Description
1. Commercial Introduction
The absorbable multifilament polyglycolic acid (PGA) suture from the Premisut range, reference 118HR4502 (SKU AM118HR4502), is the reference surgical closure solution for hospitals, private clinics, general surgery operating theatres and purchasing centres seeking a high-performance synthetic absorbable thread with an excellent price-quality ratio in B2B procurement and tenders.
It incorporates a ½ circle needle with cylindrical body and Taperpoint conical point of 45 mm, USP 2 gauge and violet grooved thread of 70 cm: a combination designed for the approximation of soft tissues, deep planes, ligatures and general surgery. Each unit is supplied sterile and ready to use, in an individual sachet with batch traceability, in a box of 12 units.
2. Technical explanation
Polyglycolic acid (PGA) is a 100% absorbable synthetic polymer presented in a braided multifilament structure with a grooved finish and a coating of polycaprolactone and calcium stearate. This configuration provides a regular surface that facilitates the passage of the thread through tissue, reduces friction and the sawing effect, offers very smooth handling, low memory and guarantees firm and secure knot tying, even in high working gauges such as USP 2.
Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the polymer degrades into glycolic acid, a substance that the body metabolises without generating relevant tissue reaction. The ½ circle needle with Taperpoint conical point penetrates by separating the fibres without cutting them, minimising trauma to soft and friable tissues. The grooved violet colour ensures optimal visibility on the surgical field. It is a single-use product, sterile by ethylene oxide and with no possibility of re-sterilisation.
3. Technical Specifications
3.1. General characteristics (Ref. AM118HR4502)
| Parameter | Specification |
| Reference (SKU Cahnos) | AM118HR4502 |
| Factory reference | 118HR4502 |
| Range / Quality | Premisut – Normal quality |
| Thread material | 100 % polyglycolic acid (PGA), synthetic absorbable |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 2 (metric 5 – EP) |
| Thread colour | Violet grooved |
| Thread length | 70 cm (1 thread, 1 needle) |
| Needle shape | ½ circle (approx. 180°) |
| Needle body | Cylindrical |
| Needle tip | Conical (Taperpoint), tapered and atraumatic |
| Needle size | 45 mm (HR 45) |
| Absorption mechanism | Hydrolysis |
| Absorption period | 60 – 90 days |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Presentation | Box of 12 sterile units (individual pouches) |
| Minimum order | 30 units |
| Regulatory standard | Conforms to USP Pharmacopoeia / European Pharmacopoeia (EP) for synthetic absorbable sutures |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
| 2 weeks | 75% of initial strength |
| 3 weeks | 50% of initial strength |
| 4 weeks | 40% of initial strength |
| Total absorption | 60 – 90 days, by hydrolysis |
Predictable absorption by hydrolysis with minimal tissue reaction, excellent tensile strength and optimal knot tying, in accordance with the USP/European Pharmacopoeia. Absorption range and retention values copied literally from the manufacturer’s documentation, without extension of limits.
3.3. Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 1 | 4 | 0.40 – 0.499 mm |
| 2 (this product) | 5 | 0.50 – 0.599 mm |
4. Equivalences and Cross-References
Braided coated multifilament polyglycolic acid (PGA) with medium absorption (60–90 days) is one of the most widely used synthetic absorbables in general surgery and ligatures. This reference —PGA homopolymer, ½ circle Taperpoint conical needle HR 45 mm, USP 2 gauge and 70 cm violet thread— shares polymer, coating and clinical profile with the homologous 100% PGA lines on the market: Safil® (B. Braun), Surgicryl® PGA (SMI), DemeSORB® PGA (DemeTECH), AssuCryl® (Assut) and the historical Dexon®/Dexon II® range (Medtronic/Covidien).
The copolymeric polyglactin 910 (PGLA) ranges, such as Coated Vicryl® (Ethicon), Polysorb® (Medtronic) or Novosyn® (B. Braun), share the category of coated braided absorbable multifilament, but differ in chemical composition and absorption window (56–72 days versus 60–90 days): they are comparable at category level, not interchangeable 1:1 without validation from the pharmacy service or the technical department. No closed code from another brand has been found in official sources that matches field by field (USP 2 gauge + 45 mm conical needle + 70 cm violet thread + PGA homopolymer) with this exact configuration. There is no equivalence table validated by the manufacturer; consult the technical department before using in a tender. Before substituting this reference in a specification, check one by one: material, USP gauge, needle size (45 mm), ½ curvature, point family (conical/taperpoint), single needle attachment, thread length (70 cm) and colour. The cited brands belong to their respective manufacturers and are mentioned solely as comparative reference.
5. Selection tips and variants of interest
USP 2 gauge (metric 5) is indicated for tissues requiring high initial resistance: deep planes, fascia, aponeurosis, high-tension closures and ligatures in general and digestive surgery. The ½ circle needle with 45 mm conical point is recommended for soft, friable or lax tissues, where the atraumatic point separates the fibres without cutting them; if the tissue is dense or fibrous, consider the trocar point (tapercut), and for very firm structures with clean cutting, the reverse cutting triangular point (HS).
| Reference (SKU) | Variant of interest |
| AM118HR4502 | PGA USP 2, HR 45 mm conical needle, violet, 70 cm (current product) |
| AMP118HR4502 | PGA Prime quality, same HR 45 mm needle, USP 2, violet, 70 cm (premium range) |
| AM113HR4502 | PGA USP 2, HR 45 mm needle, violet, 90 cm (longer thread length) |
| AM118HR4002 | PGA USP 2, HR 40 mm needle, violet, 70 cm (shorter needle) |
| AM118HR4802 | PGA USP 2, HR 48 mm needle, violet, 70 cm (longer needle) |
| AM118HR4501 | PGA USP 1, same HR 45 mm needle, violet, 70 cm (smaller gauge) |
| 1L18HR4502 | Polyglactin 910 USP 2, HR 45 mm needle, violet, 70 cm (absorption 56–72 days) |
All references are supplied in boxes of 12 units; the minimum order is 30 units. For longer force retention needs (180–210 days), the polydioxanone (PDO) family is the recommended alternative. As this is sterile single-use material, check the integrity of the packaging and the expiry date before each procedure, and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions by volume, hospital supply contracts or confirmation of equivalences for tenders, contact our technical-commercial team.





Reviews