Description
1. Product Presentation
The Helix Test PCD Kit KH2X15F1P from the CHEMDYE® range (Pergut Medical) is the process control solution that every sterilisation centre needs to validate low-temperature vaporised formaldehyde (VFBT / LTSF) cycles. It is specifically designed for the safe release of hollow loads, the most demanding and critical load profile of any low-temperature steriliser.
Each kit includes the process challenge device Helix-PCD (reusable for up to 250 cycles) together with 250 self-adhesive test strips with a Type 2 chemical indicator, in sheets of 10 units. A complete system, ready for daily use in hospitals, clinics, pharmaceutical industry and medical device processing companies.
2. Technical explanation of the procedure
The Helix-PCD is a process challenge device (PCD) that reproduces the worst-case scenario for sterilant penetration in lumen instruments: the vaporised formaldehyde must travel through a plastic tube 1.5 m long with only 2 mm internal diameter before reaching the chemical indicator. The highly durable plastic capsule houses the indicator strip, and the special silicone connector, covered by an external connector, ensures an airtight seal and a secure connection to the equipment.
The PCD-A-F1 strip is a Type 2 chemical indicator (ISO 11140-1:2014) printed with indicator ink that changes from burgundy to green only when the formaldehyde cycle has been carried out under correct conditions. If the 4 bars on the strip reach the final colour, the result is successful and the equipment is working correctly; if one or more bars do not change colour, this indicates poor penetration of the sterilant and therefore a malfunction of the steriliser. The product is free of toxic heavy metals.
3. Technical Specifications
Technical data sheet for the KH2X15F1P kit:
| Characteristic | Specification |
| Product type | Helix-PCD kit + test strips with chemical indicator for formaldehyde |
| Reference | KH2X15F1P (KH2X15-F1/P) |
| Processing | Sterilisation by low-temperature vaporised formaldehyde (VFBT / LTSF processes) |
| Application | Control of formaldehyde sterilisation processes; safe release of hollow loads |
| Helix-PCD | Highly durable plastic capsule, plastic tube 1.5 m x 2 mm internal diameter and special silicone connector with external connector |
| Chemical indicator | PCD-A-F1, Type 2 (ISO 11140-1:2014) |
| Indicator strip | Self-adhesive, 76 x 6 mm, printed with indicator ink; 10 strips per sheet (sheet of 80 x 98.5 mm) |
| Colour change | Burgundy (initial) → Green (final); correct result if the 4 bars change |
| Presentation | 250 PCD-A-F1 strips + 1 Helix-PCD device per box |
| Strips service life | 5 years |
| Helix-PCD service life | 250 cycles |
| Storage | T 10–30 ºC, RH 30–80 %, protected from light; do not store near sterilising agents |
| Regulatory standard | ISO 13485:2016, EN 867-5, EN 14180, ISO 25424, ISO 11140-1:2014 |
| Markets | Healthcare, food, pharmaceutical and medical products industries |
4. Equivalences and Cross-References
This kit belongs to the Helix process challenge device family from Pergut Medical. The following in-house references are available within the same Helix-PCD system:
| Reference | Description | Presentation |
| KH2X15F1P | Helix Test PCD kit + strips for formaldehyde (VFBT) – current product | 250 strips/box |
| KH2X15-35 | Helix Test PCD kit + simulation strips for steam | 250 strips/box |
| KH2X12-P1 | Helix Test PCD kit + strips for plasma (hydrogen peroxide) | 100 strips/box |
| KH2X15-35B | B&D type Helix test for steam (metallic device, ISO 11140-4) | On request |
| BD125X/20 | Bowie-Dick type Helix test for steam (1.5 m x 2 mm) | 20 uds/caja |
| 24102 | Bowie-Dick test pack for daily control (120 x 120 x 17 mm) | 20 uds/caja |
As complementary consumables for formaldehyde processes, Pergut Medical offers the multivariable chemical indicator CD50 (Type 4, 105 x 18 mm, for 2% and 3% formaldehyde cycles), the indicator adhesive tape CT50 for sealing wrappers, and the biological control BT100 for formaldehyde. On the international market, functionally equivalent kits are marketed under the nomenclature KH2X15-F1/P (MGM Medix, Terragene, GKE Healthcare) with identical conformity to ISO 11140-1 Type 2 and EN 867-5.
5. Selection tips and variants of interest
Hollow load control: if your low-temperature steriliser processes endoscopes, laparoscopes, cannulas or other long-lumen instruments, the Helix-type test is the right choice. The KH2X15F1P kit is the specific variant for VFBT formaldehyde; it must not be used in steam, ethylene oxide, dry heat or hydrogen peroxide cycles.
Other processes: to validate steam cycles use reference KH2X15-35 and for plasma (vH2O2) use KH2X12-P1. The kit also allows you to verify penetration of the agent at the worst point of the load before batch release.
Frequency of use: perform the test periodically following the equipment manufacturer’s protocol and the ISO 11140-1 standard. The Helix-PCD supports up to 250 cycles; replace the PCD-A-F1 strips when they run out or always verify that the colour change of the 4 bars reaches the final green.
Storage and handling: keep the product in its original packaging, protected from light, in a cool, dry place, away from sterilising agents; respect the 5-year expiry of the strips and check the integrity of the packaging before each cycle.
Professional B2B purchase: for sterilisation centres, hospitals, pharmaceutical industry and processing companies, contact our technical-commercial team for volume conditions and continuous supply.