Description
1. Commercial Introduction
The absorbable monofilament polydioxanone suture in Prime quality, reference AMP418HS4002, is the reference solution for the professional who needs a long-lasting thread, high initial strength and fully predictable behaviour in the operating theatre. Its ½ circular needle with reverse cutting triangular point of 40 mm (HS 40) offers firm and controlled penetration even in dense tissues, while the violet grooved thread of 70 cm guarantees optimal visibility and excellent handling in every stitch.
It is a synthetic absorbable medical device, sterile and single-use, with a very competitive price-quality ratio for B2B procurement of hospitals, ophthalmology clinics, private operating theatres and purchasing centres. Presentation in a box of 12 units with a minimum order of 30 boxes, which guarantees the continuous and traceable supply of your healthcare centre.
2. Technical explanation of the product and procedure
Polydioxanone is a synthetic polymer composed of poly(p-dioxanone) that is presented in a monofilament structure with a grooved finish. This single-filament configuration offers an extremely smooth and regular passage through the tissue, without sawing effect or capillarity, which minimises the dragging of microorganisms and tissue reaction, as well as a firm and secure knot tying even in larger diameter gauges such as USP 2.
Absorption occurs by hydrolysis, that is, by controlled degradation of the polymer in contact with organic fluids, over a predictable period of 180 to 210 days. This prolonged absorption profile makes it suitable for tissues that require wound support for several months: abdominal wall, intestinal anastomoses, repair of ligaments and tendons, and paediatric cardiovascular surgery. The ½ circle needle with triangular body and point of reverse cutting (Reverse Cutting) of 40 mm penetrates firm and dense tissues with controlled trauma, being especially suitable in tissue approximation, ligatures, general surgery and ophthalmology. The product is supplied sterile by ethylene oxide, single-use and without the possibility of re-sterilisation.
3. Technical Specifications
3.1. General characteristics of the reference AMP418HS4002
| Material | Polydioxanone (PDO), polymer of poly(p-dioxanone) |
| Thread structure | Monofilament with grooved finish |
| Range / Quality | Prime |
| Gauge | USP 2 (metric 5) |
| Thread colour | Violet |
| Thread length | 70 cm |
| Needle shape | ½ circle |
| Tip type | Triangular with reverse cutting (body and point) |
| Needle size | HS 40 mm |
| Absorption mechanism | Hydrolysis (180-210 days) |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures, general surgery, paediatric cardiovascular surgery and ophthalmology |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
| Reference (SKU) | AMP418HS4002 (Ref. P418HS4002) |
3.2. Tensile strength retention profile
| Time after implantation | Retained tensile strength |
| 2 weeks | 75 % |
| 4 weeks | 60-70 % |
| 8 weeks | 45 % |
| Complete absorption | 180-210 days, by hydrolysis |
Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a fully predictable absorption rate. Sterile by ethylene oxide. Price per box of 12 units.
4. Equivalences and Cross-References
Monofilament polydioxanone is the most widespread long-lasting absorbable suture family in the international market. This reference is clinically positioned in the same category as the following homologous monofilament polydioxanone lines:
| Manufacturer / Brand | Family / Model | Equivalence level | Technical differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | PDS II (polydioxanone) monofilament | Comparable by category (not interchangeable 1:1) | Same polymer (PDO) and violet monofilament structure. Verify USP 2 gauge and HS 40 mm needle in the manufacturer’s catalogue, as well as presentation and declared absorption (182-238 days) before a formal substitution. |
| B. Braun | Monodek (polydioxanone) monofilament | Comparable by category (not interchangeable 1:1) | Same PDO material family. Confirm gauge, ½ circle needle geometry with reverse cutting triangular point and thread length in the manufacturer’s official datasheet. |
| Medtronic | Range of synthetic absorbable monofilament (Biosyn type) | Comparable by category (not interchangeable 1:1) | Long-lasting absorbable monofilament alternative; the copolymer (glycomer) and the absorption profile differ, so it is not interchangeable 1:1. |
| Laboratory Aragó | Range of monofilament polydioxanone with ½ circle needles (HS / TC series) | Comparable by category (not interchangeable 1:1) | Same PDO family with needles of comparable geometry; verify reference, gauge and specific measurements in the authorised distributor’s catalogue. |
There is no equivalence table validated by the manufacturer; consult the technical department before using in tendering.
5. Selection tips and variants of interest
Choose this suture when the tissue requires prolonged wound support (absorption of 6-7 months): abdominal wall, intestinal anastomoses, repair of ligaments and tendons and paediatric cardiovascular surgery. The USP 2 gauge (metric 5) is indicated for tissues that require high initial strength: deep planes, fascia, aponeurosis and high-tension closures. The reverse cutting triangular point is suitable for penetrating firm and dense tissues without tearing them; for soft and friable tissues consider the conical point variants. If healing is shorter (2-4 weeks), consider the PGA or PGLA families from the same catalogue.
| Reference | Variant |
| AMP418HS4002 | PDO Prime USP 2, HS 40 mm needle, violet, 70 cm (current product) |
| AM418HS4002 | PDO Normal quality, same configuration USP 2, HS 40 mm, violet, 70 cm (economical alternative) |
| AM413HS4010 | PDO USP 0, HS 40 mm needle, violet, 90 cm (smaller gauge for moderate tension, more thread) |
| AM413HS4801 | PDO USP 1, HS 48 mm needle, violet, 90 cm (needle with larger radius and intermediate gauge) |
All references are supplied in a box of 12 units; the minimum order is 30 boxes to guarantee the continuous supply of your healthcare centre. As it is a sterile single-use product, check the integrity of the packaging before each procedure, respect the expiry date and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions by volume, hospital supply contracts or equivalences from other commercial houses, contact our technical-commercial team.





Reviews