Description
Absorbable Polydioxanone Prime suture with straight taper point needle of 19 mm
The Premisut Prime suture in Polydioxanone monofilament (PDO/PDS) is the reliable choice for the operating theatre when prolonged wound support is needed with predictable resorption and minimal tissue reaction. This 5/0 gauge format with a straight cylindrical needle with a taper point of 19 mm and violet thread of 30 cm is designed for surgeons who value excellent thread handling, clean passage through tissue and constant mechanical behaviour over weeks.
Technical product explanation
Polydioxanone is a synthetic polymer of poly(p-dioxanone) that degrades by hydrolysis, without the need for enzymatic action, ensuring uniform and predictable mass loss even in tissues with low metabolic activity. Being a monofilament, its smooth surface minimises the dragging of microorganisms through the tissue and reduces friction during suture passage. The straight needle with cylindrical body and taper point (Taperpoint) passes through the tissue separating the fibres without cutting them, ideal for soft tissues, fasciae, microsurgery and approximations where minimal trauma is desired. The tension retention profile (75% at 2 weeks, 60-70% at 4 and 45% at 8) and complete absorption in 180-210 days make it a prolonged absorption alternative compared to other short- or medium-profile synthetic absorbable sutures.
Technical specifications
| Parameter | Specification |
|---|---|
| Brand / line | Premisut Prime |
| Composition | Polydioxanone (PDO / PDS), synthetic monofilament of poly(p-dioxanone) |
| USP / metric gauge (EP) | 5/0 (EP 1) |
| Thread length | 30 cm |
| Thread colour | Violet |
| Needle type | Straight (straight cut), cylindrical body, taper point (Taperpoint), not grooved |
| Needle size | GR 19 mm |
| Absorption type | By hydrolysis, over a period of 180-210 days |
| Retained tensile strength | 75 % at 2 weeks; 60-70 % at 4 weeks; 45 % at 8 weeks |
| Sterilisation | Ethylene oxide |
| Clinical behaviour | Minimal tissue reaction, excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures, general surgery, paediatric cardiovascular tissues and ophthalmology |
| Presentation | Box of 12 units · Minimum order 30 boxes |
| References | P414GR1950 / AMP414GR1950 |
Equivalences and cross-references
Polydioxanone is a market standard in long-lasting absorbable sutures. Below is the cross-reference verified against public documentation from the reference manufacturer:
| Manufacturer / Brand | Family / Model | Reference code | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Ethicon (Johnson & Johnson MedTech) | PDS II (polydioxanone) | BA2932H | Comparable by category (not interchangeable 1:1) | Same chemistry (violet monofilament polydioxanone), USP gauge 5/0 and straight cylindrical taper point needle of 19 mm (ST-4). Difference: 35 cm thread versus 30 cm for PDO Prime; complete absorption per Ethicon data sheet in 182-238 days versus 180-210 days. |
Likewise, B. Braun offers its homologous line MonoPlus in violet polydioxanone with straight GR19 needles of 30 cm, although availability in gauge 5/0 with that assembly must be confirmed with the distributor (references verified in catalogue: 4/0 and 3/0 GR19 30 cm). There is no equivalence table validated by the manufacturer covering the rest of the brands on the market; consult the technical department before using any unverified substitution in a tender.
Selection tips and variants of interest
For procedures requiring prolonged absorption with a straight needle, this 5/0 and 30 cm format is suitable for soft tissue approximations, ligatures and microsurgery work. If bilateral suture support is required, consult the double straight taper point needle variant (2xGR19). For more resistant tissues or deep abdominal wall planes, consider higher gauges from the same family (4/0, 3/0, 2/0) with straight or semicircular needles depending on the accessibility of the surgical bed; and if the tissue is very friable, the Taperpoint conical point presented here is preferable to cutting points. Always check the USP gauge, thread length and needle type before replacing a reference in your stock.





Reviews