Description
1. Introduction
When the sterilisation department needs to package small and medium-width instruments —forceps, scissors, osteotomes, sets of 2–3 small pieces— the ULTRA mixed pouch 120 x 200 mm from Pergut Medical Sterimed (SPS Medical ULTRA® range) is the reference packaging solution. This compact format combines exceptional mechanical strength with full visibility of the contents thanks to its transparent face, offering a robust, verifiable response ready for daily use in hospitals, clinics, laboratories and operating theatres. An efficient choice for light and medium packs with ample space for a comfortable and secure seal.
2. Technical Explanation
The ULTRA mixed pouch is a preformed sterile barrier system (SBS) with a mixed configuration: one side consists of a reinforced non-woven polyolefin fibre sheet of 93 g/m², calendered and resistant to tearing and perforation (compliant with EN 868-9), which acts as a porous surface for the exchange of the sterilising agent and as a microbial barrier; the other side is a transparent polyester-polypropylene film that allows inspection of the contents before aseptic opening at the point of use.
The closure is performed with a heat sealer at 145–155 °C and a pressure of 90–120 N/mm², with a seal width greater than 9 mm. The pouch incorporates a type 1 chemical process indicator for steam (ISO 11140-1) printed intrafilm, which changes from pink to brown upon completion of the cycle and allows visual verification of the process. Compared to traditional porous materials, ULTRA technology offers up to 4 times greater puncture resistance (ASTM D3763 test) and excellent peelability, fibre-free and without particle emission, reducing the risks of aseptic failure upon opening. Sterility maintenance has been validated by an independent laboratory for up to 1 year under real conditions in a university hospital, in single and double packaging.
3. Technical Specifications
| Characteristic | Detail |
|---|---|
| Reference (SKU Cahnos) | 98ULT01220 |
| Official manufacturer code (SPS Medical / Pergut Medical) | SUL120X200 (verified in the official manufacturer catalogue and in authorised distribution: Dutscher art. 151507, ddbiolab, Kisker Biotech) |
| Dimensions (width x length) | 120 x 200 mm |
| Packaging type | Flat mixed pouch, without gusset, 100 % peelable, single-use |
| Porous side material | Reinforced non-woven polyolefin fibre, 93 g/m² (EN 868-9) |
| Transparent face material | Transparent polyester-polypropylene (PET/PP) film |
| Sterilisation methods | Steam, ethylene oxide (EO) and hydrogen peroxide (VH2O2); not suitable for dry heat |
| Chemical indicator | Steam, type 1 according to ISO 11140-1, intrafilm printed (pink to brown colour change) |
| Recommended sealing | 145–155 °C, 90–120 N/mm², seal width greater than 9 mm (EN 868-5) |
| Puncture resistance | Up to 4 times higher than conventional porous materials (ASTM D3763) |
| Peelability | 100% peelable, without tearing of the sheet, fibre-free and without particle emission (EN 868-5, annex C) |
| Presentation | Box of 1,200 units |
| Standards and compliance | EN 868-5; ISO 11607-1 and 2; class I medical device with CE marking |
| Use | Single-use pouch; do not re-sterilise; latex-free |
Size programme of the ULTRA range (flat mixed pouches, official Pergut Medical / SPS Medical catalogue):
| Reference | Dimensions (mm) | Units/box |
|---|---|---|
| SUL080X300 (98ULT00830) | 80 x 300 | 2,000 |
| SUL120X200 (98ULT01220, this product) | 120 x 200 | 1,200 |
| SUL120X400 (98ULT01240) | 120 x 400 | 1,000 |
| SUL160X300 (98ULT01630) | 160 x 300 | 1,200 |
| SUL160X500 (98ULT01650) | 160 x 500 | 1,000 |
| SUL220X400 (98ULT02240) | 220 x 400 | 500 |
| SUL220X600 (98ULT02260) | 220 x 600 | 400 |
| SUL250X500 (98ULT02550) | 250 x 500 | 500 |
| SUL285X450 (98ULT02845) | 285 x 450 | 500 |
| SUL285X600 (98ULT02860) | 285 x 600 | 300 |
| SUL320X550 (98ULT03255) | 320 x 550 | 300 |
| SUL380X450 (98ULT03845) | 380 x 450 | 400 |
| SUL380X650 (98ULT03865) | 380 x 650 | 300 |
| SUL380X690 (98ULT03869) | 380 x 690 | 400 |
4. Equivalences and Cross-References
The official manufacturer reference SUL120X200 (SPS Medical / Pergut Medical) has been consistently verified in the official manufacturer catalogue and in independent authorised distribution sources (Dutscher article 151507, ddbiolab, Kisker Biotech), confirming the code as valid for this 120 x 200 mm format in a box of 1,200 units. The internal code of Cahnos 98ULT01220 corresponds to this same item.
As of today, there is no manufacturer-validated cross-reference equivalence table that equates this specific reference with codes from other manufacturers. Therefore, any cross-reference with other brands (for example, reinforced polyolefin mixed pouches from other manufacturers) must be treated as a functional category comparison —same type of reinforced peelable mixed pouch— and not as an exact 1:1 interchangeable equivalent, as the grammage of the porous side, film configuration, pouch geometry and commercial presentation may differ. There is no manufacturer-validated equivalence table; consult the technical department before use in tendering.
Within the ULTRA family from Pergut Medical, the closest variants in geometry are the SUL120X400 (same width, longer format) and the SUL080X300 (narrower width for narrower instruments). For continuous custom packaging, the range includes the ULTRA mixed roll GUL120X070 (120 mm x 70 m).
5. Selection tips and variants of interest
- Choose the length according to the instruments: the 120 x 200 mm format is designed for medium-width and short-to-medium length pieces; for longer instruments consider the SUL120X400 (98ULT01240) and for larger packs the SUL160X300 (98ULT01630).
- Weight and edges: thanks to its puncture resistance (up to 4 times higher), it is the recommended option for heavy devices or those with sharp and pointed elements; for very heavy orthopaedic kits it can be complemented with SMS sheets or wrappers from the same manufacturer.
- Check compatibility: suitable for steam, ethylene oxide and hydrogen peroxide; always check the colour change of the chemical indicator (pink to brown) at the end of the cycle before considering the process valid.
- Custom format: if you work with a heat sealer and variable lengths, the ULTRA roll of 120 mm x 70 m (GUL120X070) allows cutting and sealing to the exact size without waste.
- Sealing: respect the heat sealer parameters (145–155 °C, 90–120 N/mm²) and verify a weld width greater than 9 mm to ensure the integrity of the packaging during the cycle.
- Good practices: store the bags in a cool, dry place (10–30 °C, 30–60 % relative humidity) in their original packaging, record batch, date and contents, and respect the single-use principle (do not re-sterilise).





Reviews
There are no reviews yet.