Description
1. Commercial Introduction
At Cahnos we make available to hospitals, clinics, operating theatres and purchasing centres the absorbable multifilament suture of polyglycolic acid (PGA) from the Prime range, reference AMP113HR3602. It is a high-performance synthetic absorbable surgical suture: manageable, safe, with fully predictable absorption and a very competitive quality-price ratio for institutional supply.
This presentation combines a ½ circle needle with cylindrical body and taper point of 36 mm (HR 36), a USP 2 gauge (metric 5) of high initial strength and a violet fluted braided thread of 90 cm with great visibility on the surgical field. Each unit is supplied sterile, ready to use and single-use, in a box of 12 units.
2. Technical Explanation
Polyglycolic acid (PGA) is a synthetic absorbable polymer presented in a braided multifilament structure with a fluted finish and a coating of polycaprolactone and calcium stearate. This architecture regularises the surface of the thread, favours sliding through tissue, reduces the saw effect and guarantees firm and secure knot tying even in larger diameter gauges such as USP 2.
Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity, completed between 60 and 90 days. This profile maintains the necessary tensile support during the first 2 to 4 weeks of healing, while the tissue recovers its own strength. The ½ circle cylindrical taper point needle penetrates by separating the fibres without cutting them, minimising trauma to delicate structures; it is therefore especially recommended for approximation of soft tissues, ligatures and deep planes. The violet colour guarantees optimal visibility. Product sterile by ethylene oxide (EO), single-use and not re-sterilisable.
3. Technical Specifications
3.1. Technical data sheet for reference AMP113HR3602
| Characteristic | Specification |
| Reference (SKU Cahnos) | AMP113HR3602 |
| Factory reference | P113HR3602 |
| Brand / quality | Premisut – Prime |
| Material | 100% polyglycolic acid (PGA), absorbable homopolymer |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 2 (metric 5), conforming to USP/EP Pharmacopoeia |
| Thread colour | Violet grooved |
| Thread length | 90 cm |
| Needle shape | ½ circle (1/2 circular) |
| Needle body | Cylindrical |
| Needle tip | Conical (Taperpoint) |
| Needle size | HR 36 (36 mm), 1 needle per thread |
| Absorption mechanism | Hydrolysis, completed between 60 and 90 days |
| Sterilisation | Ethylene oxide (EtO); sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Commercial presentation | Box of 12 units (individual sterile sachets) |
| Minimum order | 30 boxes |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Total absorption | 60-90 days, by hydrolysis (predictable rate) |
Predictable and controlled absorption profile, with minimal tissue reaction, excellent tensile strength and optimal knot tying.
3.3. Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 2 (this product) | 5 | 0.500 – 0.599 mm |
| 1 | 4 | 0.400 – 0.499 mm |
| 0 | 3.5 | 0.350 – 0.399 mm |
| 2/0 | 3 | 0.300 – 0.339 mm |
The thread diameter complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without exceeding their limits.
3.4. Decoding of the official reference
Official reference P113HR3602: 113 (violet thread 90 cm) + HR36 (½ circle needle with cylindrical body and taper point 36 mm) + 02 (USP 2 gauge). The official reference is marketed without altering characters: no digit may be added, removed or reordered when ordering or bidding.
4. Equivalences and Cross-References
Polyglycolic acid (PGA) is the most well-established medium-duration braided multifilament absorbable on the market and shares an absorption profile (60-90 days), coated braided structure and clinical behaviour with homologous families in the sector. There is no equivalence table validated by the manufacturer that cross-references this reference with other brands; the matches cited are functional comparisons at category level and should not be treated as 1:1 interchangeable without prior technical validation.
| Manufacturer / Brand | Family / Model | Equivalence Level | Technical Differences / Remarks |
| Ethicon (Johnson & Johnson) | Coated VICRYL® / VICRYL® Plus (braided polyglactin 910, violet) | Comparable by category (not interchangeable 1:1) | They match in braided structure, USP 2 gauge and indication; they differ in polymer (polyglactin 910, absorption 56-72 days, versus PGA 60-90 days) and retention at 4 weeks. |
| B. Braun | Safil® / Novosyn® (absorbable braided range) | Comparable by category (not interchangeable 1:1) | Same clinical indication and braided absorbable structure; verify exact material, gauge, size and needle curvature in the official data sheet. |
| Medtronic / Covidien | Polysorb® / Dexon™ II (absorbable braided range) | Comparable by category (not interchangeable 1:1) | Absorbable braided range of the segment (PGLA and PGA) with similar but not identical absorption and retention profiles; validate property by property. |
| SMI / Assut / DemeTECH | Surgicryl® PGA / Assucryl® / DemeSORB® PGA | Comparable by category (not interchangeable 1:1) | Homologous families of the braided PGA segment with medium absorption; review USP 2 gauge, HR 36 needle and 1 needle configuration in their technical data sheets. |
There is no equivalence table validated by the manufacturer; consult the technical department before use in tendering. Within the Premisut system itself from Cahnos, the HR36 / USP 2 / 90 cm geometry is fully interchangeable: reference AMP123HR3602 offers the same configuration in colourless thread and AM113HR3602 corresponds to the standard quality in violet; in the copolymer family, P1L13HR3602 (polyglactin 910) corresponds to the same format with absorption of 56-72 days.
5. Selection tips and variants of interest
USP 2 gauge (metric 5) is indicated for tissues requiring high initial strength: deep planes, fascia, aponeurosis, abdominal wall and high-tension closures, as well as ligatures. The ½ circle cylindrical taper point needle of 36 mm is atraumatic and ideal for soft and delicate tissues; if you need to penetrate firmer and denser tissues (skin, fascia, sclera), consider the reverse cutting (HS) or trocar (HRT) points available in the same range.
| Reference | Variant |
| AMP113HR3602 (P113HR3602) | PGA Prime USP 2, HR 36 mm needle, violet, 90 cm (current product) |
| AMP123HR3602 | PGA Prime USP 2, HR 36 mm needle, colourless, 90 cm (same geometry, maximum discretion) |
| AM113HR3602 (113HR3602) | PGA standard quality, HR 36 mm needle, violet, 90 cm (economical alternative) |
| AMP113HR3601 | PGA Prime USP 1, HR 36 mm needle, violet, 90 cm (smaller gauge) |
| AMP113HR3620 | PGA Prime USP 2/0, HR 36 mm needle, violet, 90 cm (fine gauge, delicate tissues) |
| AMP113HS3602 | PGA Prime USP 2, triangular reverse cutting HS 36 mm needle, violet, 90 cm (firm tissues) |
| P1L13HR3602 / 1L13HR3602 | PGLA polyglactin 910 USP 2, HR 36 mm needle, violet, 90 cm (absorption 56-72 days) |
All references are supplied in boxes of 12 units with a minimum order of 30 boxes to ensure continuous supply for your healthcare facility. If you require longer force retention (180-210 days), consider the polydioxanone (PDO) family from Cahnos. As this is a sterile single-use product, check the integrity of the pouch and the expiry date before each procedure, keep the reference together with the batch number for traceability in the operating theatre, and store in a cool, dry place, protected from sunlight. For special B2B conditions by volume or hospital supply agreements, contact the technical-commercial team at Cahnos.