Description
1. Commercial Introduction
The Absorbable Multifilament PGA Prime Suture, reference AMP113HRT4802 (factory ref. P113HRT4802), is the reference choice in the operating theatre for surgical teams, hospitals, private clinics, purchasing groups and wholesalers seeking a high-performance synthetic absorbable thread with excellent value for money. It is presented with a ½ circle needle, cylindrical body and trocar point (tapercut) of 48 mm (HRT48), USP 2 gauge and a 90 cm grooved violet thread.
Each unit is supplied sterile by ethylene oxide and ready to use, with the highest quality of the Prime range and the professional distribution guarantee of Cahnos. Especially indicated for tissue approximation, ligatures and general surgery, where precision, manageability and fully predictable absorption are required during the first weeks of healing.
2. Technical Explanation
Polyglycolic acid (PGA) is a synthetic absorbable polymer presented in a braided multifilament structure with a grooved finish and a coating of polycaprolactone and calcium stearate. This configuration provides a regular surface that facilitates the passage of the thread through tissue, reduces the sawing effect, offers smooth manageability and guarantees firm and secure knot tying, even in high working gauges such as USP 2.
Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the polymer degrades into glycolic acid, a substance that the body metabolises without generating significant tissue reaction. The tensile strength retention profile maintains the necessary support during the critical healing phase (75 % at 2 weeks, 50 % at 3 weeks and 40 % at 4 weeks), while the tissue recovers its own resistance.
The ½ circle needle with cylindrical body and trocar point (tapercut) combines a rounded body that passes through tissue with minimal injury and a cutting point that allows penetration of fibrous and dense tissues with less effort, without enlarging the exit hole. The violet colour of the thread guarantees optimal visibility over the surgical field. Single-use product, with no possibility of re-sterilisation.
3. Technical Specifications
3.1. General characteristics (Ref. AMP113HRT4802)
| Parameter | Specification |
| Reference (SKU Cahnos) | AMP113HRT4802 |
| Factory reference | P113HRT4802 |
| Range / Quality | PREMISUT – Prime (premium range) |
| Thread material | 100 % polyglycolic acid (PGA), synthetic absorbable |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 2 (metric 5) |
| Thread colour | Violet |
| Thread length | 90 cm |
| Needle shape | ½ circle (approx. 180°) |
| Needle body | Cylindrical |
| Needle tip | Trocar (tapercut) |
| Needle size | 48 mm (HRT 48 Prime), 1 needle per thread |
| Absorption mechanism | By hydrolysis |
| Absorption period | 60 – 90 days |
| Sterilisation | Ethylene oxide (EO); single-use, not re-sterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Presentation | Box of 12 sterile units |
| Minimum order | 30 boxes |
3.2. Tensile strength retention profile
| Post-implantation time | Retained tensile strength |
| 2 weeks | 75% of initial strength |
| 3 weeks | 50% of initial strength |
| 4 weeks | 40% of initial strength |
| Complete absorption | 60 – 90 days by hydrolysis |
Predictable absorption by hydrolysis with minimal tissue reaction, excellent tensile strength and optimal knot tying, in accordance with USP / European Pharmacopoeia.
4. Equivalences and Cross-References
Braided coated multifilament polyglycolic acid (PGA) with medium absorption (60–90 days) is one of the most widely used synthetic absorbables in general surgery and ligatures. The specific configuration of this datasheet —PGA homopolymer, ½ circle HRT48 needle of 48 mm with trocar point (tapercut), USP 2 gauge, 90 cm violet thread— has been compared with the main homologous families on the market:
- Ethicon / Johnson & Johnson MedTech, Coated VICRYL® (polyglactin 910): copolymer range with a shorter absorption profile (56–72 days) and different chemistry from PGA homopolymer; not interchangeable 1:1 without technical validation.
- B. Braun / Aesculap, Safil®: line of coated braided PGA with a range of ½ circle needles; the coating differs (glyconate versus polycaprolactone and calcium stearate), so it is not interchangeable 1:1.
- SMI – Surgical Materials Industries, Surgicryl® PGA: homologous family of braided PGA coated with polycaprolactone and calcium stearate, with comparable clinical behaviour; the specific code for the combination USP 2 + ½ trocar needle 48 mm + 90 cm is pending confirmation in the manufacturer’s official catalogue.
- DemeTECH, DemeSORB® PGA: homologous family of coated braided PGA with ½ circle needles; the code for this exact configuration (USP 2 + trocar 48 mm + 90 cm) has not been confirmed in official sources.
There is no manufacturer-validated equivalence table that cross-references this specific reference (PGA Prime USP 2, ½ circle HRT48 needle of 48 mm with trocar point, 90 cm violet thread) with closed codes from other brands; consult the technical department before use in tenders. Before substituting this reference in a specification, check one by one: material, USP gauge, needle size (48 mm), ½ curvature, point family (trocar/tapercut), thread length (90 cm) and colour. The cited brands belong to their respective manufacturers and are mentioned solely as comparative reference.
5. Selection tips and variants of interest
USP 2 gauge (metric 5) is indicated for tissues requiring high initial resistance: deep planes, fascia, aponeurosis, high-tension closures and ligatures in general and digestive surgery. The ½ circle needle with 48 mm trocar point is recommended for moderately dense or fibrous tissues and surgical fields with limited space; the conical point (taperpoint) is reserved for soft and loose tissues. The 90 cm thread offers sufficient strand for medium-length continuous sutures and multiple ligatures.
| Reference | Variant of interest |
| AMP113HRT4802 | PGA Prime USP 2, ½ trocar needle HRT48 mm, violet, 90 cm (current product) |
| AMP113HRT4820 | PGA Prime USP 2/0, same HRT48 mm trocar needle, violet, 90 cm (smaller gauge) |
| AMP113HRT4020 | PGA Prime USP 2/0, ½ trocar needle HRT40 mm, violet, 90 cm (shorter needle) |
| AMP113HR4802 | PGA Prime USP 2, ½ conical taperpoint needle HR48 mm, violet, 90 cm (atraumatic point) |
| P1L13HRT4820 | PGLA (polyglactin 910) Prime USP 2/0, HRT48 mm trocar needle, violet, 90 cm (absorption 56–72 days) |
All references are supplied in boxes of 12 units; the minimum order is 30 boxes to ensure continuous supply for your healthcare facility. As this is a sterile single-use product, always check the integrity of the packaging before each procedure, respect the expiry date and store it in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or enquiries about additional gauges, lengths and colours, contact our technical-commercial team.





Reviews