Description
1. Commercial Introduction
The absorbable multifilament PGA Prime suture in USP gauge 4/0 is the medium-term closure solution demanded by general surgery, gynaecology, urology and ambulatory surgery departments: a synthetic, flexible braided thread with a fully predictable absorption profile, ideal for approximation of soft tissues, ligatures and closures in deep and superficial planes. This presentation incorporates a ½ circle needle with cylindrical body and trocar point (tapercut) of 17 mm (HRT17), undyed thread of 70 cm and USP gauge 4/0, in a box of 12 sterile units with batch traceability. An excellent option for B2B supply to hospitals, clinics and purchasing centres, with a minimum order of 30 boxes.
2. Technical Explanation
Polyglycolic acid (PGA) is a 100% absorbable synthetic polymer presented in a braided multifilament structure with a fluted finish and a coating of polycaprolactone and calcium stearate. This configuration regularises the thread surface: it promotes smooth passage through tissue, reduces sawing effect, offers excellent handling and guarantees high knot security even in fine working gauges such as USP 4/0.
Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the polymer degrades into glycolic acid, a substance that the body metabolises without generating significant tissue reaction, with absorption completed between 60 and 90 days. The thread retains the necessary mechanical support during the critical healing phase, maintaining 75% of initial tensile strength at 2 weeks, 50% at 3 weeks and 40% at 4 weeks.
The HRT17 needle features a ½ circle curvature with cylindrical body and trocar point (tapercut geometry): a point with a converging triangular section that combines firm penetration in moderately tense or slightly fibrous tissues with minimal tissue tearing when passing the thread. The undyed colour facilitates visualisation of the surgical field and its aesthetic use in superficial planes. The product is supplied sterile by ethylene oxide (EtO), is single-use and does not allow re-sterilisation.
3. Technical Specifications
3.1. Technical data sheet for reference 128HRT1740 (SKU AMP128HRT1740)
| Parameter | Specification |
|---|---|
| Factory reference / SKU | 128HRT1740 · AMP128HRT1740 |
| Range / quality | PGA Prime (premium quality) |
| Material | 100 % synthetic absorbable polyglycolic acid (PGA) |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 4/0 (metric 1.5), conforming to USP/EP Pharmacopoeia |
| Approximate diameter | 0.15 – 0.199 mm (indicative according to Pharmacopoeia) |
| Thread colour | Undyed (violet version available in the range) |
| Thread length | 70 cm, 1 needle per suture |
| Needle shape | ½ circle (1/2 circular) |
| Needle body | Cylindrical |
| Needle tip | Tapercut (converging triangular cross-section) |
| Needle size | HRT17 (17 mm) |
| Absorption mechanism | Hydrolysis, completed between 60 and 90 days |
| Sterilisation | Ethylene oxide (EtO); sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Commercial presentation | Box of 12 units; minimum order of 30 boxes |
| Regulatory standard | Gauges in accordance with USP and European Pharmacopoeia (EP) |
3.2. Tensile strength retention profile after implantation
| Period after implantation | Retained tensile strength |
|---|---|
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Total absorption | 60-90 days, by hydrolysis |
Complete absorption by hydrolysis between 60 and 90 days, with minimal tissue reaction, excellent initial tensile strength and optimal knot tying, with a fully predictable and controlled absorption rate.
3.3. Gauge equivalence according to USP/EP Pharmacopoeia (synthetic absorbable suture)
| USP size | Metric size | Approximate diameter |
|---|---|---|
| 4/0 (this product) | 1.5 | 0.15 – 0.199 mm (indicative according to Pharmacopoeia) |
| 3/0 | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.349 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
4. Equivalences and Cross-References
Coated braided multifilament polyglycolic acid (PGA) is the most widely used medium-term synthetic absorbable in the surgical market, with an absorption profile of 60 to 90 days and a coating of polycaprolactone and calcium stearate. For the exact configuration of this reference (HRT17 needle of ½ circle and trocar point of 17 mm, USP gauge 4/0, undyed thread of 70 cm) no verified catalogue correspondence has been found in primary source that reproduces point by point all its characteristics (gauge, undyed colour, curvature and needle size), therefore no exact 1:1 equivalence is declared against third-party brands.
There is no equivalence table validated by the manufacturer for this specific configuration; consult with the technical department before use in tendering. As homologous families on the market to be verified reference by reference on their official data sheet before substitution, the following lines are cited: 100% coated braided PGA —Safil (B. Braun), Surgicryl PGA (SMI), DemeSORB PGA (DemeTECH) and Atramat PGA (Internacional Farmacéutica)— and polyglactin 910 —Vicryl (Ethicon) and Polysorb (Medtronic)—, always checking USP gauge 4/0, needle size 17 mm in ½ circle, trocar point, thread length of 70 cm and undyed colour. The copolymer ranges of polyglactin 910 (PGLA) are chemically distinct from homopolymer PGA and should not be treated as 1:1 interchangeable without validation from the pharmacy service or technical department. Likewise, the Monosyn range (B. Braun, glycolide/caprolactone copolymer monofilament) shares the HRT17 needle and gauge, but is a monofilament of different composition and does not allow 1:1 cross-reference. The brands cited belong to their respective manufacturers and are mentioned only as comparative reference.
5. Selection tips and variants of interest
USP gauge 4/0 (metric 1.5) is suitable for fine and low-to-medium tension tissues: skin approximation, subcutaneous tissue, mucosa, microsurgery and ligatures of small-calibre vessels. The ½ circle trocar point (tapercut) HRT17 needle is recommended for firm or slightly fibrous tissues where clean penetration without tearing is required; if the tissue is soft or friable, consider the conical point (HR) variant from the same PGA family.
| Reference (SKU) | Variant of interest |
|---|---|
| AMP128HRT1740 | PGA Prime, HRT17 trocar point needle, USP 4/0, undyed, 70 cm (current product) |
| AMP118HRT1740 | PGA Prime, same HRT17 trocar point needle, USP 4/0, fluted violet, 70 cm (direct colour equivalent for greater visibility) |
| AMP118HRT1730 | PGA Prime, HRT17 trocar point needle, USP 3/0, violet, 70 cm (same grip, larger gauge) |
| AMP128HR1740 | PGA Prime, HR17 conical point needle, USP 4/0, undyed, 70 cm (atraumatic point for delicate tissues) |
| P1L28HRT1740 | Polyglactin 910 (PGLA) Prime, HRT17 trocar point, USP 4/0, undyed, 70 cm (absorption 56-72 days, different copolymer) |
| P1L18HRT1740 | Polyglactin 910 (PGLA) Prime, HRT17 trocar point, USP 4/0, violet, 70 cm (different copolymer, violet colour) |
All references are supplied in boxes of 12 units; the minimum order is 30 boxes to ensure continuous supply for your healthcare facility. As this is a sterile single-use product, check the integrity of the pouch and the expiry date before each procedure, handle the strand with gloves to avoid altering its mechanical properties, and keep the reference together with the batch number for traceability in the operating theatre. For special B2B conditions, volume orders or other gauges, lengths and colours made to order, contact the technical-commercial team at Cahnos.





Reviews