Description
1. Commercial presentation
The absorbable monofilament polydioxanone (PDO/PDS) suture from the Prime range is the reference technical choice when the tissue needs prolonged support during healing and fully predictable reabsorption. This presentation combines a 75 cm violet grooved thread in USP 2/0 gauge with a ⅜ circle needle with a reverse cutting triangular point of 22 mm (DS22), a geometry that passes through fibrous tissue with a clean and controlled passage.
It is the B2B supply solution for operating theatres, general surgery departments, paediatric cardiovascular surgery units, ophthalmology services and hospital purchasing centres seeking quality homogeneity with CE marking. Synthetic absorbable medical device, sterile and single-use, presented in a box of 12 units.
2. Technical Explanation
Polydioxanone is a synthetic absorbable polyester —a poly(p-dioxanone) polymer— that degrades by hydrolysis, a mechanism independent of the patient’s enzymatic activity that guarantees a highly reproducible absorption kinetics. Being a monofilament, the thread has no wicking effect or capillarity: it reduces bacterial adherence, facilitates sliding through tissue and minimises the inflammatory response compared with multifilament threads. Its firm feel and excellent knotting capacity make it suitable for ligatures and sutures of planes subjected to sustained tension during the critical healing phase.
The ⅜ circle needle (approximate curvature of 135°) with a reverse cutting triangular point concentrates the cutting edge on the outer face of the curve, avoiding tearing of the superficial tissue. It is especially suitable for skin, subcutaneous tissue, aponeurosis, fascia and fibrous planes. The grooved violet colour improves visibility on the surgical field and knot holding. It is supplied sterile by ethylene oxide, single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. General characteristics
| Material | Polydioxanone (PDO/PDS), absorbable synthetic monofilament polymer |
| Thread structure | Fluted monofilament (not braided) |
| Quality / range | Prime (range PREMISUT) |
| Gauge / size | USP 2/0 (metric 3) |
| Approximate diameter | 0.30 – 0.349 mm (indicative according to Pharmacopoeia) |
| Thread colour | High-visibility violet, fluted |
| Thread length | 75 cm |
| Swaged | 1 needle per pack (single strand) |
| Needle shape | 3/8 circle (approx. 135°) |
| Needle body and point | Triangular, reverse cutting point, grooved |
| Needle size | DS22 Prime – 22 mm |
| Absorption mechanism | Hydrolysis (180–210 days) |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures, fascia closure, general surgery, paediatric cardiovascular tissues and ophthalmology |
| Presentation | Box of 12 sterile units |
| Factory reference / SKU | P412DS2220 / AMP412DS2220 |
3.2. Tensile strength retention profile
| Period after implantation | Tensile strength retention |
| 2 weeks | 75 % |
| 4 weeks | 60–70 % |
| 8 weeks | 45 % |
| Complete absorption | 180–210 days by hydrolysis, with minimal tissue reaction |
This prolonged strength retention profile is maintained during the critical healing phase of slow-healing tissues, which is why it is recommended for use in closures of fascia, aponeurosis and structures subjected to sustained traction.
4. Equivalences and Cross-References
Monofilament polydioxanone is the material basis of the market reference lines for long-lasting absorbable sutures (PDS-type family). Based on the verified configuration of this data sheet (⅜ circle reverse cutting triangular needle of 22 mm, USP 2/0 gauge, 75 cm violet thread), the following commercial reference from another manufacturer has been cross-checked in primary sources (official manufacturer catalogue):
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| B. Braun / Aesculap | MonoPlus® (polydioxanone) uncoated monofilament, violet | C0024990 (USP 2/0 / EP 3, 70 cm, violet, DS24 ⅜ circle reverse cutting body needle 24 mm) | Comparable by category (not interchangeable 1:1) | Same PDO monofilament material, same USP 2/0 gauge (EP 3), same violet colour and same DS needle family of ⅜ circle reverse cutting. Differences: needle size (DS24 of 24 mm vs DS22 of 22 mm) and thread length (70 cm vs 75 cm). Confirmed in the official catalogue of B. Braun. |
There is no equivalence table validated by the manufacturer that officially cross-references this product with the commercial ranges of other brands; consult the technical department before use in tenders. As homologous market families, to be verified code by code in their official technical data sheet before any substitution, the following are cited: the PDS II® and PDS Plus® lines from Ethicon (the latter with antibacterial triclosan coating, indicated only where its use is approved), the MonoPlus® line from B. Braun/Aesculap, the Surgicryl® PDO line from SMI and the Maxon line from Medtronic/Covidien (made of polyglyconate, a different polymer, so its absorption profile must be verified). All must be checked field by field in their official data sheet: USP gauge, thread length, colour, curvature, body, point and needle size, as well as the attachment. The cited brands belong to their respective manufacturers and are mentioned only as comparative reference.
5. Selection tips and variants of interest
When choosing a PDO suture, always check the USP gauge according to the tissue plane: fine gauges (4/0 to 6/0) for skin and subcutaneous tissue, 3/0 and 2/0 for general soft tissue —as is the case with this reference— and USP 0 and above for fascia and planes under tension. Needle geometry is also decisive: the reverse cutting triangular point (DS/HS) is indicated for fibrous and inelastic tissues (skin, aponeurosis, fascia), while the conical point (HR/SH) is reserved for delicate and well-perfused tissues (vessels, parenchyma).
Within the Cahnos catalogue, the PDO/PDS family of ⅜ circle with reverse cutting triangular point includes variants of interest:
| Reference | Variant of interest |
| AM412DS2220 | Standard quality PDO, same DS22 configuration, USP 2/0, violet, 75 cm (range alternative) |
| AMP412DS2420 | PDO Prime, DS24 needle of 24 mm, USP 2/0, violet, 75 cm (longer needle travel) |
| AMP412DS2620 | PDO Prime, DS26 needle of 26 mm, USP 2/0, violet, 75 cm (longer needle, same gauge) |
| AMP412DS2430 | PDO Prime, DS24 needle of 24 mm, USP 3/0, violet, 75 cm (finer gauge) |
| AMP412DR2020 | PDO Prime, ⅜ circle DR20 needle with tapered point, USP 2/0, violet, 75 cm (tapered point for delicate tissues) |
As this is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or other gauges, lengths and colours on request, contact the technical-commercial team at Cahnos.