Description
1. Product Presentation
The absorbable monofilament polydioxanone (PDO/PDS) suture in Prime quality (Ref. AMP412HR3010) is the long-lasting synthetic reabsorbable solution your operating theatre needs when the tissue requires prolonged tensile support during healing. Manufactured with next-generation polydioxanone —a poly(p-dioxanone) polymer—, it offers predictable behaviour, excellent initial tensile strength, smooth passage through tissue and great security in knot tying.
This presentation incorporates an atraumatic ½ circle needle, cylindrical body and Taperpoint conical point of 30 mm (HR30), USP 0 gauge (metric 3.5), violet grooved thread of 75 cm and Prime quality, in professional box format of 12 units, designed for stock management in operating theatres, hospitals and purchasing centres.
2. Technical explanation of the material and procedure
Polydioxanone (PDO) is a synthetic absorbable polymer presented in monofilament structure: a single smooth strand without bacterial anchoring zones or wicking effect, with very low memory, excellent handleability and smooth sliding through tissue. As it has no coating, it maintains a uniform surface that reduces drag and minimises tissue reaction, being the reference configuration when prolonged support is required in tissues subjected to sustained tension.
Its absorption occurs by hydrolysis, a uniform process independent of the patient’s enzymatic activity, with total absorption between 180 and 210 days. The ½ circle needle with conical point (Taperpoint geometry) of 30 mm separates the fibres without cutting effect: it is the most atraumatic configuration for delicate and friable tissues, with controlled passage and minimal injury along the penetration path. The violet colour guarantees optimal visibility on the surgical field. The product is sterilised by ethylene oxide (EO), is single-use and does not allow re-sterilisation; always check the integrity of the packaging before each procedure.
3. Technical Specifications
3.1. Technical data sheet for reference AMP412HR3010
| Characteristic | Specification |
|---|---|
| Reference (SKU) | AMP412HR3010 |
| Factory reference | P412HR3010 |
| Quality | Prime |
| Suture type | Synthetic absorbable monofilament |
| Composition | Polydioxanone (PDO/PDS), polymer of poly(p-dioxanone) |
| Thread structure | Grooved monofilament |
| USP / metric gauge | 0 / 3.5 |
| Thread colour | Violet (high visibility) |
| Thread length | 75 cm |
| Needle shape | ½ circle (half circumference) |
| Needle body | Cylindrical |
| Needle tip | Conical Taperpoint, atraumatic |
| Needle size | 30 mm (HR30 Prime) |
| Absorption mechanism | Hydrolysis |
| Total absorption | 180–210 days |
| Sterilisation | Ethylene oxide (EO), sterile, single-use |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures, general surgery, paediatric cardiovascular and ophthalmology |
| Presentation | Box of 12 units |
| Regulatory standard | USP/EP |
3.2. Tensile strength retention and absorption profile
| Period after implantation | Retained tensile strength |
|---|---|
| 2 weeks | 75 % of initial tensile strength |
| 4 weeks | 60–70 % of the initial tensile strength |
| 8 weeks | 45 % of the initial tensile strength |
| Total absorption | 180–210 days, by hydrolysis |
Predictable and controlled absorption profile, with minimal tissue reaction, excellent tensile strength and optimal knot tying. Price per box of 12 units.
4. Equivalences and Cross-References
Monofilament polydioxanone is the most well-established long-lasting absorbable raw material in surgical practice (classic ‘PDS’ type family), with hydrolysis absorption of 180–210 days and tensile strength retention of 75/60–70/45% at 2/4/8 weeks. On the verified configuration of this data sheet (½ circle cylindrical needle with conical point of 30 mm, USP 0 gauge, violet thread of 75 cm), the following commercial reference from another manufacturer has been cross-checked in primary sources:
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| Johnson & Johnson MedTech (Ethicon) | PDS II® (polydioxanone) violet monofilament, CT-1 needle of ½ circle conical point | Z740D | Comparable by category (not interchangeable 1:1) | Same PDO monofilament material, same USP 0 gauge and violet colour. They differ in needle size (36 mm CT-1 versus 30 mm HR30) and thread length (45 cm versus 75 cm). Confirmed in Ethicon product data sheet and authorised distributors. |
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. As homologous market families to verify code by code before a formal substitution: the PDS II® and PDS Plus® lines from Ethicon (the second with antibacterial triclosan coating, indicated only where its use is approved), the MonoPlus® line from B. Braun/Aesculap (violet polydioxanone monofilament), the Surgicryl® Monofilament PDO line from SMI, DemeDIOX® from DemeTECH and the Maxon line from Medtronic/Covidien. All must be checked in their official technical data sheet (USP gauge, thread length, colour, curvature, body, point and needle size) before proposing a substitution.
5. Selection tips and variants of interest
USP 0 gauge (metric 3.5) provides high initial tensile strength, indicated for medium-high tension tissues, ligatures of medium-calibre vascular structures, closures of muscular planes and general surgery. The ½ circle needle with Taperpoint conical point of 30 mm is the most atraumatic option for soft and well-perfused tissues; if the procedure requires a greater penetration arc, 40 mm needles (HR40) from the same family can be considered, and for finer or paediatric tissues, smaller gauges with needles of smaller radius (HR13 or HR17).
Variants from the same Cahnos catalogue (PDO/PDS Prime family of ½ circle with conical point): AMP418HR1310 (USP 0, HR13 needle of 13 mm, violet, 70 cm); AMP418HR2610 (USP 0, HR26 needle of 26 mm, violet, 70 cm); AMP412HR1720 (USP 2/0, HR17 needle of 17 mm, violet, 75 cm); AM412HR2630 (USP 3/0, HR26 needle of 26 mm, violet, 75 cm); AM413HR4810 (USP 0, HR48 needle of 48 mm, violet, 90 cm) for deep planes; and the double thread presentation AMP413HR4010.2 (USP 0, HR40, 2 x 90 cm). All references are supplied in boxes of 12 units. As this is sterile single-use material, respect the expiry date and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions by volume, hospital supply contracts or equivalences from other commercial houses, contact our technical-commercial team.





Reviews