Description
1. Product Presentation
The absorbable monofilament polydioxanone (PDO/PDS) suture from the Prime range, reference AMP418HR1730, is the professional solution for the surgeon who needs a thread with prolonged absorption, excellent handling and maximum knot security during the critical phase of healing. It incorporates a ½ circle needle with cylindrical body and conical Taperpoint tip of 17 mm (HR17), grooved violet thread of 75 cm and USP gauge 3/0.
It is a synthetic absorbable medical device, sterile and single-use, with a very competitive price-quality ratio for B2B supply to operating theatres, general surgery departments, paediatric cardiovascular surgery units, ophthalmology services and purchasing centres. Commercial presentation in a box of 12 units.
2. Technical Explanation
Polydioxanone is a synthetic absorbable polymer (poly(p-dioxanone)) presented in a monofilament structure, which gives it a smooth surface that reduces friction when passing through tissue, minimises the sawing effect and hinders bacterial adhesion as there are no interstitial spaces. Its degradation occurs by hydrolysis, i.e., by scission of the polymer chains in contact with organic fluids, over a predictable period of 180 to 210 days, independent of the patient’s enzymatic activity.
This prolonged absorption profile maintains the necessary tensile strength during the critical phases of healing, being especially useful in tissues that require prolonged support. The ½ circle needle with cylindrical body and conical Taperpoint tip (HR17) of 17 mm passes through tissue atraumatically, separating fibres without cutting effect: it is the reference configuration for delicate and friable tissues, with controlled passage and minimal damage along the penetration path. The grooved violet thread ensures optimal visibility on the surgical field. The product is sterilised by ethylene oxide (EO), is single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. Technical data sheet for reference AMP418HR1730
| Parameter | Specification |
| Reference (SKU) | AMP418HR1730 |
| Factory reference | P418HR1730 |
| Quality / Brand | Prime |
| Suture type | Synthetic absorbable monofilament |
| Material | 100 % polydioxanone (PDO/PDS), poly(p-dioxanone) polymer |
| Thread structure | Smooth, grooved monofilament |
| USP / metric gauge | 3/0 (metric 2) |
| Thread diameter | 0.20 – 0.249 mm (USP/EP standard) |
| Thread colour | Violet grooved |
| Thread length | 75 cm (one needle) |
| Needle shape | ½ circle |
| Needle body | Cylindrical |
| Needle tip | Conical Taperpoint, atraumatic |
| Needle size | 17 mm (HR17), one needle |
| Absorption mechanism | Hydrolysis |
| Total absorption | 180–210 days |
| Sterilisation | Ethylene oxide (EO), sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures, general surgery, paediatric cardiovascular tissues and ophthalmology |
| Presentation | Box of 12 units |
| Regulatory standard | USP/EP |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
| 2 weeks | 75% of initial strength |
| 4 weeks | 60–70% of the initial strength |
| 8 weeks | 45% of initial strength |
| Total absorption | 180–210 days, by hydrolysis |
Predictable and controlled absorption profile, with minimal tissue reaction, excellent tensile strength and optimal knot tying. Price per box of 12 units.
4. Equivalences and Cross-References
Polydioxanone monofilament is the most well-established synthetic absorbable suture material with prolonged absorption on the market (PDS-type family). The USP 3/0 configuration, violet thread of 75 cm and ½ circle needle with conical tip of 17 mm corresponds to international reference lines; however, the presentation in a box of 12 units and the finish of the strand are specific to this reference. Below is indicated only the reference verified as comparable at category level, not interchangeable point by point:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | PDS II (polydioxanone monofilament, USP 3/0, violet, RB-1 ½ circle taper point needle of 17 mm) | Z305H | Comparable by category (not interchangeable 1:1) | Confirmed in official Ethicon data sheet (EPC) and authorised distributors: polymer, gauge, colour, shape and needle size (17 mm) match; they differ in strand length (70 cm vs 75 cm), tip finish (conventional taper point vs Taperpoint) and presentation (36 units vs 12 units). |
There is no equivalence table validated by the manufacturer for an exact 1:1 cross-reference of the USP 3/0 / HR17 conical tip needle / 75 cm configuration; consult the technical department before use in tenders. As category equivalents (same indication and same polymer, without code verified in primary source), the violet monofilament PDO/PDS suture families from other laboratories are cited, such as the MonoPlus line from B. Braun; in all cases, the USP gauge, needle size and single-needle configuration must be verified in the manufacturer’s technical data sheet before substituting.
5. Selection tips and variants of interest
The USP 3/0 gauge is suitable for tissues with moderate tension: skin, subcutaneous tissue, mucosa, fascia, muscle and ligation of medium-fine calibre vessels. The 17 mm conical tip needle is recommended for soft, well-perfused tissues where a fully atraumatic passage is sought, such as subcutaneous tissue, peritoneum, paediatric cardiovascular approximations and ophthalmology; if a controlled cutting tip is preferred for denser or fibrous tissues, consider the trocar tip (HRT) versions of the same family.
| Reference | Description |
| AM418HR1730 | PDO Normal quality, HR17 needle of 17 mm, USP 3/0, violet, 70 cm (economical alternative, same configuration) |
| AMP418HR1730-2 | PDO Prime, double needle 2xHR17 of 17 mm, USP 3/0, violet, 2×70 cm (continuous suture with anchoring at both ends) |
| AMP418HR1330 | PDO Prime, HR13 needle of 13 mm, USP 3/0, violet, 70 cm (shorter needle, superficial planes) |
| AMP418HR1740 | PDO Prime, HR17 needle of 17 mm, USP 4/0, violet, 70 cm (finer gauge, greater precision) |
| AMP418HRT1730 | PDO Prime, HRT17 needle of 17 mm trocar tip (tapercut), USP 3/0, violet, 70 cm (controlled cutting tip) |
| AMP418HRP1730 | PDO Prime, HRP17 needle of 17 mm conical Taperpoint Ultra Plus tip, USP 3/0, violet, 70 cm (premium tip finish) |
All references are supplied in a box of 12 units. Keep the original packaging in a cool, dry place, respect the expiry date and handle the strand with gloves to avoid altering its mechanical properties. For special B2B conditions by volume or equivalents from other commercial houses, contact our technical-commercial team.