Description
1. About this suture
The PGA Prime Multifilament Absorbable Suture from the PREMISUT range, tariff reference P128DS1940FN (SKU AMP128DS1940FN), is the reference solution for B2B supply to operating theatres, purchasing centres and healthcare distributors. It combines a synthetic absorbable polyglycolic acid (PGA) thread with hydrolysis-controlled absorption, exceptional handling and fully predictable clinical behaviour.
Mounted on a ⅜ circle needle with a reverse cutting triangular point (inverted triangle) of 19 mm (DS19 Fine), it has USP gauge 4/0 (metric 1.5) and a colourless, non-grooved thread of 70 cm. It is indicated for tissue approximation, ligatures and general surgery, and is supplied as a sterile single-use medical device in a box of 12 units.
2. Technical explanation
Polyglycolic acid (PGA) is a synthetic absorbable polymer widely used in surgical sutures, presented here in a braided multifilament structure coated with polycaprolactone and calcium stearate. This configuration facilitates the passage of the thread through tissue, reduces the sawing effect, offers smooth sliding and guarantees excellent knot security even in fine gauges such as USP 4/0 (approximate diameter 0.15-0.199 mm).
Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the polymer degrades into glycolic acid, a substance metabolised by the body without causing significant tissue reaction. The tensile strength retention profile maintains the necessary support during the critical phase of healing, while the tissue recovers its own resistance.
The ⅜ circle needle with a 19 mm reverse cutting triangular point (Fine model) penetrates firm, fibrous and non-compliant tissues with controlled trauma and without tearing, making it particularly suitable for skin closures, fascia, aponeurosis, ligatures and reduced surgical fields. The colourless colour minimises visual interference on the surgical field, an advantage appreciated in aesthetic and reconstructive surgery. Sterilisation by ethylene oxide (EO); does not allow re-sterilisation.
3. Technical Specifications
3.1. General characteristics of reference P128DS1940FN
| Parameter | Specification |
|---|---|
| Brand / Range | PREMISUT (Prime quality) |
| Catalogue reference | P128DS1940FN |
| SKU Cahnos | AMP128DS1940FN |
| Material | 100 % absorbable polyglycolic acid (PGA) |
| Thread structure | Braided multifilament, smooth finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 4/0 (metric 1.5) |
| Approximate diameter | 0.15 – 0.199 mm (USP/EP standard) |
| Thread colour | Colourless |
| Thread length | 70 cm |
| Needle shape | 3/8 circle (approx. 135°) |
| Tip type | Triangular with reverse cutting (inverted triangular / reverse cutting) Fine model |
| Needle size | DS19 (19 mm) Fine |
| Absorption mechanism | Hydrolysis (60-90 days) |
| Sterilisation | Ethylene oxide (EO), single-use, not re-sterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Regulatory standard | USP/EP |
| Presentation | Box of 12 units (minimum order 30 boxes) |
3.2. Tensile strength retention profile
| Period | Tensile strength retention |
|---|---|
| 2 weeks | 75% of initial strength |
| 3 weeks | 50% of initial strength |
| 4 weeks | 40% of initial strength |
| Complete absorption | 60-90 days by hydrolysis |
Excellent initial tensile strength, optimal knot tying and fully predictable absorption rate, with minimal tissue reaction.
