Description
1. Introduction
The mixed Tyvek® Esterivec roll of 250 mm x 70 m (Ref. 310820) is the packaging solution your sterilisation department needs to protect medical devices processed in low-temperature cycles using gas plasma and hydrogen peroxide (H2O2). Manufactured within the Esterivec range from Pergut Medical, this roll combines a porous side of polyolefin fibre with a Tyvek® base and a side of transparent plastic film, allowing visual verification of the contents and ensuring a reliable sterile barrier until the moment of use. Its presentation in a box of 2 units is designed for continuous supply in sterilisation departments, operating theatres and healthcare services.
2. Technical explanation
Sterilisation by gas plasma and hydrogen peroxide (H2O2) is a low-temperature process indicated for heat-sensitive medical instruments and devices that cannot tolerate steam autoclaves. In these cycles, the sterile barrier system (SBS) must allow the entry and evacuation of the gaseous sterilising agent while preventing the passage of microorganisms once the process is complete.
The Esterivec mixed rolls fulfil this dual function: the microporous polyolefin sheet (Tyvek® base) acts as a microbial filter that allows vaporised H2O2 and gas plasma to diffuse, while the transparent film permits visual inspection of the device without opening the package. They incorporate a Type e1 chemical process indicator (ISO 11140-1) printed under the seals, which changes from red to yellow upon confirmed exposure to the agent, facilitating verification of each cycle. The material is cut to size with the sealer and closed by heat sealing, forming pouches adapted to the length of each device.
3. Technical Specifications
| Characteristic | Specification |
|---|---|
| Material | Microporous polyolefin fibre (Tyvek® base) and transparent plastic film |
| Measurements | 250 mm wide x 70 m long |
| Compatibility | Hydrogen Peroxide (H2O2) and Gas Plasma |
| Process indicator | Chemical indicator Type e1 (ISO 11140-1) – changes from red to yellow |
| Application | Packaging of medical devices for low-temperature chemical sterilisation |
| Format | Flat mixed roll supplied on a reel |
| Presentation | Box of 2 units |
| Regulatory standard | ISO 11607-1/2:2019; EN 868-9:2019 |
| CE marking | Yes – Class I medical device (RD 192/2023 · MDR (EU) 2017/745) |
Technical features of the Esterivec mixed roll range (Pergut Medical technical data sheet):
| Property | Standard | Value |
|---|---|---|
| Polyethylene fibre sheet (Tyvek® base) | ISO 536 | 60.5 g/m² |
| Polyester – polyethylene film (12/50 µm) | ISO 536 | 64 g/m² |
| Film delamination | EN 868-5 Annex A | Exempt |
| Seal width | EN 868-5 annex C | > 9 mm |
| H2O2 – Plasma process indicator (Type e1) | ISO 11140-1 | Red – yellow |
| Dry seal strength | ISO 1924-2 | > 2.5 N/15 mm |
| Wet seal strength | ISO 1924-2 | > 2.2 N/15 mm |
| Peelability (fibre tear) | EN 868-5 annex C | < 10 mm |
| Dimensional tolerance (width) | EN 868-5 annex E | ± 3 mm |
| Dimensional tolerance (length) | EN 868-5 annex E | ± 10 mm |
| Sealing temperature | EN 868-5 · 4.7.a | 115-126 °C |
| Expiry | — | 5 years from date of manufacture |
| Recommended storage | — | 10-30 °C · 30-60 % relative humidity |
Usage recommendations: seal with an approved heat-sealing machine respecting the temperature (115-126 °C) and pressure indicated by the sealer manufacturer, leaving a seal width greater than 9 mm. Store the roll in a cool, dry place, in its original packaging and away from sterilising agents until use.
4. Equivalences and Cross-References
Reference equivalence within the same range: this roll belongs to the Esterivec flat mixed reel range from Pergut Medical, compatible with the main plasma and hydrogen peroxide sterilisers on the market (STERRAD, V-PRO, LTSF and other low-temperature systems). The other widths in the same family (75, 100, 150, 200, 300, 350, 400 and 500 mm, all 70 m) share the same construction and materials, differing only in size, so the selection depends on the size of the device to be packaged.
In the Esterivec range, the official construction of manufacturer references for neighbouring formats has been verified in the Pergut Medical technical data sheet: the 300 mm roll corresponds to references 135310821 (standard Pergut program code) and 18G2EST018 (new factory code), and the 350 mm roll to references 135310824 and 18G2EST021. For the 250 mm width (SKU Cahnos 310820), the official manufacturer code is pending confirmation with Pergut Medical; no unverified code is indicated.
Comparables at category level (not 1:1 interchangeable): due to its Tyvek®/film composition and its H2O2 and plasma indicator, this roll is comparable to other hydrogen peroxide vapour (vH2O2) barrier systems based on Tyvek 1073B with film, such as the Tyvek ranges from Bolsaplast (e.g. MRT1176 and MRT1177) or the Tyvek rolls from Igaltex. The comparison is functional: exact dimensions, Tyvek grammage and manufacturer validation must be confirmed before substituting one reference for another. There is no official equivalence table validated by Pergut Medical for 1:1 cross-referencing with other brands; consult our technical department before use in tenders or when standardising a packaging park.
Do not confuse with other agents: the Esterivec range is specific to low-temperature processes (plasma and hydrogen peroxide). For steam, ethylene oxide or formaldehyde cycles, the mixed paper and film reels from the BOP TI series by Pergut Medical (G2T references) are used, with chemical indicators specific to each agent.
5. Selection tips and variants of interest
Select the roll width according to the size of the device: the packaging must exceed the size of the instrument leaving a minimum margin of 10 mm on each side, so that the seal is not under tension. The 250 mm x 70 m format is suitable for trays, surgical sets, flexible endoscopes and medium-to-large devices, allowing custom pouches to be made and optimising material consumption.
Respect the sealing parameters (temperature 115-126 °C) and check that the seal width is greater than 9 mm to ensure the integrity of the sterile barrier. Keep the roll in a cool, dry place (10-30 °C, 30-60 % relative humidity), in its original packaging and away from sterilising agents until use.
Variants of the same range of 70 m Esterivec rolls available at Cahnos: Ref. 310816 (75 mm), Ref. 310817 (100 mm), Ref. 310818 (150 mm), Ref. 310819 (200 mm), Ref. 310821 (300 mm), Ref. 310824 (350 mm), Ref. 310822 (400 mm) and Ref. 310823 (500 mm). There are also preformed Esterivec mixed pouches (Ref. 310806, 310807 and 310809) for direct packaging without a sealer, as well as CT40 indicator tape for sealing and control of plasma and hydrogen peroxide packs.