Description
1. Commercial Introduction
The absorbable multifilament polyglycolic acid (PGA) suture, reference 113HRT4802 (SKU AM113HRT4802), is the reference choice in the operating theatre for hospitals, private clinics, purchasing groups and wholesalers seeking a high-performance synthetic absorbable thread with an excellent price-quality ratio in B2B procurement. It features a ½ circle needle with cylindrical body and trocar point (tapercut) of 48 mm (HRT48), USP 2 gauge (metric 5) and a 90 cm violet fluted thread, particularly indicated for tissue approximation, ligatures and general surgery.
Each unit is supplied sterile by ethylene oxide and ready to use, in a box of 12 units with batch traceability, with the professional distribution warranty of Cahnos.
2. Technical Explanation
Polyglycolic acid (PGA) is a 100% absorbable synthetic polymer presented in a braided multifilament structure with a fluted finish and a coating of polycaprolactone and calcium stearate. This configuration facilitates the passage of the thread through tissue, reduces friction and the sawing effect, offers smooth handling and guarantees firm and secure knot tying, even in high working gauges such as USP 2.
Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the polymer degrades into glycolic acid, a substance metabolised by the body without causing significant tissue reaction. The tensile strength retention profile maintains the necessary support during the critical healing phase (75% at 2 weeks, 50% at 3 and 40% at 4).
The ½ circle needle (approx. 180°) with cylindrical body and trocar point (tapercut) combines a rounded body that passes through tissue separating the fibres with minimal trauma and a contained cutting point that allows penetration of fibrous and dense tissues — fascia, aponeurosis, muscle planes — with less effort and without enlarging the exit hole. The violet colour ensures optimal visibility on the surgical field. Single-use product, not suitable for re-sterilisation.
3. Technical Specifications
3.1. General characteristics (Ref. 113HRT4802)
| Parameter | Specification |
|---|---|
| Factory reference | 113HRT4802 |
| Reference (SKU Cahnos) | AM113HRT4802 |
| Range / Quality | PREMISUT – Normal quality |
| Thread material | 100 % polyglycolic acid (PGA), synthetic absorbable |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 2 (metric 5 – EP) |
| Thread colour | Violet grooved |
| Thread length | 90 cm |
| Needle shape | ½ circle (approx. 180°) |
| Needle body | Cylindrical |
| Needle tip | Trocar (tapercut) |
| Needle size | HRT 48 (48 mm), 1 needle per suture |
| Absorption mechanism | Hydrolysis (60–90 days) |
| Sterilisation | Ethylene oxide (EO); single-use, non-reesterilizable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Presentation | Box of 12 sterile units (individual pouches) |
| Minimum order | 30 boxes |
| Regulatory standard | Conforms to USP/EP Pharmacopoeia for synthetic absorbable sutures |
3.2. Tensile strength retention profile
| Post-implantation time | Retained tensile strength |
|---|---|
| 2 weeks | 75% of initial strength |
| 3 weeks | 50% of initial strength |
| 4 weeks | 40% of initial strength |
| Complete absorption | 60–90 days by hydrolysis |
Predictable absorption by hydrolysis with minimal tissue reaction, excellent tensile strength and optimal knot tying, in accordance with USP / European Pharmacopoeia.
3.3. Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric measurement (EP) | Approximate thread diameter |
|---|---|---|
| 2/0 | 3 | 0.30 – 0.349 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
| 1 | 4 | 0.40 – 0.499 mm |
| 2 (this product) | 5 | 0.50 – 0.599 mm |
The thread diameter complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without exceeding its limits.
4. Equivalences and Cross-References
Polyglycolic acid (PGA) braided multifilament coated with medium absorption (60–90 days) is one of the most widely used synthetic absorbables in general surgery and ligatures. The specific configuration of this data sheet — PGA homopolymer, ½ circle HRT48 needle of 48 mm with trocar point (tapercut), USP 2 gauge, 90 cm violet thread — has been compared with the main homologous families on the market. The following external reference is the closest located and verified in terms of gauge, length and needle geometry:
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| Johnson & Johnson MedTech – Ethicon | Coated VICRYL® (polyglactin 910, braided coated, violet, TP-1 needle with conical point 48 mm, ½ circle, 90 cm) | J585G (GTIN 30705031467600) | Comparable by category (not interchangeable 1:1) | They match in USP 2 gauge, 90 cm length, violet colour, 48 mm ½ circle needle and braided coated structure. They differ in the polymer (polyglactin 910 copolymer versus PGA homopolymer), the absorption profile (56–70 days versus 60–90 days) and the needle point (conical TP-1 versus trocar/tapercut). Code confirmed in authorised distribution; consult the manufacturer’s technical data sheet before substituting in tenders. |
Other commercial houses with homologous families of medium-absorption braided absorbable suture — B. Braun with Safil® (coated braided PGA, violet; glyconate coating versus polycaprolactone and calcium stearate), SMI with Surgicryl® PGA or DemeTECH with DemeSORB® PGA — offer products of the same clinical category, but no confirmed code has been found in official sources for the exact configuration USP 2 + trocar point needle HRT 48 mm + 90 cm thread + violet. There is no equivalence table validated by the manufacturer that cross-references this specific reference with closed codes from other brands; consult the technical department before using in tenders. Before substituting this reference in a specification, check one by one: material, USP gauge, needle size (48 mm), ½ curvature, point family (trocar/tapercut), thread length (90 cm) and colour. The brands cited belong to their respective manufacturers and are mentioned only as comparative reference.
5. Selection tips and variants of interest
USP 2 gauge (metric 5) is indicated for tissues requiring high initial strength: deep planes, fascia, aponeurosis, high-tension closures and ligatures in general and digestive surgery. The ½ circle needle with trocar point of 48 mm is recommended for moderately dense or fibrous tissues and surgical fields with limited space; the conical point (taperpoint) is reserved for soft and loose tissues, and the blunt point for maximum atraumaticity. The 90 cm thread offers sufficient strand for medium-length continuous sutures and multiple ligatures.
| Reference (SKU) | Variant of interest |
|---|---|
| AM113HRT4802 | PGA USP 2, trocar needle HRT 48 mm, violet, 90 cm (current product) |
| AMP113HRT4802 | PGA Prime USP 2, same HRT 48 mm needle, violet, 90 cm (higher quality) |
| AM113HRT4820 | PGA USP 2/0, same HRT 48 mm trocar needle, violet, 90 cm (smaller gauge) |
| AM113HRT4020 | PGA USP 2/0, trocar needle HRT 40 mm, violet, 90 cm (shorter needle) |
| AM113HR4802 | PGA USP 2, conical needle HR 48 mm taperpoint, violet, 90 cm (atraumatic point for soft tissues) |
| 1L13HRT4820 | PGLA 910 USP 2/0, HRT 48 mm trocar needle, violet, 90 cm (absorption 56–72 days) |
All references are supplied in boxes of 12 units; the minimum order is 30 boxes to ensure continuous supply for your healthcare facility. As this is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and store in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or confirmation of equivalences for tenders, contact our technical-commercial team.





Reviews