Description
The absorbable multifilament PGA suture from the PREMISUT range, reference 118HR2640, is the resorbable closure solution your centre needs for soft tissue approximation, ligatures and general surgery: a synthetic braided thread with predictable behaviour, excellent handling, high knot security and a very competitive price-performance ratio for B2B procurement for operating theatres, clinics, purchasing centres and surgical material distributors. Each unit is supplied sterile, ready to use and single-use, in a box of 12 units with batch traceability.
Technical product explanation
Polyglycolic acid (PGA) is a 100% absorbable synthetic homopolymer presented in a braided multifilament structure with a grooved finish and a coating of polycaprolactone and calcium stearate. This architecture regularises the surface of the strand, reduces friction when passing through tissues (less sawing effect), provides smooth handling, low memory and guarantees firm and secure knot tying, even in fine gauges such as USP 4/0. Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity, completed between 60 and 90 days: during the first weeks of healing, the thread maintains the necessary tensile support while the tissue recovers its own resistance. The ½ circle needle with cylindrical body and 26 mm Taperpoint conical point penetrates by separating the fibres without cutting effect: it is the most atraumatic configuration for soft and friable tissues, with minimal damage along the puncture path. The violet colour offers high visibility of the strand on the surgical field. Product sterile by ethylene oxide, single-use and non-reesterilizable.
Technical specifications
| Parameter | Specification |
| Factory reference | 118HR2640 (PREMISUT range) |
| SKU Cahnos | AM118HR2640 |
| Thread material | 100 % polyglycolic acid (PGA), synthetic absorbable |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 4/0 (metric 1.5), conforming to USP/EP Pharmacopoeia |
| Approximate thread diameter | 0.15 – 0.199 mm |
| Thread colour | Violet grooved |
| Thread length | 70 cm (1 needle per suture) |
| Needle shape | ½ circle (1/2 circular) |
| Needle body | Cylindrical |
| Needle tip | Conical (Taperpoint), atraumatic |
| Needle size | HR 26 mm (26 mm) |
| Absorption mechanism | Hydrolysis, completed between 60 and 90 days |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Tissue reaction | Minimal |
| Indications | Approximation of soft tissues, ligatures and general surgery |
| Commercial presentation | Box of 12 units (individual sterile sachets) |
| Minimum order | 30 boxes |
Tensile strength retention profile after implantation
| Period after implantation | Retained tensile strength |
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Total absorption | 60 – 90 days, by hydrolysis (predictable rate) |
Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 4/0 (this product) | 1.5 | 0.15 – 0.199 mm |
| 3/0 | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
Equivalences and cross-references
The braided and coated multifilament polyglycolic acid (PGA) is the most widely used medium-duration synthetic absorbable in the surgical market, with an absorption profile of 60 to 90 days and a coating of polycaprolactone and calcium stearate, identical to that of this reference. For the USP 4/0 configuration, ½ circle cylindrical needle with conical point of 26 mm and violet strand of 70 cm, the following correspondence has been verified against the manufacturer’s official catalogue and authorised distribution sources:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| B. Braun (Aesculap) | Safil® Violet – coated braided multifilament PGA, USP 4/0 (metric 1.5), 70 cm, HR26 cylindrical ½ circle needle | C1048040 | Comparable by category (not interchangeable 1:1) | Confirmed in the official B. Braun catalogue and authorised distribution (Safil Violet 4/0 (1.5) 70 cm HR26): material (PGA), braided and coated multifilament structure, USP 4/0 gauge, violet colour, 70 cm length and ½ circle cylindrical needle of 26 mm all match. The coating (B. Braun’s own commercial impregnation) and commercial presentation differ. Validate in the technical data sheet/IFU before substituting in a tender. |
There is no equivalence table validated by the manufacturer that cross-references this reference with other brands; consult the technical department of Cahnos before using in a tender. The indicated equivalences are functional category comparisons, not 1:1 interchangeable: before substituting this reference in a specification, each one must be verified individually in the corresponding manufacturer’s technical data sheet: material, USP gauge, needle size (26 mm), ½ curvature, point family (Taperpoint conical), thread length (70 cm), colour and coating. Other homologous PGA families to verify case by case in their official data sheet: SMI Surgicryl® PGA, DemeTECH DemeSORB® PGA, Assut Assucryl® PGA, Atramat® / Resorba® PGA. Copolymer poliglactin 910 (PGLA) ranges, such as Coated VICRYL® (Ethicon), Safil® 910 / Novosyn® (B. Braun) or Polysorb® (Medtronic), are chemically different from PGA homopolymer (absorption of 56 to 72 days versus 60-90 days) and should not be treated as 1:1 interchangeable without validation from the pharmacy service or technical department. The cited brands belong to their respective manufacturers and are mentioned only as comparative reference.
Selection tips and variants of interest
The USP 4/0 gauge (metric 1.5) is the standard for fine and low-tension tissues: skin, mucosa, subcutaneous tissue and delicate planes where aesthetic healing and minimal marking are sought, as well as soft tissue approximation and ligation of fine-calibre vessels. The HR 26 ½ circle needle with Taperpoint conical point is recommended for soft or friable tissues where an atraumatic passage without cutting effect is desired; if the tissue is firm (skin, fascia, aponeurosis), consider the reverse cutting triangular point variant (HS) of the same family, and if you prefer a smaller needle radius, the HR 22 mm version. The Prime version (non-grooved finish) and VisiBlack (maximum contrast needle) are available on request. All references are supplied in boxes of 12 units; the minimum order is 30 boxes to guarantee continuous supply for your healthcare centre. As it is a sterile single-use product, always check the integrity of the pouch and the expiry date before each procedure, keep reference 118HR2640 together with the batch number for traceability in the operating theatre and store in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or validation of equivalences with other commercial houses, contact the technical-commercial team of Cahnos.





Reviews