Description
1. Product Presentation
The ready-to-use UPSTAR Bowie & Dick test is the practical solution that every sterilisation department needs to quickly and reliably verify the correct removal of air and the penetration of saturated steam in steam autoclaves. Supplied as a single-use test pack, it requires no prior assembly: it is ready to be placed directly in the steriliser chamber and to check at the end of the cycle whether the process conditions have been adequate. It combines high sensitivity, stable colour change and immediate, objective interpretation of the result, ideal for daily routine monitoring in hospitals, clinics and medical device processing companies.
2. Technical explanation of the procedure
The Bowie & Dick test is a steam penetration test that checks the air removal efficiency of steam sterilisers with dynamic air removal (pre-vacuum and vacuum pump). If residual air pockets remain in the chamber, the steam does not reach the load under saturated conditions and the sterilisation process may not be effective, even if the temperature and time records are correct.
The UPSTAR pack consists of a test sheet located in the centre of several layers of cardboard. This sheet is coated with a thermoreactive chemical pigment that changes colour according to three parameters: temperature (134 °C), time (3 min 30 s) and presence of saturated steam. When the cycle runs correctly, the result is a uniform green colour; if the time, temperature or steam pressure are incorrect, a central red colour appears on the sheet, indicating that the load must not be released and that the equipment must be checked.
3. Technical Specifications
Product data
| Manufacturer | GIMA |
|---|---|
| Sales unit | 1 pc |
| Device type | It is not a medical device (no medical device) |
| EAN | 3760243050039 |
Technical data sheet for the UPSTAR test (Ref. 35853)
| SKU | 35853 |
|---|---|
| Model | UPSTAR – Bowie & Dick Indicator ready for use |
| Product type | Bowie & Dick test pack ready to use (single-use) |
| Process controlled | Steam sterilisation in autoclaves |
| Parameters verified | Temperature: 134 °C · Time: 3 min 30 s · Saturated steam |
| Composition | Test sheet coated with a thermoreactive chemical pigment located in the centre of several layers of cardboard |
| Satisfactory result | Uniform green colour |
| Unsatisfactory result | Central red colour: incorrect time and/or temperature and/or steam pressure |
| Features | Stability of the colour change · High sensitivity, detects any anomaly |
| Packaging | Impermeable aluminium pouch |
| Application | Control of the correct operation of steam autoclaves and sterilisation of porous devices (compresses, drapes, gowns, woven or non-woven fabrics) |
4. Equivalences and Cross-References
This pack belongs to the family of steam penetration tests (Bowie & Dick) for routine autoclave monitoring. In the Cahnos catalogue it is linked to the following references of interest from the same family of process controls:
| Reference | Description | Presentation |
|---|---|---|
| BD125X/20 | Bowie-Dick Helix-type test (lumen 1.5 m x 2 mm) for hollow load | 20 uds/caja |
| 24102 | Bowie-Dick test package for daily control (120 x 120 x 17 mm) | 20 uds/caja |
| BD125X/1 | Bowie-Dick test package with pre-notice sheet | 20 uds/caja |
| KH2X15-35B | Bowie & Dick Helix-type test for steam (metal device) | On request |
| KH2X15-35 | Helix Test PCD kit + simulation strips for steam | 250 strips/box |
| KH2X12-P1 | Helix Test PCD kit + strips for plasma (hydrogen peroxide) | 100 strips/box |
| KH2X15F1P | Helix Test PCD kit + strips for vaporised formaldehyde | 250 strips/box |
At a functional level, this test is interchangeable with other Bowie & Dick test packs conforming to standard process control practice for the validation of air removal in steam autoclaves with vacuum pump. To confirm an exact 1:1 equivalence in a tender or specification, consult our technical-commercial department.
5. Selection tips and variants of interest
- Daily monitoring of steam autoclaves: carry out the Bowie & Dick test at least once a day, with an empty chamber and following the equipment manufacturer’s protocol and the ISO 11140-1 standard. The UPSTAR pack is ready to use, which streamlines the sterilisation department’s routine.
- Validation of porous loads: this test is the one indicated for verifying the sterilisation of compresses, drapes, gowns and fabrics (woven or non-woven), the load profile for which the Bowie & Dick test is designed.
- Hollow load: if your autoclave processes long-lumen instruments (endoscopy, laparoscopy, cannulas, etc.), consider the Helix-type test (BD125X/20) as a complement for the worst-case penetration scenario.
- Other processes: for the validation of hydrogen peroxide/gas plasma or vaporised formaldehyde cycles, the references KH2X12-P1 and KH2X15F1P are available, specific to low temperature.
- Handling and storage: keep the pack in its original impermeable aluminium pouch until the moment of use. It is a single-use product: it must not be re-sterilised. Always check the integrity of the pouch before each cycle.
- Professional B2B purchasing: for sterilisation departments, hospitals and processing companies, contact our technical-commercial team for volume conditions and continuous supply.