Description
1. Commercial Introduction
The absorbable multifilament polyglycolic acid (PGA) suture reference 112DS3030 (SKU AM112DS3030), from the PREMISUT range in standard quality, is the professional choice for operating theatre services, purchasing centres and hospital distribution seeking a synthetic absorbable thread with predictable behaviour, excellent handling and maximum knot security. Its ⅜ circle needle with reverse cutting triangular point of 30 mm (DS 30) allows firm and controlled penetration in consistent tissues, while the USP 3/0 gauge offers a very versatile balance between initial strength and manageability. The 75 cm violet fluted thread guarantees optimal visibility over the surgical field and smooth sliding during suturing. It is a synthetic absorbable medical device, sterile and single-use, with an excellent quality-price ratio for B2B procurement.
2. Technical explanation
Polyglycolic acid (PGA) is a synthetic absorbable polymer presented in a braided multifilament structure with a fluted finish and a coating of polycaprolactone and calcium stearate. This configuration favours smooth passage of the thread through tissue, reduces the sawing effect, offers excellent manageability and guarantees great knot security, even in fine working gauges such as USP 3/0 (metric 2).
Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the polymer degrades into glycolic acid, a substance that the body metabolises without generating relevant tissue reaction. The tensile strength retention profile maintains the necessary support during the critical phase of healing while the tissue recovers its own resistance. The ⅜ circle needle with reverse cutting triangular point of 30 mm penetrates firm, fibrous and non-compliant tissues with controlled trauma and without tearing, which is especially suitable for skin closures, fascia, aponeurosis and ligatures. The violet colour improves visibility of the thread over the surgical field. The product is supplied sterile by ethylene oxide (EO), is single-use and does not allow re-sterilisation.
3. Technical Specifications
3.1. General characteristics of reference 112DS3030
| Brand / Range | PREMISUT (standard quality) |
| Catalogue reference | 112DS3030 |
| SKU Cahnos | AM112DS3030 |
| Material | 100 % synthetic absorbable polyglycolic acid (PGA) |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 3/0 (metric 2) |
| Thread colour | Fluted violet (also available colourless in the range) |
| Thread length | 75 cm |
| Needle shape | 3/8 circle (approx. 135°) |
| Tip type | Reverse cutting triangular (inverted triangular) |
| Needle size | DS 30 mm |
| Absorption mechanism | Hydrolysis, between 60 and 90 days |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Presentation | Box of 12 units |
3.2. Tensile strength retention profile
| Period | Tensile strength retention |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 40 % |
| Complete absorption | 60 – 90 days by hydrolysis |
Excellent initial tensile strength, optimal knot tying and fully predictable absorption rate, with minimal tissue reaction.
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
| 3/0 | 2 | 0.20 – 0.249 mm (indicative according to USP/EP Pharmacopoeia) |
4. Equivalences and Cross-References
Braided coated multifilament polyglycolic acid (PGA) with medium absorption (60-90 days) is one of the most widespread synthetic absorbables in the surgical market and corresponds to the homologous families of pure PGA from the main manufacturers. The following references have been cross-checked against the manufacturer’s official catalogue and regulatory database (GUDID/FDA):
| Manufacturer / Brand | Family / Range | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| SMI (Belgium) | Surgicryl PGA (polyglycolic acid, coated braided multifilament) | 11301530 (violet) | Comparable by category (not interchangeable 1:1) | Surgicryl PGA USP 3/0 with DS 30 needle of ⅜ circle and reverse cutting of 30 mm, violet thread of 75 cm, box of 12 units; it matches in material, coating (polycaprolactone and calcium stearate 1 %), gauge, needle, length and colour. Verified in SMI official data sheet; the definitive 1:1 validation corresponds to the technical department. |
| B. Braun (Aesculap) | Safil (pure polyglycolic acid, coated braided multifilament) | C1048246 (box of 12) / G1048246 | Comparable by category (not interchangeable 1:1) | Safil Violet USP 3/0 with ⅜ circle needle DS 30 (M) of 30 mm; the thread length differs (70 cm vs 75 cm) and the coating (polyglyconate vs polycaprolactone). Verified in GUDID/FDA and official catalogue of B. Braun. |
Other brands with homologous ranges of pure PGA or comparable glycolide copolymers by clinical category, without code verified in primary sources for this exact configuration: Ethicon / Johnson & Johnson (Vicryl, polyglactin 910), Medtronic / Covidien (Polysorb, Lactomer), DemeTECH (DemeSORB, PGA) and Laboratorio Aragó (PGA series). Polyglactins 910 (PGLA) are copolymers of glycolide and L-lactide chemically different from PGA homopolymer, so they should not be treated as interchangeable 1:1 without validation from the pharmacy service or the technical department.
There is no equivalence table validated by the manufacturer for this specific configuration (multifilament PGA, ⅜ circle needle DS 30 mm, reverse cutting triangular point, USP 3/0 gauge, 75 cm violet thread); consult the technical department before using in tenders. Before substituting this reference in a specification, check one by one: material, USP gauge, needle size (30 mm), ⅜ curvature, point family (reverse cutting), thread length (75 cm) and colour.
5. Selection tips and variants of interest
The USP 3/0 gauge (metric 2) offers an ideal balance between initial strength and manageability for tissues with moderate mechanical demand: approximation of soft tissues, skin closures, subcutaneous tissue and ligatures. The DS 30 mm needle with reverse cutting triangular point is indicated when firm tissues need to be penetrated without tearing (skin, fascia, aponeurosis), and the violet thread provides maximum visibility over the surgical field; if maximum aesthetic discretion is preferred, there is the colourless version of the same range.
| Reference | Variant of interest |
| AM112DS3030 | PGA standard quality, DS 30, USP 3/0, violet, 75 cm (current product) |
| AMP112DS3030 | PGA Prime, same configuration DS 30, USP 3/0, violet, 75 cm (higher range) |
| AM122DS3030 | PGA, DS 30, USP 3/0, colourless thread, 75 cm (same configuration in colourless) |
| AM112DS3020 | PGA, DS 30, USP 2/0, violet, 75 cm (thicker gauge) |
| AM112DS3010 | PGA, DS 30, USP 0, violet, 75 cm (higher initial strength) |
| AM112DS3620 | PGA, DS 36, USP 2/0, violet, 75 cm (longer needle) |
| 1L12DS3030 | PGLA (polyglactin 910), DS 30, USP 3/0, violet, 75 cm (absorption 56-72 days) |
All references are supplied in a box of 12 units. As this is sterile single-use material, check the integrity of the packaging before each procedure and store it in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or enquiries about other gauges, lengths and colours on request, contact our technical-commercial team at Cahnos.