Description
1. Product Presentation
The absorbable monofilament polydioxanone (PDO/PDS) suture is the reference choice when the tissue needs prolonged support during healing and fully predictable reabsorption. This presentation combines a 70 cm violet grooved thread in USP 3/0 gauge with double needle of ½ circle, cylindrical body and 22 mm taper point conical tip (2xHR22), a configuration designed for continuous suturing techniques, anastomoses and procedures where two active ends are required without repositioning the thread. Available in a box of 12 sterile units, it is a reliable and cost-effective solution for general surgery operating theatres, paediatric cardiovascular surgery units, ophthalmology services and hospital purchasing centres.
2. Technical Explanation
Polydioxanone is a synthetic absorbable polyester —a polymer of poly(p-dioxanone)— that degrades by hydrolysis, a mechanism independent of the patient’s enzymatic activity that guarantees a highly reproducible absorption kinetics between individuals. Being a monofilament, the thread does not present a wick effect or capillarity, which reduces bacterial adherence, facilitates sliding through tissue and minimises the inflammatory response compared to multifilaments. Its firm feel and excellent knotting capacity make it ideal for ligatures and sutures of planes subjected to sustained tension during the critical healing phase.
The ½ circle needle with cylindrical body and conical point (taper point) separates and slides through soft tissue with the least possible trauma, without cutting effect; it is the geometry of choice for vessels, parenchyma and friable tissues. With 22 mm length per active point, the set offers a good balance between reach and precision in medium working fields. The grooved violet colour improves thread visibility on the surgical field. It is supplied sterile by ethylene oxide, single-use and without the possibility of re-sterilisation.
3. Technical Specifications
3.1. General characteristics
| Parameter | Specification |
|---|---|
| Material | Polydioxanone (PDO/PDS), absorbable synthetic monofilament polymer |
| Thread structure | Monofilament, non-woven, grooved |
| Gauge | USP 3/0 (metric 2) |
| Thread length | 70 cm |
| Thread colour | Violet (high visibility) |
| Needle shape | ½ circle (half circumference) |
| Needle body and point | Cylindrical, taperpoint conical point |
| Needle size | 2xHR22 – 22 mm each needle |
| Swaged | Double needle (2x) |
| Absorption mechanism | Hydrolysis (180–210 days) |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal, with excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures, general surgery, paediatric cardiovascular tissues and ophthalmology |
| Presentation | Box of 12 sterile units |
| Factory reference / SKU | 418HR2230-2 / AM418HR2230-2 |
3.2. Tensile strength retention profile
| Time | Tensile strength retention |
|---|---|
| 2 weeks | 75 % |
| 4 weeks | 60–70 % |
| 8 weeks | 45 % |
| Complete absorption | 180–210 days by hydrolysis |
This prolonged force retention profile is maintained during the critical healing phase of slow-healing tissues, which is why its use is recommended in closures of fascia, aponeurosis and structures subjected to sustained traction, as well as in continuous sutures where the double needle-to-needle configuration avoids repositioning the thread.
4. Equivalences and Cross-References
Polydioxanone monofilament is the material basis of the market reference lines for long-lasting absorbable sutures (PDS type). However, the configuration of this reference —double ½ circle needle with 22 mm conical point, USP 3/0 gauge, 70 cm violet thread— does not match any verified commercial code from another manufacturer in all the required fields (material, gauge, double needle configuration, curvature, point family and needle size within tolerance), therefore there is no equivalence table validated by the manufacturer; consult the technical department before use in tenders.
As homologous market families to verify code by code before any substitution: the PDS II® and PDS Plus® lines from Ethicon (Johnson \u0026 Johnson MedTech) —the latter with antibacterial triclosan coating, indicated only where its use is approved; their single needle configurations in 3/0 gauge, such as the CT-2 series of 26 mm or SH, are comparable at category level, not interchangeable 1:1 because they differ in double needle configuration and needle size—, the MonoPlus® line from B. Braun/Aesculap (violet polydioxanone monofilament), the Surgicryl PDO line from SMI and the Maxon line from Medtronic/Covidien. All must be checked in their official technical data sheet (USP gauge, thread length, colour, curvature, body, point, size and number of needles) before substitution. The cited brands belong to their respective manufacturers and are mentioned only as comparative reference.
5. Selection tips and variants of interest
When choosing a PDO suture, always check the USP gauge according to the tissue plane: fine gauges (4/0 to 6/0) for skin and subcutaneous tissue, 3/0 and 2/0 for general soft tissue —like this reference— and USP 0 and above for fascia and planes under tension. Needle geometry is also decisive: the conical point (HR) is indicated for delicate and well-perfused tissues (vessels, parenchyma), while the trocar point (HRT) is reserved for firmer tissues and the reverse cutting triangular point (HS) for fibrous tissues.
The double configuration of this reference (2xHR22) is especially useful in continuous sutures, anastomoses and ligatures where it is desired to have a needle at both ends of the thread. Within the Cahnos catalogue, the PDO/PDS family includes other variants of interest: AM412HR2630 (HR26 needle of 26 mm, USP 3/0, 75 cm) for single needle configurations; AM418HRT2630 (trocar point HRT26, USP 3/0, 70 cm) for more resistant tissues; AMP418HRP3020 (Ultra Plus conical point HRP30, USP 2/0, 70 cm) for high-demand protocols; and AM418HR1350-2 (double needle 2xHR13, USP 5/0, 70 cm) for finer tissues with the same double configuration concept. Always check USP gauge, size and needle geometry against your surgical protocol before substituting one reference for another.
As it is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or other gauges, lengths and colours on request, contact the technical-commercial team of Cahnos.





Reviews