Description
1. Product Presentation
The Prime line absorbable monofilament polydioxanone (PDO/PDS) suture is the reference technical choice when the tissue needs prolonged support during healing with fully predictable resorption and minimal tissue reaction. This presentation combines a 75 cm violet grooved thread in USP 2/0 gauge with a ⅜ circle needle, triangular body and reverse cutting point of 26 mm (DS26), a geometry designed to pass through fibrous tissue with a clean and controlled passage. Each pack includes 2 threads with needle, which optimises performance per unit in the operating theatre.
Available in a box of 12 sterile units, it is a reliable and cost-effective solution for general surgery departments, paediatric cardiovascular surgery operating theatres, ophthalmology units and hospital purchasing centres seeking quality consistency with CE marking.
2. Technical Explanation
Polydioxanone is a synthetic absorbable polyester —a polymer of poly(p-dioxanone)— that degrades by hydrolysis, a mechanism independent of the patient’s enzymatic activity that guarantees a highly reproducible absorption kinetics between individuals. Being a monofilament, the thread has no wick effect or capillarity, which reduces bacterial adherence, facilitates sliding through tissue and minimises the inflammatory response compared with multifilaments. Its firm feel and excellent knotting capacity make it suitable for ligatures and sutures of planes subjected to sustained tension during the critical healing phase.
The ⅜ circle needle with reverse cutting triangular point has the cutting edge on the outer face of the curve, which prevents tearing of superficial tissue and makes it especially suitable for skin, subcutaneous tissue, aponeurosis and other fibrous planes. With a length of 26 mm (DS26) and USP 2/0 gauge (metric 3), this configuration offers an excellent balance between strength and delicacy. The grooved violet colour improves thread visibility on the surgical field and knot security. It is supplied sterile by ethylene oxide, single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. General characteristics
| Parameter | Specification |
| Material | Polydioxanone (PDO/PDS), absorbable synthetic monofilament polymer |
| Thread structure | Grooved monofilament |
| Quality | Prime |
| Gauge / size | USP 2/0 (metric 3) |
| Thread colour | Violet (high visibility) |
| Thread length | 75 cm per thread (2 x 75 cm per pack) |
| Needle shape | ⅜ circle |
| Needle body and point | Triangular, reverse cutting point, grooved |
| Needle size | DS26 – 26 mm |
| Swaged | 2 needles per pack (double thread) |
| Absorption mechanism | Hydrolysis (180 – 210 days) |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures, general surgery, fascia closure, paediatric cardiovascular tissues and ophthalmology |
| Presentation | Box of 12 sterile units (individual pack) |
| Factory reference / SKU | P412DS2620.2 / AMP412DS2620.2 |
3.2. Tensile strength retention profile
| Time | Tensile strength retention |
| 2 weeks | 75 % |
| 4 weeks | 60 – 70 % |
| 8 weeks | 45 % |
| Complete absorption | 180 – 210 days by hydrolysis |
This force retention profile is maintained during the critical healing phase of slow-healing tissues, which recommends its use in closures of fascia, aponeurosis and structures subjected to sustained traction.
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
| 2/0 | 3 | 0.30 – 0.339 mm (indicative according to Pharmacopoeia) |
4. Equivalences and Cross-References
Monofilament polydioxanone is the material basis of the market reference lines for long-lasting absorbable sutures (PDS-type family). Based on the verified configuration of this data sheet (⅜ circle reverse cutting triangular needle of 26 mm, USP 2/0 gauge, violet thread of 2 x 75 cm), commercial references from other manufacturers verified in primary sources are cited below.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | PDS II® (polydioxanone) violet monofilament | Z317H (USP 2/0, 70 cm, violet, SH needle 26 mm ½ circle taper point) | Comparable by category (not interchangeable 1:1) | Same PDO monofilament material, same USP 2/0 gauge, same violet colour and same needle size (26 mm). Differences: curvature (½ circle SH vs ⅜ circle DS26), point (tapered vs reverse cutting triangular), length (70 cm vs 2 x 75 cm double thread) and presentation (box of 36 units). Confirmed in official catalogues and authorised distributors. |
| B. Braun / Aesculap | MonoPlus® (polydioxanone) uncoated monofilament, violet | C0024086 (USP 2/0 / EP 3, 70 cm, violet, HR26S needle ½ circle) | Comparable by category (not interchangeable 1:1) | Same PDO monofilament material, same USP 2/0 gauge (EP 3), same violet colour and same needle size (26 mm). Differences: curvature (½ circle HR26S vs ⅜ circle DS26), point family (tapered vs reverse cutting) and length (70 cm vs 2 x 75 cm double thread). Confirmed in official catalogue B. Braun. |
There is no equivalence table validated by the manufacturer that officially cross-references this product with the commercial series of other brands; consult the technical department before use in tenders. As homologous market families to verify code by code before substitution: the PDS II® and PDS Plus® lines from Ethicon (the latter with antibacterial triclosan coating, indicated only where its use is approved), the MonoPlus® line from B. Braun/Aesculap, the Surgicryl® PDO line from SMI and the Maxon line from Medtronic/Covidien. All must be checked in their official technical data sheet (USP gauge, thread length, colour, curvature, body, point and needle size). The cited brands belong to their respective manufacturers and are mentioned only for comparative reference.
5. Selection tips and variants of interest
When choosing a PDO suture, always check the USP gauge according to the tissue plane: fine gauges (4/0 to 6/0) for skin and subcutaneous tissue, 3/0 and 2/0 for general soft tissue —as is the case with this reference— and USP 0 and above for fascia and planes under tension. Needle geometry is also decisive: the reverse cutting triangular point (DS/HS) is indicated for fibrous and inelastic tissues (skin, aponeurosis), while the tapered point (HR/SH) is reserved for delicate and well-perfused tissues (vessels, parenchyma). The double thread format (2 x 75 cm) is especially efficient in prolonged continuous sutures and in procedures where both threads per pack are used.
Within the Cahnos catalogue, the PDO/PDS family of ⅜ circle with reverse cutting triangular point includes variants of interest: AMP412DS3010.2 (DS30 needle of 30 mm, USP 0, 2 x 75 cm) for a larger gauge while maintaining the double thread; AMP418HS2620 (½ circle HS26 mm needle, USP 2/0, 70 cm) if the protocol prefers a half-circumference curvature; and AM41-2.7020 (USP 2/0, without needle, 2 x 70 cm) if loose thread is required for ligatures with your own needle or thread passers. Always verify the USP gauge, the measurement and the geometry of the needle against your surgical protocol before substituting one reference for another.
As this is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or other gauges, lengths and colours made to order, contact the technical-commercial team at Cahnos.





Reviews