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
|---|---|---|
| 4/0 | 1.5 | 0.15 – 0.199 mm (indicative according to Pharmacopoeia) |
4. Equivalences and Cross-References
Coated braided multifilament polyglycolic acid (PGA) is the most established medium-absorption synthetic absorbable in surgical practice (classic ‘Dexon’ type family). This Prime reference, with hydrolysis absorption of 60-90 days, tensile retention of 75/50/40% at 2/3/4 weeks and a ⅜ circle needle with a 19 mm reverse cutting triangular point (DS19 Fine), is comparable at category level (not 1:1 interchangeable) with the following homologous lines from other manufacturers, verified against official sources (manufacturer’s catalogue, regulated repertoires or authorised distributors):
| Manufacturer / Brand | Family / Range | Code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| B. Braun (Aesculap) | Safil® — pure polyglycolic acid, coated braided multifilament | C1049420 | Comparable by category (not interchangeable 1:1) | Safil colourless USP 4/0 (EP 1.5), 70 cm, DS19 reverse cutting needle. It matches in material, gauge, needle, length and colour; the coating differs (glyconate vs. polycaprolactone) and the commercial presentation (box of 36 vs. 12). Code verified in medical device repertoires and authorised distributors. |
| B. Braun (Aesculap) | Novosyn® — polyglactin 910, coated braided multifilament | C0069420 | Comparable by category (not interchangeable 1:1) | Novosyn colourless 4/0 (1.5), 70 cm, DS19. The material differs: polyglactin 910 (copolymer 90% glycolide / 10% L-lactide, absorption 56-72 days) vs. pure PGA with absorption of 60-90 days. Code verified in the official B. Braun catalogue (Aesculap). |
| SMI (Surgical Materials Industries) | Surgicryl® PGA — polyglycolic acid, braided multifilament coated with polycaprolactone and calcium stearate | 11151519U | Comparable by category (not interchangeable 1:1) | Surgicryl PGA USP 4/0 (1.5), colourless (undyed), DS19 reverse cutting, 75 cm. It matches in material, coating, gauge, needle and colour; the thread length differs (75 cm vs. 70 cm) and the commercial presentation. Code verified in the official SMI catalogue. |
| DemeTECH (USA) | DemeSORB® PGA — polyglycolic acid, braided coated with polycaprolactone and calcium stearate | PGA274019F4P (DFS-2, 19 mm) | Comparable by category (not interchangeable 1:1) | DemeSORB PGA USP 4/0, colourless (undyed), 3/8 circle reverse cutting needle DFS-2 of 19 mm, thread of 75 cm. It matches in material, coating and gauge; the length differs (75 cm vs. 70 cm) and the needle geometry (DFS-2 vs. DS19 Fine). Configuration confirmed in the official DemeTECH store and in GUDID/FDA (DI 30652927185288). |
There is no equivalence table validated by the manufacturer for this specific configuration (PGA multifilament Prime, ⅜ circle DS19 Fine reverse cutting needle, USP gauge 4/0, colourless thread of 70 cm); consult the technical department of Cahnos before use in tenders. Before substituting this reference in a specification, check one by one: material, USP gauge, needle size (19 mm), ⅜ curvature, point family (reverse cutting), thread length (70 cm) and colour. Copolymer ranges of polyglactin 910 (PGLA), such as Vicryl® from Ethicon or Novosyn® from B. Braun, are chemically different from PGA homopolymer and are not 1:1 interchangeable without validation from the technical department. The cited brands belong to their respective manufacturers and are mentioned only as an equivalence reference.
5. Selection tips and variants of interest
USP gauge 4/0 (metric 1.5) is indicated for tissues under moderate tension: skin, subcutaneous tissue, mucous membranes, fascia and standard healing planes. The ⅜ circle needle with reverse cutting triangular point is recommended when firm tissues need to be penetrated without tearing (skin, fascia, aponeurosis). The colourless thread is chosen when maximum aesthetic discretion is sought; if maximum visibility in the surgical field is preferred, the same configuration is available in violet.
| Reference | Variant |
|---|---|
| AMP128DS1940FN | PGA Prime, DS19 Fine, USP 4/0, colourless, 70 cm (current product) |
| AMP118DS1940FN | PGA Prime, DS19 Fine, USP 4/0, violet, 70 cm (maximum visibility) |
| AMP121DS1940FN | PGA Prime, DS19 Fine, USP 4/0, colourless, 45 cm (shorter length) |
| AMP111DS1940FN | PGA Prime, DS19 Fine, USP 4/0, violet, 45 cm |
| AMP122DS1940 | PGA Prime, DS19, USP 4/0, colourless, 75 cm (longer length) |
| AMP112DS1940 | PGA Prime, DS19, USP 4/0, violet, 75 cm |
All references are supplied in boxes of 12 units, with a minimum order of 30 boxes. As this is sterile single-use material, check the integrity of the packaging before each procedure and store it in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or enquiries about additional gauges, lengths and colours on request, contact the technical-commercial team at Cahnos.