Description
1. Commercial Introduction
The absorbable multifilament PGA Prime suture from the Premisut range is the surgical closure solution your centre needs when a synthetic resorbable thread with fully predictable behaviour, excellent handling and maximum knot security is required. It combines a ½ circle cylindrical body needle with a blunt point of 48 mm (HRN 48) and a 90 cm violet fluted braided thread in USP 1 gauge, designed for the approximation of soft tissues, ligatures and medium-high tension closures in general surgery.
It is a synthetic absorbable medical device, sterile and single-use, presented in B2B format in a box of 12 units with a minimum order of 30 boxes, which guarantees continuous and traceable supply for hospitals, clinics, operating theatres, distributors and purchasing centres.
2. Technical Explanation
Polyglycolic acid (PGA) is a synthetic absorbable polymer composed of 100% glycolide units, presented in a braided multifilament structure with a fluted finish and coated with polycaprolactone and calcium stearate. This configuration provides great flexibility, low shape memory and a contained friction coefficient that favours the sliding of the thread through tissue, manageability and firm, secure knot tying, with high initial tensile strength.
Absorption occurs by hydrolysis, that is, by degradation of the polymer in contact with organic fluids, over a predictable period of 60 to 90 days, independent of the patient’s enzymatic activity and with minimal tissue reaction. The thread retains 75% of tensile strength at 2 weeks, 50% at 3 and 40% at 4, an ideal profile for tissues requiring support during 2 to 4 weeks of healing.
The ½ circle needle with cylindrical body and blunt point of 48 mm passes through tissues separating the fibres without cutting effect, which minimises trauma to delicate and friable structures. The violet colour of the thread guarantees optimal visibility on the surgical field. Sterilisation by ethylene oxide; sterile single-use product, with no possibility of re-sterilisation.
3. Technical Specifications
3.1. Technical data sheet
| Characteristic | Specification |
|---|---|
| Reference (SKU) | AMP113HRN4801 |
| Range / Quality | Premisut, Prime quality |
| Material | 100% polyglycolic acid (PGA), absorbable by hydrolysis, with polycaprolactone and calcium stearate coating |
| Thread structure | Braided multifilament with grooved finish |
| Thread colour | Violet grooved |
| Gauge / size | USP 1 (metric 4), conforming to USP/EP Pharmacopoeia |
| Thread length | 90 cm, 1 needle per thread |
| Needle shape | ½ circle (1/2 circular) |
| Needle body | Cylindrical |
| Needle tip | Roma (blunt point) |
| Needle reference | HRN 48 mm, Prime quality |
| Needle size | 48 mm |
| Absorption mechanism | Hydrolysis, completed between 60 and 90 days |
| Sterilisation | Ethylene oxide (EtO); sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal, promoting comfortable recovery |
| Indications | Tissue approximation, ligatures and general surgery |
| Commercial presentation | Box of 12 units; minimum order of 30 boxes |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
|---|---|
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Total absorption | 60-90 days, by hydrolysis |
Excellent resistance during the critical healing period, smooth passage through tissues, uniform and atraumatic surface, knot security and comfortable handling.
4. Equivalences and Cross-References
Braided polyglycolic acid (PGA) with 60-90 days of absorption is one of the most proven synthetic absorbable multifilament suture raw materials of medium absorption on the international market. There is no equivalence table validated by the manufacturer for a 1:1 cross-reference between brands in the exact configuration of this reference (½ circle needle, cylindrical body, blunt point HRN 48 mm, USP 1 gauge, 90 cm violet thread); consult the technical department of Cahnos before use in tenders.
At family level, the homologous reference lines in the sector are those of 100% braided coated PGA —Safil (B. Braun), Dexon II / Dexon (Medtronic/Covidien), Surgicryl PGA (SMI) and DemeSORB PGA (DemeTECH)— and those of poliglactin 910 copolymer —Vicryl / Vicryl Plus (Ethicon), Polysorb (Medtronic) and Novosyn (B. Braun)—, always checking one by one the polymer and coating, USP 1 gauge, the configuration (single needle), the ½ circle curvature, the point family (blunt, blunt point), the 48 mm needle size and the 90 cm thread length, since small variations in radius or point profile prevent direct equivalence. These comparisons are at category level, not interchangeable 1:1; the cited brands belong to their respective manufacturers and are mentioned only as comparative reference.
5. Selection tips and variants of interest
USP 1 gauge (metric 4) is indicated for tissues that support moderate-high tension: fascia, aponeurosis, muscle planes, abdominal wall and ligatures of medium-calibre vessels, with predictable absorption during a healing period of 2 to 4 weeks. The blunt point (HRN) is the most atraumatic option for soft and friable tissues; if you need greater ease of penetration in somewhat denser tissues, consider the conical point (HR) versions of the same PGA family, and for fibrous structures the trocar point (HRT).
Within the Cahnos catalogue, the reference that shares the complete configuration in poliglactin 910 is P1L13HRN4801 (PGLA Prime, same HRN 48 mm needle, USP 1, violet, 90 cm), which differs only in the polymer and in the absorption period (56-72 days). As variants of the same PGA Prime family with blunt point: AMP113HRN4810 (USP 0, same needle), AMP117HRN6501 (USP 1, HRN 65 mm needle, 100 cm thread), AMP113HRN4001H (Prime Heavy, USP 1, HRN 40 mm needle) and AM113HRN4801 (same configuration in normal quality). If the procedure requires a longer wound support (180-210 days), the polydioxanone (PDO) family is the recommended alternative. Always check the integrity of the pouch and the expiry date before each intervention and keep the reference together with the batch number for traceability in the operating theatre. For special B2B conditions by volume, hospital supply contracts or equivalences with other commercial houses, contact the technical-commercial team of Cahnos.