Description
1. Commercial Introduction
The absorbable multifilament suture of polyglycolic acid (PGA) in Prime quality from the Premisut range is the trusted choice for the professional seeking a synthetic absorbable thread with fully predictable behaviour, excellent handling and maximum knot security. This reference mounts a ½ circle needle with cylindrical body and conical Taperpoint point of 36 mm (HR 36), USP 0 gauge and violet grooved thread of 90 cm, a highly versatile configuration for the approximation of soft tissues, ligatures and general surgery.
It is a synthetic, absorbable, sterile and single-use medical device, with an excellent quality-price ratio for B2B supply to operating theatres, clinics, purchasing centres and distributors of surgical material. Commercial presentation in a box of 12 units with a minimum order of 30 units.
2. Technical Explanation
Polyglycolic acid (PGA) is a 100% absorbable synthetic polymer presented in a braided multifilament structure with a grooved finish and a coating of polycaprolactone and calcium stearate. This configuration regularises the surface of the thread, favours sliding through tissue, reduces the sawing effect and guarantees firm and secure knot tying, even in medium working gauges such as USP 0 (metric 3.5).
Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the polymer degrades into glycolic acid, which the body metabolises without generating relevant tissue reaction, completing between 60 and 90 days. This profile maintains the necessary tensile strength during the first weeks of healing (75% at 2 weeks, 50% at 3 and 40% at 4), while the tissue recovers its own resistance.
The HR 36 needle has a ½ circle curvature with a cylindrical body and conical point (Taperpoint geometry): it passes through tissues separating the fibres without cutting effect, which makes it the most atraumatic option for soft, friable and well-irrigated tissues, with minimal injury along the penetration path. The grooved violet colour guarantees optimal visibility on the surgical field. The product is supplied sterile by ethylene oxide (EO), is single-use and does not allow re-sterilisation.
3. Technical Specifications
3.1. Technical data sheet for reference P113HR3610 (SKU AMP113HR3610)
| Parameter | Specification |
|---|---|
| Factory reference / SKU | P113HR3610 · AMP113HR3610 |
| Range / quality | Premisut – Prime (premium range) |
| Material | 100 % synthetic absorbable polyglycolic acid (PGA) |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 0 (metric 3.5), conforming to USP/EP Pharmacopoeia |
| Approximate diameter | 0.35 – 0.399 mm (indicative according to Pharmacopoeia) |
| Thread colour | Fluted violet (colourless version available in the range) |
| Thread length | 90 cm, 1 needle per thread |
| Needle shape | ½ circle (1/2 circular) |
| Needle body | Cylindrical |
| Needle tip | Conical (Taperpoint), atraumatic |
| Needle size | HR 36 mm (36 mm) |
| Absorption mechanism | Hydrolysis, completed between 60 and 90 days |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Tissue reaction | Minimal |
| Indications | Approximation of soft tissues, ligatures and general surgery |
| Commercial presentation | Box of 12 units; minimum order of 30 units |
| Regulatory standard | USP / European Pharmacopoeia (EP) for synthetic absorbable sutures |
3.2. Tensile strength retention profile after implantation
| Period after implantation | Retained tensile strength |
|---|---|
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Total absorption | 60 – 90 days, by hydrolysis |
Complete absorption by hydrolysis between 60 and 90 days, with minimal tissue reaction, excellent initial tensile strength and optimal knot tying, with a fully predictable and controlled absorption rate.
3.3. Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric size | Approximate diameter |
|---|---|---|
| 0 (this product) | 3.5 | 0.35 – 0.399 mm |
| 2/0 | 3 | 0.30 – 0.349 mm |
| 1 | 4 | 0.40 – 0.499 mm |
4. Equivalences and Cross-References
Polyglycolic acid (PGA) braided coated multifilament of medium absorption (60–90 days) is one of the most widely used synthetic absorbables in general surgery and ligatures (classic family type «Dexon»). Having verified the documentation of the main homologous manufacturers, no official equivalence table from the manufacturer has been found for a 1:1 cross-reference with the exact configuration of this reference (HR 36 mm Taperpoint needle, USP 0 gauge, 90 cm violet thread); the reference cited below is a catalogue comparison confirmed against authorised distribution, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Ethicon (Johnson & Johnson MedTech) | VICRYL PLUS (polyglactin 910 braided coated with triclosan), violet, USP 0, 90 cm, CT-1 needle of ½ circle conical point of 36 mm | VCP346H | Comparable by category (not interchangeable 1:1) | Confirmed in catalogue and authorised distribution: USP 0 gauge, violet colour, 90 cm length, ½ circle curvature, conical point family and needle size (CT-1 ≈ 36 mm) coincide; the polymer (polyglactin 910 vs 100% PGA), the presence of antibacterial triclosan and the absorption profile (56–72 days vs 60–90 days) differ. |
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. As homologous market families to verify reference by reference in their official data sheet before substitution, the following lines are cited: 100% braided coated PGA —Safil (B. Braun), Surgicryl PGA (SMI), DemeSORB PGA (DemeTECH), Assucryl PGA (Assut), Atramat PGA (Internacional Farmacéutica) and the historical Dexon (limited availability)— and those of polyglactin 910 —Vicryl / Vicryl Plus (Ethicon), Polysorb (Medtronic), Novosyn (B. Braun)—, always checking USP 0 gauge, 36 mm needle size in ½ circle, conical point, 90 cm thread length, colour and configuration (single needle). The copolymer ranges of polyglactin 910 (PGLA) are chemically different from PGA homopolymer and should not be treated as interchangeable 1:1 without validation from the pharmacy service or the technical department. The brands cited belong to their respective manufacturers and are mentioned only as comparative reference.
5. Selection tips and variants of interest
USP 0 gauge (metric 3.5) offers an ideal balance between initial strength and handling for medium-thickness tissues with moderate tension: subcutaneous tissue, fascia, mucosa, approximation of soft tissues and ligatures of medium-calibre vessels in general surgery. The ½ circle conical point Taperpoint HR 36 needle is the appropriate choice for soft, lax or friable tissues where an atraumatic passage without cutting effect is sought; if the tissue is firm or fibrous (fascia, aponeurosis), consider the trocar point (Tapercut) HRT 36 variant of the same family, and for very dense planes, the reverse cutting triangular point (HS).
| Reference (SKU) | Variant of interest |
|---|---|
| AMP113HR3610 | PGA Prime, HR 36 mm conical point needle, USP 0, violet, 90 cm (current product) |
| AM113HRT3620 | Standard PGA, HRT 36 mm trocar point needle, USP 2/0, violet, 90 cm (firm tissues) |
| AM113HR4010 | Standard PGA, HR 40 mm conical point needle, USP 0, violet, 90 cm (larger radius needle) |
| AMP113HR4001 | PGA Prime, HR 40 mm conical point needle, USP 1, violet, 90 cm (larger gauge) |
| AMP113HR4501 | PGA Prime, HR 45 mm taper point needle, USP 1, violet, 90 cm (deeper fields) |
| AMP113HR4801 | PGA Prime, HR 48 mm taper point needle, USP 1, violet, 90 cm (larger needle radius) |
All references are supplied in boxes of 12 units; the minimum order is 30 units. As this is a sterile single-use product, check the integrity of the pouch and the expiry date before each procedure, handle the strand with gloves to avoid altering its mechanical properties, and keep the reference together with the batch number for traceability in the operating theatre. For special B2B conditions, volume orders or other gauges, lengths and colours made to order, contact the technical-commercial team at Cahnos.