also available undyed in the range) |
| Thread length | 75 cm |
| Needle shape | 3/8 circle (approx. 135°) |
| Tip type | Triangular with reverse cutting (inverted triangular / reverse cutting) |
| Needle size | DS 19 (19 mm) |
| Absorption mechanism | By hydrolysis, between 60 and 90 days |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Presentation | Box of 12 units |
3.2. Tensile strength retention profile after implantation
| Time point | Tensile strength retention |
| 2 weeks | 75% of initial strength |
| 3 weeks | 50% of initial strength |
| 4 weeks | 40% of initial strength |
| Complete absorption | 60-90 days by hydrolysis |
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
| 5/0 | 1 | 0.10 – 0.149 mm (indicative according to USP/EP) |
4. Equivalences and Cross-References
The coated braided multifilament polyglycolic acid (PGA) with medium absorption (60-90 days) is one of the most widely used synthetic absorbables in general surgery. The homologous families by composition (PGA homopolymer coated with polycaprolactone and calcium stearate), absorption profile, calibre, needle and colour that have been verified with reference codes in public sources are as follows. As they are other manufacturers, none is declared interchangeable 1:1 without validation by the technical service.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
| SMI (Surgical Materials Industries, Belgium) | Surgicryl® PGA (polyglycolic acid, coated braided multifilament) | 11101519 / SM11101519 (DS-19, 19 mm) | Comparable by category (not interchangeable 1:1) | Material matches (PGA coated with 1% polycaprolactone and calcium stearate), USP 5/0 calibre, violet colour, 75 cm thread and ⅜ circle reverse cutting needle DS19 of 19 mm, with absorption of 60 to 90 days. Configuration confirmed in SMI catalogues and authorised distributors; verify needle geometry before substituting. |
| DemeTECH (USA) | DemeSORB® PGA (glycolide homopolymer, braided coated) | PGA285019F4P (DFS-2, 19 mm) | Comparable by category (not interchangeable 1:1) | Material matches (glycolide homopolymer coated with polycaprolactone and calcium stearate), USP 5/0 calibre, violet colour, 75 cm thread (30″) and 3/8 circle reverse cutting needle DFS-2 of 19 mm. Official coding pattern confirmed in the DemeTECH store for 3-0 (PGA283019F4P) and 4-0 (PGA284019F4P); the 5-0/75 cm/violet configuration appears published by authorised distributors (MPN DT-421-1). Confirm the exact code for the 5/0 calibre before bidding. The retention profile differs (84 % at 2 weeks / 23 % at 4 weeks) compared to the Premisult profile. |
There is no equivalence table validated by the manufacturer for this specific configuration (multifilament PGA, ⅜ circle needle DS19 of 19 mm, reverse cutting triangular point, USP 5/0 calibre, 75 cm violet thread); consult the technical department of Cahnos before using in bidding. Before substituting this reference in a specification, check one by one: material, USP calibre, needle size (19 mm), ⅜ curvature, point family (reverse cutting), thread length (75 cm) and colour. Other PGA families with which it is advisable to compare technical data sheets: Safil® from B. Braun (pure polyglycolic acid, DS19 needle series in lengths of 45 and 70 cm, not 75 cm) and the PGA line from Laboratorio Aragó (Spain). Copolymer ranges of polyglactin 910 (PGLA), such as Vicryl® from Ethicon, are chemically different from PGA homopolymer and are not interchangeable 1:1 without validation by the technical department. The brands cited belong to their respective manufacturers and are mentioned solely as an equivalence reference.
5. Selection tips and variants of interest
The USP 5/0 (metric 1) gauge is indicated for fine tissues and low to moderate tension: approximation of soft tissues, skin closures, subcutaneous layers, microsurgery, and ligation of small-calibre vessels. The DS19 needle of 19 mm with reverse cutting triangular point is recommended when firm tissues need to be passed through without tearing in medium or small surgical fields. The violet colour enhances thread visibility during the procedure; if maximum aesthetic discretion is preferred, a colourless version of the same configuration is available.
| Reference | Variant |
| AM112DS1950 (112DS1950) | PGA USP 5/0, ⅜ triangular needle DS19 mm, violet, 75 cm (current product) |
| AM122DS1950 | PGA USP 5/0, same DS19 mm needle, colourless fluted, 75 cm (colourless variant) |
| AMP112DS1950 | PGA Prime quality USP 5/0, DS19 mm needle, violet, 75 cm (superior quality) |
| AM112DS1650 | PGA USP 5/0, DS16 mm needle, violet, 75 cm (shorter needle) |
| AM111DS1950 | PGA USP 5/0, DS19 mm needle, violet, 45 cm (shorter thread) |
| AM112DS1940 | PGA USP 4/0, DS19 mm needle, violet, 75 cm (inferior gauge) |
| AM112DS1930 | PGA USP 3/0, DS19 mm needle, violet, 75 cm (superior gauge) |
| 1L12DS1950 | PGLA (polyglactin 910) USP 5/0, DS19 mm needle, violet, 75 cm (absorption 56-72 days) |
All references are supplied in boxes of 12 units; the minimum order is 30 boxes to ensure continuous supply for your healthcare centre. As this is a sterile single-use product, always check the integrity of the packaging before each procedure, observe the expiry date and store in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or enquiry about additional gauges, lengths and colours, contact the technical-commercial team of Cahnos.





Reviews